Installation
The chiller is water-cooled and requires a cooling water supply with a minimum differential pressure of 1 bar and a maximum pressure of 6 bar.
Ensures the site cooling water loop can provide adequate flow and pressure without overpressurizing the unit.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water inlet temperature must be between 5°C and 20°C, and pipes must be insulated if below the condensation point.
Defines the permissible cooling water supply conditions and condensation prevention measures at installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling power drops as cooling water temperature rises, and cooling water flow rate may need to increase.
Affects sizing of the cooling water supply and expected performance at the installation site.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and material compatibility to avoid corrosion or damage to the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.
Determines whether the site power supply quality is acceptable for warranty and reliable operation.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at ordering.
Prevents ordering the wrong plug configuration and ensures the unit can be connected on site.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Operation
The unit has an autostart function for power failure, so it may restart automatically after power is restored.
Important for process safety and for assessing whether automatic restart is acceptable in the application.
Source
“Autostart function for power failure”
Safety
The unit has adjustable overtemperature protection according to DIN 12876 and safety classification III / FL.
Confirms built-in safety protection for the temperature control loop and flammable liquid classification.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Selection
The refrigeration machine is water-cooled and uses CFC- and HCFC-free refrigerant R-449A with GWP 1397.
Relevant for environmental compliance and refrigerant handling requirements at the site.
Source
“Refrigeration machine water-cooled, CFC- and HCFC-free”
Installation
The pump connection is G3/4 male and the cooling water connection is G1/2 male.
Needed to select matching hoses and fittings before installation.
Source
“Pump connection G3/4 male”
Operation
The unit is rated IP20 and must be operated within an ambient temperature range of 5°C to 40°C.
Defines the acceptable operating environment and protection against ingress.
Source
“min. ambient temperature 5 °C”
Selection
The optional Pt100 sensor connection is factory-fitted only and can display but not control process temperature.
Must be decided at ordering and clarifies that it is monitoring only, not closed-loop control.
Source
“Option: Pt100 sensor connection #10519 to display (not control) e.g. of the process temperature (only available factory fitted, additional charge)”