Unistat 905w Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat 905w unit
  2. Spec pp. 1-2
    Technical performance data and operating specifications per DIN 12876
  3. p. 2 table, 21 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 0,4 bar36 l/min
Delivery at 0,6 bar28 l/min
Delivery at 0,8 bar17 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C580 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,2 l
Filling capacity expansion tank5,3 l
Overall dimensions WxDxH **540x654x1738 mm
Net weight388 kg
Power supply requirement208V 3~ 60Hz
max. current34 A
Fuse3x40A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S471454

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat 905w — Technical data according to DIN 12876

ParameterValue
Order-No.1054.0008.01
Operating temperature range-90...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C4,5 kW
Cooling power with Thermooil at 200°C4,5 kW
Cooling power with Thermooil at 100°C4,5 kW
Cooling power with Ethanol at 0°C4,5 kW
Cooling power with Ethanol at -20°C4,5 kW
Cooling power with Ethanol at -40°C4 kW
Cooling power with Ethanol at -60°C2,5 kW
Cooling power with Ethanol at -80°C0,7 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,4 kg
CO2 equivalent3,0 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage0,75 kg
CO2 equivalent 2nd stage11,1 t
Circulation pump:MK pump
max. delivery48 l/min
max. delivery pressure0,9 bar
Delivery at 0,2 bar43 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a water-cooled refrigeration machine, so a cooling water supply with at least 1 bar differential pressure and 580 l/h flow at 15°C must be available at the installation site.

Cooling water capacity and pressure determine whether the site can support the unit without additional pumps or chillers.

Source
“Consumption at water 15°C, flow 20°C 580 l/h”
Installation

The cooling water inlet temperature must be kept between 5°C and 20°C; below the condensation point the cooling water pipes must be insulated.

Operating outside this range risks condensation, reduced cooling power, or equipment damage.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly at low temperatures, from 4.5 kW at -20°C to only 0.7 kW at -80°C, which must be considered when sizing for deep-cold applications.

The available cooling capacity at the target process temperature determines achievable pull-down and heat removal.

Source
“at -80°C 0,7 kW”
Installation

The electrical supply must be 208V 3~ 60Hz with a maximum current of 34 A and 3x40A fusing, and external branch circuit protection according to UL 489 is required.

The site electrical infrastructure and protection devices must be rated for the unit's current and comply with UL requirements.

Source
“External branch circuit protection according to UL 489 required.”
Safety

The unit has a leakage current greater than 3.5 mA, which may require additional protective measures such as a higher-rated RCD.

Standard 30 mA RCDs may nuisance-trip; the electrical design must account for this leakage current.

Source
“Attention: leakage current > 3,5mA”
Installation

Voltage and frequency tolerances are ±5% and ±2% respectively, but they must not be applied in opposite directions simultaneously.

Supply quality must be verified; a combination such as -5% voltage with +2% frequency is not permitted.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices with current consumption less than 63A are delivered with cable but without a plug, so a suitable plug must be provided on site.

The installer must plan for a compatible plug and receptacle, as none is included.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Selection

The pump delivers 48 l/min at zero head but only 17 l/min at 0.8 bar, so external circuit pressure drop must be checked against the pump curve.

Insufficient flow at the required pressure can cause poor temperature control in external closed or open applications.

Source
“Delivery at 0,8 bar 17 l/min”
Selection

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, which limits the choice of fluid at low temperatures.

Using a fluid that is too viscous at the operating temperature can overload the pump and reduce performance.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the cooling water must be compatible with these materials.

Aggressive or scaling cooling water can corrode or foul these materials, reducing cooling performance and causing leaks.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Safety

The refrigerant R-452A has a GWP of 2141 and the second-stage R-23 has a GWP of 14800, giving a total CO2 equivalent of about 14.1 t, which matters for leak reporting and environmental compliance.

High-GWP refrigerants are subject to F-gas regulations, leak checks, and end-of-life recovery obligations.

Source
“Global Warming Potential (GWP) 14800”
Operation

The unit is rated IP20 and must be operated between 5°C and 40°C ambient temperature, so it cannot be installed in wet or dusty environments or outside this ambient range.

Exceeding the ambient limit or exposing the unit to moisture/dust can cause overheating or electrical hazards.

Source
“max. ambient temperature 40 °C”
Selection

The minimum filling capacity is 3.2 l and the expansion tank holds 5.3 l, which sets the minimum external circuit volume and expansion allowance.

External circuits smaller than the minimum filling capacity may not allow stable temperature control or proper pump operation.

Source
“min. filling capacity 3,2 l”
Cumplimiento

The equipment is classified as EN55011 Class A, Group 1, so it is intended for industrial environments and may cause radio interference in residential areas.

Class A equipment may not be suitable for residential use without additional EMC measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”