Unichiller 800Tw-H Datasheet

Contenido

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data and performance specifications per DIN 12876
  3. p. 2 table, 9 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Net weight851 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)50 A
Fuse (3 phase)3x63A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S433261

Technical data according to DIN 12876 — Unichiller 800Tw-H (Order-No.: 3076.0004.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at 40°C80 kW
Cooling power at 15°C80 kW
Cooling power at 0°C60 kW
Cooling power at -10°C40 kW
Cooling power at -20°C20 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity5 kg
CO2 equivalent7,0 t
Circulation pumpD3
max. delivery196 l/min
max. delivery pressure5 bar
Delivery at 1,0 bar175 l/min
Delivery at 2,0 bar150 l/min
Delivery at 3,0 bar122 l/min
Delivery at 4,0 bar74 l/min
Pump connectionG1 1/4 male
Cooling water connectionG1 1/4 male
Consumption at water 15°C, flow 15°C3480 l/h
min. cooling water differential pressure2,5 bar
max. cooling water pressure6 bar
min. filling capacity12,7 l
Volume of expansion7 l
Overall dimensions WxDxH **1000x1600x1620 mm

** Overall dimensions WxDxH

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply with a 63 A fuse per phase, so the site electrical infrastructure must be sized accordingly.

Determines upstream breaker/fuse rating and cable sizing for installation.

Source
“Fuse (3 phase) 3x63A”
Installation

The unit is water-cooled and needs a cooling water differential pressure of at least 2.5 bar and no more than 6 bar at the connections.

Cooling water supply pressure must be verified before installation to avoid insufficient cooling or overpressure damage.

Source
“min. cooling water differential pressure 2,5 bar max. cooling water pressure 6 bar”
Installation

Cooling water inlet temperature must be kept between 5 °C and 20 °C, and pipes must be insulated if condensation can occur.

Defines the acceptable cooling water supply conditions and condensation protection measures.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity is rated at 80 kW at 40 °C and drops to 20 kW at -20 °C, so capacity must be checked at the actual operating temperature.

Prevents oversizing or undersizing the chiller for the required process temperature.

Source
“at 40°C 80 kW at 15°C 80 kW at 0°C 60 kW at -10°C 40 kW at -20°C 20 kW”
Performance

The circulation pump delivers 196 l/min maximum and 74 l/min at 4.0 bar, so system pressure drop must be matched to the pump curve.

Ensures the pump can overcome the external circuit pressure drop at the required flow.

Source
“max. delivery 196 l/min max. delivery pressure 5 bar Delivery at 1,0 bar 175 l/min Delivery at 2,0 bar 150 l/min Delivery at 3,0 bar 122 l/min Delivery at 4,0 bar 74 l/min”
Installation

The unit has IP20 protection and must be installed in an ambient temperature between 5 °C and 40 °C.

Restricts installation location to indoor, dry environments within the specified ambient range.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Safety

The refrigerant R-449A has a GHS hazard statement H280, so refrigerant handling and ventilation precautions are required.

Informs site safety planning for refrigerant containment and emergency procedures.

Source
“R-449A (A1, H280)”
Cumplimiento

The equipment emits Class A, Group 1 disturbances under EN55011, so it may require EMC mitigation in residential or sensitive environments.

Class A equipment is not intended for residential use without additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

Leakage current exceeds 3.5 mA, so protective earth connection and residual current device selection must account for this.

Affects earth bonding and RCD type/rating selection during electrical installation.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so a suitable plug must be provided on site.

Prevents delays at commissioning due to missing power connector.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Defines acceptable supply quality for reliable operation and warranty compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to avoid corrosion or scaling in the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Selection

The minimum filling capacity is 12.7 l and expansion volume is 7 l, so the external circuit volume must be considered for proper expansion tank sizing.

Ensures the expansion vessel can accommodate thermal expansion of the filled system.

Source
“min. filling capacity 12,7 l Volume of expansion 7 l”