Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power distribution, cable and breaker sizing before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 1 bar minimum differential pressure and must not exceed 6 bar.
Ensures adequate cooling water flow and protects the water-cooled condenser from overpressure.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water intake temperature must stay between 5 °C and 20 °C, and pipes must be insulated if condensation occurs.
Defines the permissible cooling water supply conditions and condensation protection measures.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance
Cooling capacity drops significantly at low temperatures with ethanol, from 3.5 kW at 0 °C to only 0.3 kW at -40 °C.
Critical for sizing the unit when low-temperature operation near the lower limit is required.
Source
“at -40°C 0,3 kW”
Performance
The circulation pump delivers 91 l/min maximum and 1.5 bar maximum pressure, with flow falling as pressure rises.
Needed to match the pump curve against the pressure drop of the external circuit.
Source
“max. delivery 91 l/min”
Safety
The unit has a leakage current above 3.5 mA, which affects residual current device selection.
Requires appropriate RCD type and rating to avoid nuisance tripping and ensure protection.
Source
“Attention: leakage current > 3,5mA”
Cumplimiento
Voltage and frequency tolerances are linked: -5% voltage with +2% frequency is not allowed.
Prevents operation outside the EN60034-1 tolerance envelope that could damage the compressor or pump.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so a plug must be provided on site.
Affects site electrical preparation and commissioning schedule.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit is rated for ambient temperatures from 5 °C to 40 °C and IP20 protection, so it must be installed indoors in a dry location.
Defines the allowable room conditions and excludes wet or dusty environments.
Source
“max. ambient temperature 40 °C”
Maintenance
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Guides water quality and treatment to prevent corrosion and fouling of the condenser.
Source
“Please use suitable cooling water.”
Selection
The refrigerant R-452A has a GWP of 2141 and a charge of 0.97 kg, giving 2.1 t CO2 equivalent.
Relevant for refrigerant regulations, leak monitoring and environmental reporting.
Source
“Global Warming Potential (GWP) 2141”
Installation
The pump connection is M30x1.5 male with a Y bore shape per DIN 3863, so matching ball-socket hoses and sleeve nuts are required.
Ensures correct mating of external tempering hoses and avoids leaks.
Source
“Pump connection M30x1,5 male”