Installation
The chiller requires a three-phase 400 V supply with a 3x20 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply, cable cross-section and protection device needed before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
The unit is water-cooled and needs a cooling water supply with at least 1 bar differential pressure and no more than 6 bar.
Cooling water pressure must be verified on site to guarantee rated cooling performance and avoid damage.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water inlet temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation occurs.
Out-of-range cooling water temperature reduces cooling power and can cause condensation and corrosion.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops from 10 kW at 0 °C to 3 kW at -20 °C, so low-temperature duty must be checked against the load.
Ensures the selected chiller can actually remove the process heat at the lowest required operating temperature.
Source
“at -20°C 3 kW”
Performance
The circulation pump delivers 96 l/min at zero head but only 24 l/min at 5.0 bar, so system pressure drop must be matched to the pump curve.
Prevents undersizing of flow in circuits with high pressure loss and confirms achievable flow at the application's back pressure.
Source
“max. delivery 96 l/min”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 1.75 kg corresponding to 2.4 t CO2 equivalent.
Affects refrigerant handling, leak reporting obligations and room ventilation requirements.
Source
“R-449A (A1, H280)”
Operation
The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.
Defines the acceptable installation environment and excludes wet or dusty locations without additional protection.
Source
“max. ambient temperature 40 °C”
Installation
The chiller needs a minimum filling capacity of 6.5 l and has 5.5 l expansion volume, which sets the heat transfer fluid charge.
Determines the volume of heat transfer fluid to prepare and the expansion tank sizing for the loop.
Source
“min. filling capacity 6,5 l”
Installation
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality selection to avoid corrosion and fouling of the wetted materials.
Source
“Please use suitable cooling water.”
Cumplimiento
Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and combined deviations outside this are not allowed.
Protects the compressor and electronics from supply quality issues at the installation site.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices below 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.
Avoids delays at commissioning because the customer must supply the mains connector.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit emits a sound pressure level of 65 dB(A) ±4 dB(A), which must be considered for the installation location.
Relevant for workplace noise limits and placement near occupied areas.
Source
“sound pressure level +/- 4 dB(A) 65 dB(A)”
Selection
The chiller has a safety classification of III / FL according to DIN 12876 and adjustable overtemperature protection.
Confirms the safety class and protective function required for flammable or sensitive bath fluids.
Source
“With adjustable overtemperature protection according to DIN 12876.”