Unimotive 26w Thermostatic Bath/Circulator Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unimotive 26w thermostatic bath/circulator unit
  2. Spec pp. 1-2
    Operating limits, temperature ranges, and performance data per DIN 12876
  3. Spec p. 2
    Physical dimensions, fittings, and connection details
  4. p. 2 table, 25 rows
  5. p. 1 table, 27 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity200 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C1260 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight462 kg
Power supply requirement400V 3~ 50Hz
max. current38 A
Fuse3x40A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S551405

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Order-No.5005.0001.01
ModelUnimotive 26w
Operating temperature range-45...95 °C
Temperature stability at -10°C0,05 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Water/ Glysantin 40/60 at 20°C28 kW
Cooling power with Water/ Glysantin 40/60 at 0°C25 kW
at -20°C14,5 kW
at -30°C6,5 kW
at -40°C2,6 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity3,6 kg
CO2 equivalent7,7 t
Circulation pump: MK pump max. delivery201 l/min
max. delivery pressure5,3 bar

Engineering Facts

13 facts · every claim linked to a page
Selection

The unit is intended for water-ethylene glycol mixtures with corrosion protection down to -45°C, so it suits automotive temperature simulation and material testing rather than general lab use.

Confirms the heat transfer fluid and application envelope before specifying the unit.

Source
“The temperature control systems are constructed for operation with Water-ethylene glycol mixture incl. corrosion protection (for example Glysantin®) down to -45°C.”
Installation

The cooling water supply must be between 5°C and 20°C, and pipes must be insulated if the water temperature can fall below the dew point.

Defines the site cooling water requirements and condensation risk during installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water differential pressure must stay between 1 bar and 4 bar, with a maximum supply pressure of 6 bar.

Ensures the facility cooling loop can provide adequate flow without overpressurizing the unit.

Source
“min. cooling water differential pressure 1 bar”
Installation

The unit requires a 400 V 3~ 50 Hz supply with a maximum current of 38 A and 3x40 A fusing.

Determines electrical supply, cable sizing, and protection device selection.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide the plug and connection.

Affects electrical installation planning and who supplies the plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

Leakage current exceeds 3.5 mA, which may require additional protective measures such as a higher-rated RCD or fixed connection.

Critical for electrical safety design and RCD selection.

Source
“Attention: leakage current > 3,5mA”
Operation

Cooling capacity drops significantly at lower temperatures, from 28 kW at 20°C to 2.6 kW at -40°C, so low-temperature loads must be sized accordingly.

Prevents oversizing expectations for deep-cooling applications.

Source
“at -40°C 2,6 kW”
Operation

The pump delivers 201 l/min at zero head but only 42 l/min at 5 bar, so system pressure drop directly limits available flow.

Needed to match pump performance to the external circuit resistance.

Source
“max. delivery 201 l/min”
Operation

Maximum permissible kinematic viscosity of the heat transfer fluid is 200 mm²/s, limiting use of very viscous fluids at low temperature.

Constrains fluid selection, especially for low-temperature operation.

Source
“max. permissible kin. viscosity 200 mm²/s”
Installation

The unit is IP20 and must be installed where ambient temperature stays between 5°C and 40°C.

Defines environmental limits and protection against dust and water ingress.

Source
“Degree of Protection IP20”
Cumplimiento

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

Important for sites with unstable power quality to avoid invalid operating conditions.

Source
“-5% voltage and +2% frequency -> not allowed!”
Selection

The refrigerant R-452A has a GWP of 2141 and a charge of 3.6 kg, giving a CO2 equivalent of 7.7 t, which matters for environmental reporting and leak management.

Relevant for F-gas compliance and environmental impact assessment.

Source
“Global Warming Potential (GWP) 2141”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and corrosion control to protect the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”