Product name and general introduction to the Unistat 650w unit
Specp. 2
Technical data per DIN 12876 including performance and dimensions
p. 1table, 33 rows
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-60...200 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
48 kW
Cooling power with Thermooil at 200°C
65 kW
Cooling power with Thermooil at 100°C
65 kW
Cooling power with Ethanol at 0°C
65 kW
Cooling power with Ethanol at -20°C
56 kW
Cooling power with Ethanol at -40°C
28 kW
Cooling power with Ethanol at -60°C
9,5 kW
Refrigeration machine
water-cooled, natural refrigerant
Refrigerant stage 1.1 (ASHRAE, GHS)
R-1270 (A3, H220)
Refrigerant quantity stage 1.1
2,8 kg
CO2-Äquivalent Stufe 1.1
0,0 t
Refrigerant stage 1.2 (ASHRAE, GHS)
R-1270 (A3, H220)
Refrigerant quantity stage 1.2
2,8 kg
CO2 equivalent stage 1.2
0,0 t
UN-number
UN 3358
Circulation pump: MK pump
max. delivery
346 l/min
max. delivery pressure
4 bar
Delivery at 1,0 bar
325 l/min
Delivery at 1,5 bar
310 l/min
Delivery at 2,0 bar
290 l/min
Delivery at 2,5 bar
270 l/min
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires a 400 V 3-phase 50 Hz supply with a maximum current of 105 A and 3x125 A fusing, so the site electrical infrastructure must be sized accordingly.
Determines electrical supply, cable sizing, and breaker/fuse rating before installation.
Source
“Power supply requirement 400V 3~ 50Hz max. current 105 A Fuse 3x125A”
Installation
Because the current exceeds 63 A, the unit is delivered without cable and without plug, so the customer must provide and terminate the power cable on site.
Affects installation planning, cable procurement, and electrical connection responsibility.
Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation
The cooling water connection is G1 1/4 male and requires a minimum differential pressure of 1 bar and maximum pressure of 6 bar, so the facility water loop must meet these conditions.
Needed to verify facility cooling water supply pressure and fitting compatibility.
Source
“Cooling water connection G1 1/4 male ... min. cooling water differential pressure 1 bar max. cooling water pressure 6 bar”
Installation
The unit is designed for cooling water intake between 5 °C and 20 °C, and if water temperature falls below the condensation point the cooling water pipes must be insulated.
Prevents condensation and ensures valid performance data; affects piping insulation and water supply design.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling capacity is rated at 65 kW at 200 °C and 100 °C with Thermooil, dropping to 9.5 kW at -60 °C with ethanol, so process cooling capacity must be checked against the actual operating temperature.
Critical for selecting the unit against process heat load at the required temperature.
Source
“Cooling power with Thermooil at 200°C 65 kW at 100°C 65 kW Cooling power with Ethanol at 0°C 65 kW at -20°C 56 kW at -40°C 28 kW at -60°C 9,5 kW”
Performance
The circulation pump delivers 346 l/min maximum and 4 bar maximum delivery pressure, with flow falling to 173 l/min at 3.5 bar, so external circuit pressure drop must be checked against the pump curve.
Ensures the pump can overcome the external loop resistance at the required flow.
Source
“max. delivery 346 l/min max. delivery pressure 4 bar ... Delivery at 3,5 bar 173 l/min”
Installation
The pump connection is a DN32 PN10/40 flange per EN1092-1, and pipework must use ball socket per DIN 3863 and sleeve nut per DIN 3870, so mating flanges and fittings must be procured accordingly.
Determines the mechanical interface and fitting standards for connecting the external circuit.
Source
“Pump connection flange DN32 PN10/40 EN1092-1 ... pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Safety
The refrigerant is R-1270 (A3, H220), a flammable natural refrigerant, with 2.8 kg in each of two stages, so the installation area must account for A3 refrigerant safety requirements.
The leakage current exceeds 3.5 mA, so protective earthing and residual current device selection must be suitable for high-leakage equipment.
Affects electrical protection design and RCD type/rating selection.
Source
“Attention: leakage current > 3,5mA”
Installation
The unit has IP20 protection and must be operated between 5 °C and 40 °C ambient, so it must be installed indoors in a dry, temperature-controlled area.
Restricts installation location and environmental conditions.
Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation
The unit weighs 1617 kg and measures 1850x1348x2167 mm, so floor loading, transport route, and space requirements must be verified before delivery.
Needed for rigging, floor loading, and room layout planning.
Source
“Overall dimensions WxDxH ** 1850x1348x2167 mm Net weight 1617 kg”
Operation
The sound pressure level is 73 dB(A) ±4 dB(A), so personnel noise exposure and acoustic treatment may need consideration.
Affects workplace noise assessment and possible hearing protection requirements.
Source
“sound pressure level +/- 4 dB(A) 73 dB(A)”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and combinations such as -5% voltage with +2% frequency are not allowed, so supply quality must be verified.
Prevents invalid supply conditions that could damage the unit or void compliance.
Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% ... -5% voltage and +2% frequency -> not allowed!”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water quality may be used to avoid corrosion.
Guides water quality specification to prevent corrosion and fouling of the cooling circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”