Contenido

  1. Overview p. 1
    Product introduction and key features of the chiller
  2. Spec pp. 1-2
    Technical data and performance parameters per DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 38 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S459713

Technical data according to DIN 12876

ParameterValue
Operating temperature range-20...100 °C
temperature set point / displaydigital
Internal temperature sensorPt100
Resolution of display0,1 K
Interface digitalUSB (Device), RS232
Interface
Temperature stability at -10°C0,2 K
Alarm messageoptic, acoustic
Safety classificationIII / FL
Heating power at 240V2,1 kW
Heating power at 230V2 kW
Heating power at 220V1,8 kW
Cooling power at 15°C1 kW
Cooling power at 0°C0,8 kW
Cooling power at -10°C0,5 kW
Cooling power at -20°C0,15 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,33 kg
CO2 equivalent0,5 t
Circulation pumpB
at 0,5 bar21 l/min
at 1,0 bar17 l/min
at 1,5 bar11 l/min
at 2,0 bar6 l/min
max. delivery25 l/min
max. delivery pressure2,5 bar
Pump connectionG3/4 male
min. filling capacity3,8 l
expansion tank1,7 l
Overall dimensions WxDxH **350x496x622 mm
Net weight53 kg
sound pressure level +/- 4 dB(A)57 dB(A)
Power supply requirement220-240V 1~/2~ 50/60Hz
max. current14,5 A
Fuse16 A
Pressure equipment categoryArt. 4.3 PED

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is rated for an operating temperature range of -20 to 100 °C, so it must be matched to the process temperature window before selection.

Defines whether the chiller can cover the required process temperatures at all.

Source
“Operating temperature range -20...100 °C”
Performance

Cooling capacity falls from 1 kW at 15 °C to only 0.15 kW at -20 °C, so low-temperature duty needs a much larger unit or longer pull-down.

Prevents undersizing when the application runs near the cold end of the range.

Source
“at 15°C 1 kW at 0°C 0,8 kW at -10°C 0,5 kW at -20°C 0,15 kW”
Installation

The refrigeration machine is air-cooled, so sufficient room ventilation and respect for space requirements are mandatory at the installation site.

Air-cooled condensers reject heat to the room and need adequate airflow clearance.

Source
“Refrigeration machine air-cooled, CFC- and HCFC-free”
Installation

The power supply requirement is 220-240 V single/two-phase 50/60 Hz with a maximum current of 14.5 A and a 16 A fuse, which dictates the site electrical supply.

The electrical connection, cable and protection must be sized for 14.5 A and a 16 A fuse.

Source
“Power supply requirement 220-240V 1~/2~ 50/60Hz max. current 14,5 A Fuse 16 A”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but the combination of -5% voltage with +2% frequency is explicitly not allowed.

Site power quality must be checked; an out-of-tolerance combination can invalidate operation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Single/two-phase devices are delivered with a power cable and country-specific plug, which must be specified at ordering.

The correct plug type must be stated on the order to avoid a non-connectable unit.

Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation

The plastic pump is not resistant to acetone and polyglycols, so these media must be avoided or a more resistant plastic must be ordered.

Prevents chemical attack and pump failure when selecting the heat transfer fluid.

Source
“The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer).”
Operation

For freeze protection water should be mixed with glysantine or ethylene glycol rather than used pure.

Defines the correct heat transfer fluid preparation for sub-zero operation.

Source
“It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Performance

The circulation pump delivers 21 l/min at 0.5 bar but only 6 l/min at 2.0 bar, so high-pressure-drop circuits reduce available flow.

The pump curve must be checked against the pressure drop of the external circuit.

Source
“at 0,5 bar 21 l/min at 1,0 bar 17 l/min at 1,5 bar 11 l/min at 2,0 bar 6 l/min”
Performance

Maximum delivery is 25 l/min at a maximum delivery pressure of 2.5 bar, setting the hydraulic limit of the unit.

Defines the upper bound of flow and pressure available to the application.

Source
“max. delivery 25 l/min max. delivery pressure 2,5 bar”
Installation

The pump connection is G3/4 male, so mating hoses and fittings must be G3/4 female.

Determines the connection hardware needed for the external circuit.

Source
“Pump connection G3/4 male”
Operation

Minimum filling capacity is 3.8 l and the expansion tank is 1.7 l, which sets the minimum fluid charge and expansion allowance.

Needed to charge the system correctly and allow for thermal expansion.

Source
“min. filling capacity 3,8 l expansion tank 1,7 l”
Operation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in wet or hot environments.

Site conditions must stay within these limits to avoid damage or derating.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Output data are valid at 20 °C room temperature and cooling capacity may drop if ambient temperature rises.

Cooling capacity must be derated for warm rooms when sizing the unit.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”