Unichiller P050w-H Recirculating Chiller Datasheet

Contenido

  1. Overview p. 1
    Product introduction and main features of the chiller
  2. Spec pp. 1-2
    Technical data, performance and operating parameters
  3. Spec p. 2
    Detailed specifications according to DIN 12876 standard
  4. p. 2 table, 20 rows
  5. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C240 l/h
Consumption at water 15°C, flow 0°C216 l/h
Consumption at water 15°C, flow -10°C167 l/h
Consumption at water 15°C, flow -20°C137 l/h
min. cooling water differential pressure0,4 bar
max. cooling water pressure6 bar
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **740x1160x1050 mm
Net weight307 kg
sound pressure level +/- 4 dB(A)62 dB(A)
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)13,5 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S421182

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unichiller P050w-H (Order-No. 3038.0059.01)

ParameterValue
Operating temperature range-20...100 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,2 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationIII / FL
Heating power2 kW
Cooling power at ambient temperature +20°C at 100°C5 kW
Cooling power at ambient temperature +20°C at 20°C5 kW
Cooling power at ambient temperature +20°C at 0°C3,4 kW
Cooling power at ambient temperature +20°C at -10°C2,1 kW
Cooling power at ambient temperature +20°C at -20°C0,8 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity0,85 kg
CO2 equivalent1,2 t
Circulation pumpE2
max. delivery130 l/min
max. delivery pressure5,7 bar
Delivery at 2,4 bar116 l/min
Delivery at 2,5 bar115 l/min
Delivery at 3,0 bar108 l/min
Delivery at 3,5 bar100 l/min
Delivery at 4,0 bar92 l/min
Delivery at 4,5 bar80 l/min
Delivery at 5,0 bar59 l/min
Delivery at 5,5 bar20 l/min
Pump connectionG1 1/4 male

Cooling power values are given at ambient temperature +20°C for the listed temperatures (100°C, 20°C, 0°C, -10°C, -20°C).

Engineering Facts

13 facts · every claim linked to a page
Installation

The chiller requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and protection device rating before installation.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

The unit is water-cooled and needs a cooling water connection of G1/2 male with a minimum differential pressure of 0.4 bar.

Cooling water supply pressure and connection size must be verified during site preparation.

Source
“min. cooling water differential pressure 0,4 bar”
Installation

Cooling water pressure must not exceed 6 bar, so a pressure reducer may be needed on the building water line.

Prevents damage to the cooling water circuit from overpressure.

Source
“max. cooling water pressure 6 bar”
Operation

The unit is designed for cooling water intake between 5 °C and 20 °C, and lower temperatures require insulating the cooling water pipes.

Defines the acceptable cooling water supply range and condensation protection measures.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Affects sizing of the cooling water supply and expected performance at site conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be combined in opposite directions.

Ensures the site power quality is within the allowed envelope for safe operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the need to procure a matching plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The refrigerant R-449A is classified A1 with GHS hazard H280, so refrigerant handling and ventilation precautions apply.

Informs safety measures for refrigerant handling and room ventilation.

Source
“R-449A (A1, H280)”
Cumplimiento

The unit is classified as Class A, Group 1 for electromagnetic disturbance, which has implications for the installation environment.

Class A equipment may require additional EMC measures in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor location protected from water and dust ingress.

Restricts the installation location to dry, protected indoor areas.

Source
“Degree of Protection IP20”
Operation

The ambient temperature must stay between 5 °C and 40 °C for correct operation.

Defines the allowable room temperature range at the installation site.

Source
“min. ambient temperature 5 °C”
Selection

The pump delivers a maximum of 130 l/min at 5.7 bar, but flow falls sharply at higher pressures, which matters for high-pressure-drop applications.

Helps verify the pump can overcome the pressure drop of the external circuit.

Source
“max. delivery pressure 5,7 bar”
Maintenance

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to avoid corrosion and fouling of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”