Product introduction and main features of the chiller
Specpp. 1-2
Technical data, performance and dimensions per DIN 12876
p. 2table, 14 rows
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Consumption at water 15°C, flow -20°C
445 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
15 l
Volume of expansion
22 l
Overall dimensions WxDxH **
760x800x1615 mm
Net weight
374 kg
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
17,5 A
Fuse (3 phase)
3x20A
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S361439
** Please respect space requirements. See operating conditions at www.huber-online.com
Technical data according to DIN 12876 — Unichiller 200Tw-H (Order-No.: 3028.0113.01)
Parameter
Value
Operating temperature range
-20...100 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
III / FL
Heating power
2 kW
Cooling power at 15°C
20 kW
Cooling power at 0°C
11 kW
Cooling power at -10°C
5 kW
Cooling power at -20°C
3 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Global Warming Potential (GWP)
1397
Refrigerant quantity
2,35 kg
CO2 equivalent
3,3 t
Circulation pump
D3
max. delivery
210 l/min
max. delivery pressure
4,7 bar
Delivery at 1,1 bar
202 l/min
Delivery at 1,5 bar
193 l/min
Delivery at 2,0 bar
179 l/min
Delivery at 2,5 bar
159 l/min
Delivery at 3,0 bar
140 l/min
Delivery at 3,5 bar
115 l/min
Delivery at 4,0 bar
77 l/min
Delivery at 4,5 bar
31 l/min
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 15°C
725 l/h
Consumption at water 15°C, flow 0°C
860 l/h
Consumption at water 15°C, flow -10°C
610 l/h
Engineering Facts
13 facts · every claim linked to a page
Installation
The chiller is water-cooled and requires a cooling water supply with a differential pressure between 1 and 6 bar, so the site must provide a suitable cooling water circuit.
Determines whether the facility's cooling water supply is adequate and what pressure regulation may be needed.
Source
“min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar”
Installation
The unit is designed for cooling water intake between 5°C and 20°C, and if the cooling water falls below the condensation point the cooling water pipes inside the unit must be insulated.
Defines the allowable cooling water temperature window and a mandatory insulation measure to prevent condensation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase, so performance must be derated for warmer cooling water.
Prevents oversizing assumptions and ensures the cooling water supply can handle increased flow at higher temperatures.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate
possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility to avoid corrosion or damage to the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento
Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is not allowed.
Ensures the electrical supply stays within the permitted combined tolerance envelope to avoid malfunction.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices drawing less than 63 A are supplied with a cable but without a plug, so a suitable plug must be provided on site.
Affects electrical installation planning and the need to procure and fit a compatible plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The refrigerant R-449A is classified A1 with GHS hazard H280, and the unit contains 2.35 kg corresponding to 3.3 t CO2 equivalent.
The unit requires a minimum filling capacity of 15 l and has an expansion volume of 22 l, which must be accounted for when preparing the external circuit.
Determines the minimum heat transfer fluid volume and expansion tank sizing for the external loop.
Source
“min. filling capacity
15 l
Volume of expansion
22 l”
Installation
The unit has IP20 protection and must be operated between 5°C and 40°C ambient temperature, so it needs a dry, climate-controlled location.
Constrains the installation environment and prevents ingress of water or dust that IP20 does not protect against.
Source
“Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C”
Installation
The unit is 760x800x1615 mm and weighs 374 kg net, and space requirements must be respected per the operating conditions.
Needed for floor loading, transport route and clearance planning during installation.
Source
“** Please respect space requirements. See operating conditions at www.huber-online.com”
Operation
The pump delivers 210 l/min maximum at 4.7 bar, but flow falls sharply at higher pressure (e.g. 31 l/min at 4.5 bar), so system pressure drop must be matched to the pump curve.
Ensures the external circuit pressure drop does not starve flow and that the pump operating point is acceptable.