Unichiller P180 Recirculating Chiller Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unichiller P180 recirculating chiller unit
  2. Spec pp. 1-2
    Performance, temperature, and dimensional data per DIN 12876
  3. p. 2 table, 4 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

Unichiller P180 — Technical data according to DIN 12876

ParameterValue
Order-No.3041.0027.01
Operating temperature range-20...40 °C
temperature set point / display5,7" colour Touchscreen
Internal temperature sensorPt100
Sensor external connectionPt100
Temperature stability at -10°C0,5 K
Interface digitalEthernet, USB (Host u. Device), RS232
Safety classificationI / NFL
Cooling power at ambient temperature +20°C
at 20°C18 kW
at 0°C10 kW
at -10°C6 kW
at -20°C3,5 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-449A (A1, H280)
Global Warming Potential (GWP)1397
Refrigerant quantity1,49 kg
CO2 equivalent2,1 t
Circulation pump:E2
max. delivery140 l/min
max. delivery pressure5,3 bar
Delivery at 2,8 bar126 l/min
Delivery at 3,0 bar124 l/min
Delivery at 3,5 bar115 l/min
Delivery at 4,0 bar103 l/min
Delivery at 4,5 bar83 l/min
Delivery at 5,0 bar43 l/min
Pump connectionG1 1/4 male
min. filling capacity18 l
Volume of expansion48 l
Overall dimensions WxDxH **938x1288x2003 mm
Power supply (3 Phase)460V 3~ 60Hz
Pressure equipment categoryI

** Overall dimensions WxDxH

Engineering Facts

14 facts · every claim linked to a page
Selection

The chiller is rated for an operating temperature range of -20 to 40 °C, so process requirements outside this window cannot be met.

Defines the application envelope and excludes processes needing lower or higher temperatures.

Source
“Operating temperature range -20...40 °C”
Performance

Cooling capacity is only 3.5 kW at -20 °C versus 18 kW at +20 °C, so low-temperature duty drastically reduces available heat removal.

Sizing must use capacity at the actual operating temperature, not the nominal +20 °C figure.

Source
“at -20°C 3,5 kW”
Installation

The unit requires a 460 V 3-phase 60 Hz supply, so site power must match this configuration.

Wrong voltage/frequency will prevent commissioning or damage the compressor.

Source
“Power supply (3 Phase) 460V 3~ 60Hz”
Installation

Three-phase devices below 63 A are supplied with cable but without a plug, so the installer must provide a suitable plug or hardwiring.

Affects electrical installation scope and bill of materials.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

External branch circuit protection per UL 489 is mandatory and must be sized from the electrical schematics.

Required for code compliance and to protect the chiller circuit.

Source
“External branch circuit protection according to UL 489 required.”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are explicitly not allowed.

Prevents installation on marginal power quality that would void performance or damage the unit.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

The unit is IP20 and must be installed where ambient temperature stays between 5 and 40 °C.

Site conditions outside this range affect cooling capacity and reliability.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Performance

Cooling capacity figures are valid at 20 °C room temperature and may drop as ambient rises.

Sizing must account for actual plant room temperature, not just process temperature.

Source
“If the ambient temperature rises, the cooling capacity may drop.”
Installation

The refrigerant is R-449A with GWP 1397 and a 2.1 t CO2 equivalent charge, which matters for leak reporting and environmental compliance.

Triggers F-gas/CO2-equivalent record-keeping and leak-check obligations in many jurisdictions.

Source
“CO2 equivalent 2,1 t”
Installation

The pump connection is G1 1/4 male, so mating hoses and fittings must match this thread.

Determines hose coupling selection and avoids adapters or leaks at commissioning.

Source
“Pump connection G1 1/4 male”
Selection

Minimum filling capacity is 18 l and expansion volume is 48 l, which sets the minimum loop volume and expansion allowance.

Loop volume below 18 l or insufficient expansion headroom will cause control and overflow problems.

Source
“min. filling capacity 18 l Volume of expansion 48 l”
Installation

Overall dimensions are 938x1288x2003 mm, so access and clearance must be planned for this footprint and height.

Needed for room layout, door widths and service access.

Source
“Overall dimensions WxDxH ** 938x1288x2003 mm”
Cumplimiento

The unit is EMC Class A Group 1, which is industrial-grade emission classification and may require separation from sensitive equipment.

Class A equipment is not intended for residential use and may disturb sensitive electronics nearby.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Operation

The controller supports Ethernet, USB host/device and RS232, enabling remote monitoring and data logging.

Determines integration options with BMS, SCADA or lab data systems.

Source
“Interface digital Ethernet, USB (Host u. Device), RS232”