Selection
The chiller is intended only for externally closed applications, so an open bath or vented loop is not permitted.
Defines the hydraulic circuit type allowed and prevents selecting the unit for open-tank duty.
Source
“For externally closed applications.”
Installation
The unit requires a three-phase 400 V 50 Hz supply with 8 A max current and 3x10 A fusing.
Determines site power availability, cable sizing and breaker rating before installation.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the site plug.
Affects electrical installation scope and ordering of the mating plug.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Voltage and frequency tolerances must be met simultaneously; a -5% voltage with +2% frequency is not allowed.
Prevents connecting the chiller to a supply outside the permitted combined tolerance window.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit must be installed where ambient temperature stays between 5 °C and 40 °C.
Room conditions outside this range invalidate performance and may damage the refrigeration circuit.
Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance
Cooling capacity is rated at 20 °C room temperature and drops if ambient temperature rises.
Selection must account for actual plant room temperature, not just the nominal 5 kW.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Selection
The pump delivers 130 l/min maximum and 5.7 bar maximum pressure, with flow falling sharply above 5.0 bar.
Needed to check whether the external circuit pressure drop can be overcome at the required flow.
Source
“max. delivery 130 l/min max. delivery pressure 5.7 bar”
Selection
The expansion tank volume is 48 l and minimum filling capacity is 18 l.
Determines the minimum heat-transfer fluid charge and expansion allowance for the loop.
Source
“min. filling capacity 18 l Volume of expansion 48 l”
Safety
The refrigerant is R-449A with GHS classification H280 and GWP 1397, quantity 1.03 kg.
Affects refrigerant handling, leak response and environmental reporting obligations.
Source
“R-449A (A1, H280)”
Cumplimiento
The unit is classified as pressure equipment category Art. 4.3 PED.
Defines the pressure-equipment compliance route and documentation required at site.
Source
“Pressure equipment category Art. 4.3 PED”
Cumplimiento
The chiller is EMC Class A Group 1, which is not intended for residential environments.
Class A equipment may cause radio interference in domestic areas and needs industrial installation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The enclosure is IP20, so it must be protected from water and dust ingress at the installation site.
Prevents placing the unit in wash-down or dusty areas without additional protection.
Source
“Degree of Protection IP20”
Operation
The sound pressure level is 62 dB(A) ±4 dB(A), which matters for occupied plant rooms.
Used for acoustic planning and personnel exposure assessment.
Source
“sound pressure level +/- 4 dB(A) 62 dB(A)”
Maintenance
The unit has a digital level indicator and an atmospheric open expansion tank that require periodic checking.
Level monitoring is part of routine operation to avoid pump cavitation and air ingress.
Source
“With digital level indicator.”