Induztr
Resumen técnico Visor de documentos Modelos relacionados

Unichiller 110T Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller 110T chiller and its main features
  2. Spec pp. 1-2
    Lists performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/20

Technical data according to DIN 12876 — Unichiller 110T (Order-No.: 3027.0078.01)

ParameterValue
Operating temperature range-10...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
Temperature stability at -10°C0,5 K
Safety classificationI / NFL
Cooling power at 15°C11 kW
Cooling power at 0°C6 kW
Cooling power at -10°C2,7 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pump:C3
max. delivery90 l/min
max. delivery pressure5,6 bar
Delivery at 0,5 bar87 l/min
Delivery at 1,0 bar81 l/min
Delivery at 1,5 bar76 l/min
Delivery at 2,0 bar70 l/min
Delivery at 2,5 bar64 l/min
Delivery at 3,0 bar58 l/min
Delivery at 3,5 bar50 l/min
Delivery at 4,0 bar42 l/min
Delivery at 4,5 bar33 l/min
Delivery at 5,0 bar22 l/min
Pump connectionG1 1/4 male
min. filling capacity8,36 l
Volume of expansion7 l
Overall dimensions WxDxH **600x790x1614 mm
Net weight230 kg
sound pressure level +/- 4 dB(A)74 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI

** см. сноску в оригинале (Overall dimensions WxDxH **)

Engineering Facts

12 facts · every claim linked to a page
Selection

The chiller is intended only for externally closed applications, so the process loop must be closed and not open to atmosphere.

Determines whether the unit can be used with an open bath or only with a closed external circuit.

Source
“For externally closed applications.”
Installation

The unit requires a three-phase 400 V 3~ 50 Hz supply, so site power must be three-phase.

Electrical supply must be verified before ordering and installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and install the plug.

Affects electrical installation scope and requires a suitable plug on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and some combinations are not allowed.

Supply quality must be checked; out-of-tolerance combinations can invalidate operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

The unit must be installed where ambient temperature stays between 5 °C and 40 °C.

Ambient conditions affect cooling capacity and safe operation.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises.

Selection must account for actual site ambient temperature, not just nominal capacity.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The pump connection is G1 1/4 male, so mating hoses or fittings must match this thread.

Ensures correct process plumbing connection.

Source
“Pump connection G1 1/4 male”
Installation

Space requirements must be respected; dimensions are 600x790x1614 mm and the unit weighs 230 kg.

Needed for floor loading, access, and clearance planning.

Source
“Please respect space requirements.”
Safety

The refrigerant is R-449A with GHS classification A1, H280, and GWP 1397.

Refrigerant safety and environmental data are needed for risk assessment and compliance.

Source
“R-449A (A1, H280)”
Cumplimiento

The equipment is EMC Class A, Group 1, which may require additional measures in residential or sensitive environments.

Affects installation location and EMC compatibility with nearby equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

The unit has IP20 protection, so it must be protected from water and dust ingress.

Determines suitable environment and cleaning/installation precautions.

Source
“Degree of Protection IP20”
Maintenance

The expansion tank is atmospheric open and has a cover as standard equipment, which must be handled correctly during operation.

Open expansion tank design affects filling, venting, and maintenance.

Source
“cover expansion tank*”