Unichiller 700Tw-H Technical Datasheet

Contenido

  1. Overview p. 1
    Product introduction and general description of the chiller unit
  2. Spec pp. 1-2
    Technical data, performance figures and operating parameters per DIN 12876
  3. p. 2 table, 10 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
min. filling capacity12,7 l
Volume of expansion7 l
Overall dimensions WxDxH **1000x1100x1635 mm
Net weight634 kg
Power supply (3 Phase)400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/21

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at 15°C70 kW
Cooling power at 0°C50 kW
Cooling power at -10°C30 kW
Cooling power at -20°C20 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Circulation pumpD3
max. delivery234 l/min
max. delivery pressure4,9 bar
Delivery at 1,3 bar197 l/min
Delivery at 1,5 bar192 l/min
Delivery at 2,0 bar178 l/min
Delivery at 2,5 bar158 l/min
Delivery at 3,0 bar137 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar86 l/min
Delivery at 4,5 bar40 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1 1/4 male
Consumption at water 15°C, flow 15°C3780 l/h
Consumption at water 15°C, flow 0°C2940 l/h
Consumption at water 15°C, flow -10°C1980 l/h
min. cooling water differential pressure1,5 bar
max. cooling water pressure6 bar

Order-No.: 3032.0006.01; Unichiller 700Tw-H

Engineering Facts

14 facts · every claim linked to a page
Installation

The chiller is water-cooled and requires cooling water inlet temperature between 5°C and 20°C, with insulation of cooling water pipes mandatory if below the condensation point.

Determines cooling water supply requirements and pipe insulation needs during installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water differential pressure must be at least 1.5 bar and must not exceed 6 bar.

Defines the required cooling water supply pressure range for proper operation.

Source
“min. cooling water differential pressure 1,5 bar”
Installation

Three-phase devices with current consumption less than 63A are supplied with cable but without plug, so the customer must provide a suitable plug.

Affects electrical installation planning and procurement of the mating plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination of -5% voltage with +2% frequency is not allowed.

Ensures the site power supply stays within safe limits for the compressor and electronics.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit has IP20 protection and must be installed in an environment with ambient temperature between 5°C and 40°C.

Restricts installation location to dry indoor areas with adequate ventilation and temperature control.

Source
“Degree of Protection IP20”
Selection

Cooling capacity drops as cooling water temperature rises, and cooling water flow rate may need to increase accordingly.

Affects sizing of the cooling water supply and expected performance at site conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Selection

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality requirements to avoid corrosion and fouling of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Safety

The refrigerant R-449A has safety group A1 and GHS hazard H280, and the unit has safety classification III / FL.

Informs risk assessment, ventilation requirements and refrigerant handling procedures.

Source
“R-449A (A1, H280)”
Safety

The unit includes adjustable overtemperature protection according to DIN 12876.

Provides a safety function that must be configured correctly for the application.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Operation

The unit is intended for externally closed applications and has an atmospheric open expansion tank.

Defines the permissible hydraulic circuit configuration and prevents misuse with open systems.

Source
“For externally closed applications.”
Operation

The pump connection and cooling water connection are both G1 1/4 male.

Specifies the thread type for hose couplings and piping adapters.

Source
“Pump connection G1 1/4 male”
Performance

Cooling power is 70 kW at 15°C, 50 kW at 0°C, 30 kW at -10°C and 20 kW at -20°C, valid at 20°C room temperature and 15°C cooling water inlet.

Allows correct sizing of the chiller for the required heat load at the target temperature.

Source
“at 15°C 70 kW”
Performance

Maximum pump delivery is 234 l/min and maximum delivery pressure is 4.9 bar, with delivery dropping at higher pressure.

Helps verify that the pump can overcome the pressure drop of the external circuit.

Source
“max. delivery 234 l/min”
Cumplimiento

The equipment is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.

Indicates EMC class and potential need for filtering or separation from sensitive equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”