Unimotive Operation Manual

Contenido

  1. Overview p. 7
    Lists manual chapters and page references
  2. Overview pp. 8, 32-44
    Describes operating principles and device functions
  3. Procedure pp. 9, 45-63
    Instructions for configuring parameters in setup mode
  4. Procedure pp. 10, 66-67
    Interface connections and firmware update procedures
  5. Overview p. 11
    Safety notes and manual conventions
  6. Overview pp. 12-20
    Device overview, technical data and safety information
  7. Procedure pp. 21-31
    Initial startup and commissioning steps
  8. Procedure pp. 64-65
    Standard operating procedures during normal use
  9. Procedure pp. 68-75
    Maintenance tasks and service intervals
  10. Procedure pp. 76-80
    Shutdown and decommissioning procedures
  11. pp. 81-82
    Supplementary technical data and temperature range
  12. p. 9 table, 40 rows
  13. p. 43 table, 27 rows
  14. p. 35 table, 4 rows
  15. p. 40 table, 7 rows
  16. p. 44 table, 3 rows
  17. p. 12 table, 7 rows
  18. p. 16 table, 5 rows
  19. p. 37 table, 8 rows

Содержание (оглавление) — раздел 4 Setup mode

РазделНаименованиеСтр.
3.9.5Restore factory settings42
3.9.5.1Restore to factory settings without overtemperature protection44
3.9.5.2Restore to factory settings including overtemperature protection44
4Setup mode45
4.1Setup mode45
4.1.1Activating / deactivating the emergency stop switch45
4.1.1.1Activating45
4.1.1.2Deactivating45
4.1.2Turning on the temperature control unit45
4.1.3Turning off the temperature control unit46
4.1.4Setting the overtemperature protection46
4.1.4.1General information on the overtemperature protection47
4.1.4.2Setting “OT limit: heating”47
4.1.4.3Setting “OT expansion vessel”47
4.1.4.4Setting “Process Safety”48
4.1.4.5Monitoring via “Display OT values”48
4.1.5Testing overtemperature protection for functionality48
4.1.6Adjusting the Delta T limiter49
4.1.6.1Changing the Delta T limiter49
4.2The temperature control circuit49
4.2.1Select temperature control: Internal or process50
4.2.2Temperature control to internal temperature50
4.2.3Temperature control to process temperature50
4.2.4Delta T limiter51
4.2.5Monitoring the Pt100 temperature sensors51
4.2.6Optimum control parameters for optimum temperature control51
4.2.7Sub-category: “Select auto/expert mode”.51
4.2.8Sub-category: “Configuration auto”52
4.2.8.1Sub-category: “Find parameters”52
4.2.8.2Sub-category: “Control Dynamics”53
4.2.8.3Sub-category: “Fluid Properties”55
4.2.8.4Sub-category: “Display parameters”55
4.2.9Sub-category: “Configuration manual”56
4.2.9.1Sub-category: “Change parameters”56
4.2.9.2Sub-category: “Display parameters”57
4.2.9.3Sub-category: “Control structure”57
4.2.10Sub-category: “Reset parameters”57
4.2.11Sub-category: “Display parameters”58
4.2.12Category “Freeze protection”58
4.2.13Setting the setpoint thresholds58

Это оглавление, а не матрица конфигураций; транспонирование не применялось.

