Induztr
Resumen técnico Visor de documentos Modelos relacionados

Unichiller P022w OLÉ Technical Data

Contenido

  1. Spec p. 2
    Technical specifications according to DIN 12876 for the chiller
  2. p. 2 table, 4 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.1/24

Engineering Facts

12 facts · every claim linked to a page
Installation

The chiller requires cooling water with a differential pressure of at least 1 bar and no more than 6 bar at the connection.

Ensures the facility cooling water supply can meet the required pressure range for proper operation and to avoid damage.

Source
“min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation

The unit is water-cooled and must be connected to a cooling water supply with G1/2 male connections.

Determines the type of fittings and piping needed for installation.

Source
“Cooling water connection G1/2 male”
Operation

The cooling water inlet temperature must be between 5°C and 20°C; if below the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and ensures reliable operation within specified conditions.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power decreases as cooling water temperature increases, and the required cooling water flow rate may increase.

Helps in sizing the cooling water supply and understanding performance limitations.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety

The plastic pump is not resistant to acetone and polyglycols; for freeze protection, use water mixed with glysantine or ethylene glycol instead.

Prevents damage to the pump and ensures compatibility of the heat transfer fluid.

Source
“Special Case: Acetone and Polyglycol: The plastic pump is not resistant against acetone and polyglycols (depending on the manufacturer). It is recommended that water is mixed with either glysantine or ethylene glycol for freeze protection.”
Installation

The power supply must be 208-230V single or two-phase 60Hz, with voltage tolerance ±5% and frequency tolerance ±2% simultaneously.

Ensures the electrical supply matches the unit's requirements and avoids performance issues.

Source
“Power supply requirement 208-230V 1~/2~ 60Hz”
Installation

The unit requires a minimum fuse of 2x10A and a maximum fuse of 2x16A.

Guides proper circuit protection sizing to prevent electrical hazards.

Source
“min. Fuse 2x10A max. Fuse 2x16A”
Operation

The ambient temperature for operation must be between 5°C and 40°C.

Ensures the installation location maintains suitable environmental conditions.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Cumplimiento

The equipment emits electromagnetic disturbance according to EN55011 Class A, Group 1, which may require specific installation measures.

Informs about potential electromagnetic interference and compliance requirements for the installation environment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

For single/two-phase devices, a power cable with country-specific plug is included; for three-phase devices under 63A, cable is included but no plug.

Clarifies what electrical accessories are supplied and what may need to be sourced locally.

Source
“1. Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering) 2. Three-phase devices with current consumption less than 63A --> with cable, without plug”
Maintenance

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE, and EPDM, so suitable cooling water must be used to avoid corrosion.

Helps select appropriate cooling water quality to prevent corrosion and fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Selection

The refrigeration machine uses R-449A refrigerant with a GWP of 1397, which is CFC- and HCFC-free.

Relevant for environmental regulations and refrigerant handling requirements.

Source
“Refrigerant (ASHRAE, GHS) R-449A (A1, H280) Global Warming Potential (GWP) 1397”