Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat 530w unit
  2. Spec pp. 1-2
    Technical data, performance figures and operating limits per DIN 12876
  3. p. 2 table, 21 rows
  4. p. 1 table, 33 rows

Technical data according to DIN 12876

ParameterValue
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C890 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,2 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight443 kg
sound pressure level +/- 4 dB(A)59 dB(A)
Power supply requirement460V 3~ 60Hz
max. current25 A
Fuse3x25A
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S611236

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat 530w (Order-No.: 5002.0102.01)

CharacteristicValue
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 250°C21 kW
Cooling power with Thermooil at 200°C21 kW
Cooling power with Thermooil at 100°C21 kW
Cooling power with Thermooil at 20°C21 kW
Cooling power with Ethanol at 0°C16 kW
Cooling power with Ethanol at -20°C9 kW
Cooling power with Ethanol at -40°C3 kW
Cooling power with Ethanol at -50°C1 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-1270 (A3, H220)
Global Warming Potential (GWP)0
Refrigerant quantity1,35 kg
CO2 equivalent0,0 t
Circulation pump:MK pump
max. delivery196 l/min
max. delivery pressure2,5 bar
Delivery at 0,5 bar175 l/min
Delivery at 1,0 bar152 l/min
Delivery at 1,5 bar123 l/min
Delivery at 2,0 bar86 l/min

Engineering Facts

13 facts · every claim linked to a page
Installation

The unit requires a 460 V 3-phase 60 Hz supply with a 25 A max current and 3x25 A fusing, so the site electrical infrastructure must be sized accordingly.

Determines breaker, cable and panel requirements before ordering.

Source
“Power supply requirement 460V 3~ 60Hz max. current 25 A Fuse 3x25A”
Installation

Cooling water must be supplied at 5–20 °C with 1–4 bar differential pressure and max 6 bar pressure, so the facility water system must meet these limits.

Cooling water conditions directly affect cooling capacity and safe operation.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

If cooling water temperature can fall below the dew point, the cooling water pipes inside the unit must be insulated to prevent condensation.

Prevents condensation damage and electrical risk in low-temperature water supply conditions.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity is rated at 21 kW with Thermooil from 20 to 250 °C but drops with ethanol to 1 kW at -50 °C, so fluid choice strongly affects low-temperature performance.

Needed to verify the unit can handle the required heat load at the target process temperature.

Source
“Cooling power with Ethanol at 0°C 16 kW at -20°C 9 kW at -40°C 3 kW at -50°C 1 kW”
Performance

The magnetic-coupled stainless steel pump delivers 196 l/min max and 2.5 bar max, with 86 l/min still available at 2.0 bar, so external circuit pressure drop must be checked.

Ensures adequate flow through the external application at its actual pressure drop.

Source
“max. delivery 196 l/min max. delivery pressure 2,5 bar Delivery at 2,0 bar 86 l/min”
Safety

The refrigerant is R-1270 (A3, H220) with GWP 0 and 1.35 kg charge, so A3 flammability rules and ventilation/ignition-source precautions apply.

A3 refrigerant classification drives room ventilation, ignition-source and handling requirements.

Source
“Refrigerant (ASHRAE, GHS) R-1270 (A3, H220)”
Safety

Leakage current exceeds 3.5 mA and external branch circuit protection per UL 489 is required, so a suitable upstream breaker must be installed.

Affects electrical protection design and compliance with UL installation requirements.

Source
“Attention: leakage current > 3,5mA External branch circuit protection according to UL 489 required.”
Installation

Three-phase devices under 63 A are supplied with cable but without a plug, so the customer must provide the final connection.

Prevents delays at commissioning due to missing plug or hardwiring provisions.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation

Voltage and frequency tolerances are ±5% and ±2% but must not combine in the same direction, so supply quality must be verified.

Non-compliant supply combinations can damage the compressor and electronics.

Source
“-5% voltage and +2% frequency -> not allowed!”
Operation

The unit is rated for ambient temperatures of 5–40 °C and IP20 protection, so it must be installed indoors in a dry, ventilated location.

Ambient limits affect reliability and cooling performance; IP20 excludes wet or dusty environments.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The unit has a -55 to 250 °C operating range with 0.01 K stability and Pt100 sensors, suitable for precise external closed or open applications.

Confirms the unit meets the required process temperature window and precision.

Source
“Operating temperature range -55...250 °C Temperature stability at -10°C 0,01 K”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Incompatible water chemistry can corrode or foul the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The unit weighs 443 kg and measures 730x804x1738 mm, so floor loading and access route must be planned.

Affects rigging, doorway clearance and structural floor requirements.

Source
“Overall dimensions WxDxH ** 730x804x1738 mm Net weight 443 kg”