Function description — Unimotive® 43

Description(A)(B)(C)(D)(E)(F)
AcousticsXXX–––
Display functions Warnings; messages; inactive menu items; resolution temperatureXXX–––
Display modesXXX–––
Autostart responseXXX–––
Limitations DeltaT limit; max. heating variable; max. cooling variable; max. permissible current consumption (for 230 V temperature control units)XXX–––
Com.G@te (only if Com.G@te is connected) Analogue interface (configuration input)
Meaning of AIF input; response on cable break; measurement range settingXXX––X
Calibration–––––X
In case of analog error; current / voltage switching; filter constantXXX––X
Configuration output AIF output variable; output range settingXXX––X
Calibration–––––X
RS232/RS485 (hardware RS, Baud rate, device address); ECS standby; POKO alarmXXX–––
Properties (other) Change heat transfer fluid (heat transfer fluid, liter indicator, bypass usage); cooling bath/cooling (CC-E only)XXX–––
Sensor calibration––––––
Automated compressorXXX–––
Enter programX––X––
Pump settings Pump speed setpoint; pump pressure setpoint; control modeXXX–––
Fluid properties Heat transfer fluid; liter indication; pressurisation (only with XT models)XXX–––
Configuration manual parameters
Safety function High limit alarm internal; low limit alarm internal; high limit alarm process; low limit alarm process; hydrostatic correction; warning time limit (CC-E only); min. level (for temperature control units with analogue level sensor); max. level (for temperature controllers with analogue level sensor); optional freeze protectionXXX–––
setpointXXX–––
Setpoint limitation Min. setpoint; max. setpointXXX–––
LanguageXXX–––
Temperature formatXXX–––
Temperature control modeXXX–––
Favorites menuX–––X–
Pilot ONE EthernetXXX–––

“X” = Value is reset, “–” = Value is not reset; (A) = All together; (B) = Device parameters without OT; (C) = Device Parameter; (D) Program Controller; (E) = Menu; (F) = Com.G@te

Overview of the E-grade variants

Temperature control units / E-gradeE-grade BasicE-grade ExclusiveE-grade Professional
Beer force-aging thermostats–XO
Unistat temperature control units––X
UniCAL––X
Other temperature control unitsXOO

X = standard equipment, O = optional, – = not possible

Selecting the heat transfer fluid

Freezing point °CEthylene glycolWater1. Setpoint °C1. Setpoint bar abs.2. Setpoint °C2. Setpoint bar abs.3. Setpoint °C3. Setpoint bar abs.Achievable end temperature in °C
-50 °C60.0040.00951.51202.21504.2150
-45 °C55.0045.00951.51202.41454.2145
-40 °C52.5047.50951.51202.41454.2145
-35 °C47.5052.50951.51202.61404.2140
-30 °C45.0055.00951.51202.61404.2140
-25 °C40.0060.00951.51202.61404.2140
-20 °C35.0065.00951.51202.61354.2135

Mixing ratio in percent by volume: Ethylene glycol / Water. Exemplary representation of the system pressure. Factory settings for models with pressure blanketing.

Table

Description(A)(B)(C)(D)(E)(F)
IP address; subnet mask; remote access
Time unitXXX–––
2nd setpointXXX–––

Representation of textual emphases

EmphasisDescription
AbcStep-by-step explanation of the procedure.
→Reference to information or procedures.
»Abc«Reference to a paragraph in the document.
>Abc< [123]Reference to the wiring diagram in the annex, specifying the designation and search string (number).
>Abc< [ABC]Reference to a drawing in the same paragraph with specification of designation and search string (character).
▪List, first level
‒List, second level

Overview of transportation routes

RefrigerantsRefrigerant filling ex factoryAir freightRoad transportationSea freight
UN 3358≥ 0.1 kg to ≤ 11.99 kgnot possible a)possiblepossible
UN 3358≥ 12 kgpossible b)possible b)
UN 3358 SP A103≤ 99.9 gpossiblepossiblepossible
UN 2857≤ 11.99 kgpossiblepossiblepossible
UN 2857≥ 12 kgnot possible a)possible b)possible b)

a) Filling for transportation: Nitrogen. On-site filling with refrigerants: Only permitted by partners certified by Huber (branch, service partner, dealer or end customer). b) Transportation as hazardous goods.

