Unimotive 26w-XT Thermostatic Bath/Circulator Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unimotive 26w-XT unit and its main features
  2. Spec pp. 1-2
    Operating data, temperatures, capacities and dimensions per DIN 12876
  3. p. 2 table, 22 rows
  4. p. 1 table, 26 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity200 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C1260 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure12 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/23

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-45...150 °C
Temperature stability at -10°C0,05 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationI / NFL
Heating power24 kW
Cooling power with Water/ Glysantin 40/60 at 20°C28 kW
Cooling power with Water/ Glysantin 40/60 at 0°C25 kW
at -20°C14,5 kW
at -30°C6,5 kW
at -40°C2,6 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity3,6 kg
CO2 equivalent7,7 t
Circulation pump:MK pump
max. delivery201 l/min
max. delivery pressure5,3 bar

Engineering Facts

12 facts · every claim linked to a page
Selection

The unit is designed for water-ethylene glycol mixtures with corrosion protection down to -45°C, so the heat transfer fluid must be selected accordingly.

Determines the permissible heat transfer fluid and prevents unsuitable fluids from being specified.

Source
“The temperature control systems are constructed for operation with Water-ethylene glycol mixture incl. corrosion protection (for example Glysantin®) down to -45°C.”
Installation

The cooling water supply must be between 5°C and 20°C, and pipes must be insulated if condensation can occur.

Defines the required cooling water supply conditions and condensation protection for installation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water differential pressure must stay between 1 bar and 4 bar, with a maximum cooling water pressure of 6 bar.

Sets the cooling water circuit pressure limits for correct and safe connection.

Source
“min. cooling water differential pressure 1 bar”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides cooling water quality and material compatibility to avoid corrosion or damage.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination -5% voltage with +2% frequency is not allowed.

Defines acceptable power supply tolerances and a prohibited voltage/frequency combination.

Source
“-5% voltage and +2% frequency -> not allowed!”
Safety

Leakage current exceeds 3.5 mA, which must be considered for protective earthing and RCD selection.

Affects electrical installation and protective device selection.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices below 63 A are delivered with cable but without plug, so a suitable plug must be provided on site.

Determines what electrical termination is included and what must be supplied during installation.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

The unit requires 400 V 3~ 50 Hz supply and has IP20 protection, so it must be installed in a dry indoor environment.

Defines the electrical supply and environmental protection required for installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Operation

Ambient temperature must remain between 5°C and 40°C for operation.

Sets the permissible ambient conditions for reliable operation.

Source
“min. ambient temperature 5 °C”
Performance

Cooling power drops as temperature decreases, from 28 kW at 20°C to 2.6 kW at -40°C, which must be considered for low-temperature applications.

Allows correct sizing of cooling capacity at the required operating temperature.

Source
“at -40°C 2,6 kW”
Performance

Pump delivery decreases with increasing pressure, from 201 l/min maximum to 42 l/min at 5.0 bar.

Needed to match pump performance to the external circuit pressure drop.

Source
“Delivery at 5,0 bar 42 l/min”
Cumplimiento

The equipment is classified as EN55011 Class A Group 1, which has implications for electromagnetic compatibility in the installation environment.

Determines EMC classification and suitability for the intended electromagnetic environment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”