Introduces the Unichiller 200T chiller unit and its main features
Specp. 2
Lists technical data according to DIN 12876 standard
p. 1table, 34 rows
Technical data according to DIN 12876 — Unichiller 200T (Order-No.: 3028.0148.01)
Characteristic
Value
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
Temperature stability at -10°C
0,5 K
Safety classification
I / NFL
Cooling power at 15°C
20 kW
Cooling power at 0°C
10 kW
Cooling power at -10°C
5 kW
Cooling power at -20°C
3 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Circulation pump
D3
max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 1,1 bar
202 l/min
Delivery at 1,5 bar
193 l/min
Delivery at 2,0 bar
179 l/min
Delivery at 2,5 bar
159 l/min
Delivery at 3,0 bar
140 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
77 l/min
Delivery at 4,5 bar
31 l/min
Pump connection
G1 1/4 male
min. filling capacity
14 l
Volume of expansion
39 l
Overall dimensions WxDxH **
905x1582x1902 mm
Power supply (3 Phase)
460V 3~ 60Hz
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
** Overall dimensions WxDxH
Engineering Facts
13 facts · every claim linked to a page
Selection
The chiller's cooling capacity drops as ambient temperature rises, so size the unit for the actual room temperature rather than the 20 °C catalogue values.
Prevents undersizing when the installation room runs warmer than 20 °C.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
The unit requires an external branch circuit protection device per UL 489, which must be provided by the installer.
Affects electrical panel design and compliance of the installation.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Three-phase units drawing less than 63 A are shipped with a cable but without a plug, so a suitable plug must be sourced on site.
Avoids surprises at commissioning when no mating plug is available.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances must be met simultaneously, and certain combinations such as -5% voltage with +2% frequency are not permitted.
Guides verification of the site power quality before connecting the chiller.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit is rated IP20, so it must be installed in a dry indoor location protected from water and dust ingress.
Determines suitable installation environment and enclosure requirements.
Source
“Degree of Protection
IP20”
Operation
The chiller is intended only for externally closed applications, which constrains the hydraulic circuit design.
Prevents use in open-loop or vented process circuits that would void the design intent.
Source
“For externally closed applications.”
Operation
Ambient temperature must stay between 5 °C and 40 °C for correct operation.
Defines the allowable room conditions for reliable operation.
Source
“min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.
Informs site safety assessment and service procedures for the refrigeration circuit.
Source
“R-449A (A1, H280)”
Cumplimiento
The equipment emits Class A, Group 1 disturbances per EN55011, which typically requires industrial-environment EMC measures.
Affects EMC planning and suitability for residential or sensitive environments.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Cumplimiento
The unit complies with US-SNAP for industrial process refrigeration end-use, which matters for US installations.
Confirms regulatory acceptability for the intended US application.
Source
“The US-SNAP end-use for this equipment is the industrial process refrigeration.”
Maintenance
The expansion tank is atmospherically open and has a cover included, so level checks and venting are part of routine upkeep.
Signals maintenance tasks and the need to keep the tank cover in place.
Source
“atmospheric open expansion tank”
Selection
The pump delivers 202 l/min at 1.1 bar but only 77 l/min at 4.0 bar, so system pressure drop must be matched to the pump curve.
Ensures the selected pump curve covers the actual external circuit resistance.
Source
“Delivery at 4,0 bar
77 l/min”
Installation
Overall dimensions are 905x1582x1902 mm and space requirements must be respected, so plan adequate clearance around the unit.
Prevents installation in spaces too tight for service access and airflow.