Product introduction and main features of the chiller
Specp. 2
Technical data according to DIN 12876
p. 1table, 26 rows
Unichiller GL 300w — Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-20...40 °C
temperature set point / display
5,7" colour Touchscreen
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Temperature stability at -10°C
0,2 K
Interface digital
Ethernet, USB (Host u. Device), RS232
Safety classification
I / NFL
Cooling power at 20°C
30 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)
R-744 (A1, H280)
Global Warming Potential (GWP)
1
UN-number
UN 2857
max. delivery
100 l/min
max. delivery pressure
3,5 bar
Pump connection
G1 1/4 male
Cooling water connection
G3/4 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
Volume of expansion
30 l
Overall dimensions WxDxH **
730x804x1738 mm
Power supply (3 Phase)
400V 3~ 50Hz
Pressure equipment category
II
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.
1.0/26
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
12 facts · every claim linked to a page
Installation
The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar at the connection.
Ensures the cooling water supply can meet the minimum flow requirement without exceeding pressure limits.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit is water-cooled and must be connected to a cooling water circuit with G3/4 male connections.
Determines the piping and fittings needed for installation.
Source
“Cooling water connection G3/4 male”
Operation
The chiller is designed for cooling water inlet temperatures between 5°C and 20°C, and lower temperatures require insulation of the cooling water pipes.
Operating outside this range may cause condensation and requires additional insulation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature increases, and the required cooling water flow rate may increase.
Affects sizing of the cooling water supply and expected performance.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water should be used.
Prevents corrosion or damage by ensuring water quality is compatible with these materials.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento
Voltage tolerance is ±5% with a simultaneous frequency tolerance of ±2%, but combinations like -5% voltage and +2% frequency are not allowed.
Ensures the power supply stays within safe limits to avoid damage or malfunction.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Three-phase devices with current consumption less than 63A are supplied with a cable but without a plug.
Requires the installer to provide a suitable plug or hardwired connection.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
The refrigerant is R-744 (CO2) with safety group A1 and GHS hazard H280, and the unit has UN number 2857.
Informs about refrigerant safety classification and transport regulations.
Source
“R-744 (A1, H280)”
Selection
The chiller has a maximum delivery of 100 l/min and a maximum delivery pressure of 3.5 bar.
Helps verify the pump performance matches the application's flow and pressure requirements.
Source
“max. delivery 100 l/min”
Installation
The unit has an IP20 degree of protection, so it must be installed in a dry, indoor environment.
Prevents exposure to water and dust that could damage the equipment.
Source
“Degree of Protection IP20”
Operation
The ambient temperature must be between 5°C and 40°C for proper operation.
Ensures the chiller operates within its designed environmental limits.
Source
“min. ambient temperature 5 °C”
Cumplimiento
The equipment is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011.
Indicates potential electromagnetic interference and may require filtering or separation from sensitive equipment.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”