Unichiller 230-H Technical Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller 230-H chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 9 rows
  4. p. 1 table, 35 rows

Technical data according to DIN 12876

ParameterValue
Net weight477 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)14 A
Fuse (3 phase)3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S309173

Technical data according to DIN 12876

ParameterUnichiller 230-H
Order-No.3039.0018.01
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C
at 100°C23 kW
at 20°C23 kW
at 15°C21 kW
at 0°C13,5 kW
at -10°C9 kW
at -20°C5,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,49 kg
CO2 equivalent2,1 t
Circulation pump:E1
max. delivery54 l/min
max. delivery pressure3,5 bar
Förderleistung bei 0,17 bar52 l/min
Delivery at 0,5 bar48 l/min
Delivery at 1,0 bar42 l/min
Delivery at 1,5 bar36 l/min
Delivery at 2,0 bar28 l/min
Delivery at 2,5 bar21 l/min
Delivery at 3,0 bar11 l/min
Pump connectionG1 1/4 male
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **938x1288x2003 mm

** см. сноску в оригинале (в тексте страницы пояснение к ** отсутствует)

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply with a 16 A fuse per phase, so the site electrical infrastructure must be sized accordingly.

Determines the electrical connection, cable sizing and breaker rating needed before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the site plug.

Affects installation planning and the bill of materials for the electrical hook-up.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit is rated IP20, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts the permissible installation environment and enclosure requirements.

Source
“Degree of Protection IP20”
Operation

The chiller may only be operated between 5 °C and 40 °C ambient, and cooling capacity drops if ambient rises above 20 °C.

Sets the allowable room conditions and warns that published cooling capacities are only valid at 20 °C ambient.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The unit is air-cooled and needs sufficient free space around it, so the installation layout must respect the manufacturer's space requirements.

Prevents derating or overheating caused by restricted air flow in a cramped installation.

Source
“Please respect space requirements. See operating conditions at www.huber-online.com”
Cumplimiento

Voltage and frequency tolerances are coupled: ±5 % voltage is only allowed together with ±2 % frequency, and certain combinations are prohibited.

Must be checked against the site power quality to avoid invalid supply conditions and warranty issues.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280 (pressurised gas), and the charge is 1.49 kg with 2.1 t CO2 equivalent.

Drives refrigerant handling, labelling, leak-check and F-gas reporting obligations.

Source
“R-449A (A1, H280)”
Safety

The chiller is safety class III / FL with adjustable overtemperature protection per DIN 12876, which is relevant for protecting the external process.

Confirms the built-in safety concept and the need to set the overtemperature limit for the application.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Performance

Cooling capacity falls steeply at low temperatures, from 23 kW at 20 °C to 5.5 kW at -20 °C, so the load must be checked at the lowest operating point.

Prevents oversizing assumptions based on the 20 °C figure when the process runs cold.

Source
“at -20°C 5,5 kW”
Performance

The circulation pump delivers 54 l/min at zero head but only 11 l/min at 3.0 bar, so the system pressure drop must be matched to the pump curve.

Ensures the external circuit resistance does not starve the required flow rate.

Source
“Delivery at 3,0 bar 11 l/min”
Installation

The pump connection is G1 1/4 male, so hoses and fittings must match this thread size.

Determines the connection hardware needed for the external circuit.

Source
“Pump connection G1 1/4 male”
Operation

The unit has a minimum filling capacity of 18 l and an expansion volume of 48 l, which sets the heat transfer fluid charge and expansion allowance.

Needed to size the fluid fill and confirm the expansion tank is adequate for the loop.

Source
“min. filling capacity 18 l”
Selection

The chiller is intended for externally closed applications, so an open external circuit is not a valid configuration.

Rules out use with open baths or unsealed external loops during selection.

Source
“For externally closed applications.”
Cumplimiento

The equipment is classified EN55011 Class A Group 1, which is industrial EMC classification and may require additional measures in residential areas.

Affects EMC planning and suitability of the installation location.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”