Unistat 510w Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat 510w unit
  2. Spec pp. 1-2
    Technical performance data and operating parameters per DIN 12876
  3. p. 2 table, 20 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,25 bar27 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C355 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity4,1 l
Filling capacity expansion tank3,6 l
Overall dimensions WxDxH **460x554x1453 mm
Net weight180 kg
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement400V 3~ 50Hz
max. current13,5 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S383697

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat 510w — Technical data according to DIN 12876

ParameterValue
Order-No.1070.0001.01
Operating temperature range-50...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C5,3 kW
Cooling power with Thermooil at 200°C5,3 kW
Cooling power with Thermooil at 100°C5,3 kW
Cooling power with Thermooil at 20°C5,3 kW
Cooling power with Ethanol at 0°C5,3 kW
Cooling power with Ethanol at -20°C2,8 kW
Cooling power with Ethanol at -40°C0,9 kW
Cooling power with Ethanol at -50°C0,25 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,45 kg
CO2 equivalent3,1 t
UN-numberUN 2857
Circulation pump:MK pump
max. delivery112 l/min
max. delivery pressure1,5 bar
Delivery at 0,25 bar100 l/min
Delivery at 0,5 bar87 l/min
Delivery at 0,75 bar72 l/min
Delivery at 1,0 bar54 l/min

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and breaker rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit must be supplied at 5–20 °C inlet with at least 1 bar differential pressure and no more than 6 bar.

Cooling water conditions directly affect achievable cooling power and must be verified on site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water pipes must be insulated if the water temperature can drop below the dew point to prevent condensation.

Prevents condensation damage and electrical/safety issues in the installation.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality should be used.

Guides selection of water treatment and avoids corrosion or material incompatibility.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

Voltage and frequency tolerances are ±5% and ±2% but must not be combined in the unfavorable direction.

Defines acceptable supply quality for reliable operation and warranty compliance.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The unit has a leakage current above 3.5 mA, which requires appropriate protective earthing and RCD selection.

Affects electrical protection design and residual current device selection.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Prevents delays at commissioning because the customer must supply the plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit is rated for ambient temperatures from 5 °C to 40 °C, which limits where it can be installed.

Room conditions must be checked to avoid derating or malfunction.

Source
“min. ambient temperature 5 °C”
Performance

Cooling power drops significantly at low temperatures, from 5.3 kW at 0 °C with ethanol to 0.25 kW at -50 °C.

Critical for selecting the unit for low-temperature process cooling duties.

Source
“at -50°C 0,25 kW”
Selection

The pump delivers 112 l/min maximum and 54 l/min at 1.0 bar, which must match the external circuit pressure drop.

Ensures the circulation pump can meet the flow and pressure needs of the application.

Source
“max. delivery 112 l/min”
Safety

The refrigerant R-452A is classified A1 with GHS H280, and the charge corresponds to 3.1 t CO2 equivalent.

Relevant for refrigerant handling, safety classification and environmental reporting.

Source
“R-452A (A1, H280)”
Installation

The pump connection is M30x1.5 male and the cooling water connection is G1/2 male, so matching hoses and adapters are needed.

Determines the fittings and hoses required for commissioning.

Source
“Pump connection M30x1,5 male”
Operation

The unit has IP20 protection, so it must be kept in a dry indoor environment free from water ingress.

Restricts installation location and protects against moisture damage.

Source
“Degree of Protection IP20”