Function description

FunctionE-grade BasicE-grade ExclusiveE-grade Professional
E-grade evaluation versions available (valid for 30 days)XXX
Storage/loading temperature control programs onto USB stick–XX
Copying settings––X
Service data recorders (flight recorder)XXX
PB commands1XXX
Communication watchdog––X
Process data recording directly to a USB stick: Setpoint, Actual Value Internal and Actual Value Process / Heating Capacity %, Cooling Capacity %, and Pump Pressure / Pump Speed, and VPC Pressure–/–/–X/X/–X/X/X
Calendar start––X

1 See Data Communication Manual. Anything controllable via the graphical user interface of the Pilot ONE can be controlled by PB commands.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit must not be installed or started inside an ATEX zone, so hazardous-area installations are excluded.

Determines whether the unit can be placed in an explosive atmosphere at all.

Source
“Do NOT install or start up the temperature control unit within an ATEX zone.”
Installation

Refrigerant charge and room size must be checked during installation to avoid dangerous refrigerant concentration.

Room volume and charge size govern safe placement of refrigerant-based units.

Source
“Observe the amount of refrigerant (see data sheet/rating plate) and the room size when installing the temperature control unit.”
Installation

A drip tray for condensate/heat transfer fluid must be installed below the unit by the responsible body.

Affects site preparation and containment design before commissioning.

Source
“The responsible body must install a condensation water / heat transfer fluid drip tray below the temperature control unit.”
Installation

Minimum wall clearances depend on air- or water-cooling and on whether the unit runs with or without a bath.

Clearances must be respected to avoid inadequate energy dissipation and overheating.

Source
“Values in the table: without air outlet or connections / with air outlet or connections”
Installation

The unit is a Group 1 Class A device for industrial EMC environments, so it may not be EMC-compatible elsewhere.

Determines EMC suitability of the installation environment and wiring measures.

Source
“It is a Group 1 and Class A device according to the currently applicable EN55011.”
Installation

Only drinking water or filtered river/sea water may be used as cooling water, and sea water is prohibited.

Cooling water quality directly affects warranty and heat exchanger fouling.

Source
“Sea water must not be used for water cooling.”
Operation

Cooling water supply should be shut off during downtimes because flow can occur even when the unit is switched off.

Prevents water damage and unnecessary consumption between operating periods.

Source
“Shut off the cooling water supply to the temperature control unit even for shorter downtimes (e.g. overnight).”
Installation

A backflow preventer on the building side is required to protect the drinking water supply.

Mandatory plumbing measure for water-cooled units connected to mains water.

Source
“A backflow from the cooling water line into the drinking water supply must be prevented on the building side.”
Installation

Outdoor installation requires frost-protected cooling water lines and a minimum cooling water temperature of 5 °C.

Defines winterization and heating requirements for outdoor siting.

Source
“The cooling water temperature must not fall below 5 °C.”
Safety

The unit must always be connected to a socket with protective earth, and hardwiring must be done by a certified electrician.

Electrical safety requirement governing the power connection method.

Source
“Always connect the temperature control unit to a power socket with protective earth (PE).”
Installation

Replacement power cables are length-limited to 3 m for 1–2 phase and 4 m for 3-phase connections.

Constrains cable routing and replacement during installation.

Source
“Connections with 1 or 2 phases have a maximum length of 3 meters; connections with 3 phases a maximum length of 4 meters.”
Performance

Pressure-sensitive applications such as glass reactors require accounting for the unit's maximum flow pressure and possibly an external bypass.

Prevents overpressure damage to glass apparatus when selecting the unit.

Source
“The maximum flow pressure of a temperature control unit must be taken into account in case of pressure-sensitive applications, such as glass reactors.”
Operation

Heat transfer fluid must be brought to room temperature before switching off to avoid overpressure in the system.

Shutdown procedure that protects unit and application from overpressure.

Source
“the heat transfer fluid must always be adjusted to room temperature before switching off”
Selection

Hose inside diameters must at least match the pump connections, with larger diameters for longer lines.

Sizing rule for hoses that affects pressure drop and achievable temperature control.

Source
“The inside diameters of the temperature control hoses must correspond at least to the pump connections.”