Installation
The chiller requires a three-phase 400 V supply, so the site must provide a 3~ 50 Hz feed.
Determines electrical infrastructure and whether a suitable three-phase supply is available at the installation point.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the plug.
Affects electrical installation planning and the need to procure a compatible plug or hard-wired connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit is IP20, so it must be installed in a dry indoor location protected from water and dust ingress.
Restricts installation environment and rules out washdown or outdoor locations.
Source
“Degree of Protection IP20”
Operation
Ambient temperature must stay between 5 °C and 40 °C for valid operation.
Defines the allowable room conditions and warns that cooling capacity may drop if ambient rises.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance
Cooling capacity is rated at 20 °C room temperature and may decrease at higher ambient temperatures.
Prevents oversizing assumptions and requires derating when the installation room is warmer than 20 °C.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Voltage and frequency tolerances are ±5% and ±2% respectively, but not all combinations are allowed.
Guides power quality assessment and avoids invalid supply combinations that could damage the unit.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Cumplimiento
The equipment emits Class A, Group 1 disturbances per EN55011, so EMC measures may be needed in residential-like areas.
Affects site EMC planning and suitability for sensitive or residential environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety
The refrigerant is R-449A with GHS classification H280, so refrigerant handling and ventilation precautions apply.
Informs safety planning for refrigerant storage, leak response, and room ventilation.
Source
“R-449A (A1, H280)”
Selection
The pump connection is G1 1/4 male, so external piping must match this thread.
Determines the hose couplings and adapters needed for the external circuit.
Source
“Pump connection G1 1/4 male”
Selection
The minimum filling capacity is 14 l and the expansion volume is 39 l, which sets the external loop volume limits.
Helps size the external circuit and expansion tank requirements.
Source
“min. filling capacity 14 l Volume of expansion 39 l”
Installation
The unit is air-cooled and requires respect of space requirements around it.
Ensures adequate airflow and service access, which affects cooling performance and maintenance.
Source
“Please respect space requirements.”
Operation
The unit is intended for externally closed applications, so the external circuit must be closed.
Prevents use with open baths or unsealed loops that could cause contamination or overflow.
Source
“For externally closed applications.”