Unistat 915w Temperature Control System

Contenido

  1. Overview p. 1
    Introduces the Unistat 915w temperature control unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 24 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
max. delivery pressure1,5 bar
Förderleistung bei 0,65 bar90 l/min
Delivery at 1,0 bar74 l/min
Delivery at 1,2 bar59 l/min
Delivery at 1,4 bar31 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C720 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity3,9 l
Filling capacity expansion tank6,5 l
Overall dimensions WxDxH **630x704x1565 mm
Net weight362 kg
sound pressure level +/- 4 dB(A)64 dB(A)
Power supply requirement400V 3~ 50Hz
max. current28,5 A
Fuse3x32A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S402151

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat 915w — Technical data according to DIN 12876

ParameterValue
Order-No.1080.0001.01
Operating temperature range-90...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C6,5 kW
Cooling power with Thermooil at 200°C7,5 kW
Cooling power with Thermooil at 150°C9 kW
Cooling power with Thermooil at 100°C11 kW
Cooling power with Ethanol at 0°C11 kW
Cooling power with Ethanol at -20°C11 kW
Cooling power with Ethanol at -40°C8,2 kW
Cooling power with Ethanol at -60°C4,2 kW
Cooling power with Ethanol at -80°C1,3 kW
Cooling power with Ethanol at -90°C0,6 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-507 (A1, H280)
Global Warming Potential (GWP)3985
Refrigerant quantity2,9 kg
CO2 equivalent11,6 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage1,23 kg
CO2 equivalent 2nd stage18,2 t
Circulation pump:MK pump
max. delivery110 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x32 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable cross-section and protection device needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 28,5 A Fuse 3x32A”
Installation

The cooling water circuit must be supplied at 5–20 °C with at least 1 bar differential pressure and no more than 6 bar.

Cooling water conditions directly affect achievable cooling power and safe operation of the water-cooled refrigeration machine.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation

Cooling water pipes must be insulated if the water temperature can fall below the dew point to prevent condensation.

Prevents condensation damage and electrical hazards in the installation.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling power drops as cooling water temperature rises, and the required cooling water flow rate may increase.

Needed to correctly size the cooling water supply and predict performance at the actual site conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit materials include copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to avoid corrosion or fouling of the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations are not allowed.

Defines acceptable supply quality for warranty and reliable operation of the compressor and electronics.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the need to procure a matching plug.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit is classified as safety class III / FL according to DIN 12876 and has adjustable overtemperature protection.

Relevant for risk assessment and compliance with laboratory safety requirements.

Source
“Safety classification III / FL”
Safety

The refrigerant R-507 has a GWP of 3985 and a CO2 equivalent of 11.6 t, and the second-stage R-23 has a GWP of 14800 and 18.2 t CO2 equivalent.

Important for environmental compliance, leak monitoring and refrigerant reporting obligations.

Source
“Global Warming Potential (GWP) 3985”
Operation

The pump delivers 110 l/min maximum and 31 l/min at 1.4 bar, with a maximum delivery pressure of 1.5 bar.

Needed to match the external circuit pressure drop and ensure adequate flow through the application.

Source
“max. delivery 110 l/min”
Operation

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s.

Limits the choice of heat transfer fluid, especially at low temperatures.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit has a minimum filling capacity of 3.9 l and an expansion tank capacity of 6.5 l.

Determines the minimum heat transfer fluid volume and expansion volume for the system.

Source
“min. filling capacity 3,9 l”
Installation

The unit must be operated within an ambient temperature range of 5 °C to 40 °C.

Affects room requirements and cooling performance at the installation site.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Installation

The unit has IP20 protection, so it must be installed in a dry indoor environment protected from water ingress.

Defines environmental protection requirements for the installation location.

Source
“Degree of Protection IP20”