Unistat GL P550w Technical Data Sheet

Contenido

  1. Overview p. 1
    Introduces the Unistat GL P550w unit and its main features
  2. Spec pp. 1-2
    Performance, temperature range, and technical data per DIN 12876
  3. Spec p. 2
    Physical dimensions, connections, and installation data
  4. p. 2 table, 23 rows
  5. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C3100 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Net weight799 kg
Power supply requirement400V 3~ 50Hz
max. current61,5 A
Fuse3x63A
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S573295

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat GL P550w (Order-No.: 5023.0002.01)

CharacteristicValue
Operating temperature range-50...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 200°C41 kW
Cooling power with Thermooil at 100°C41 kW
Cooling power with Thermooil at 20°C41 kW
Cooling power with Ethanol at 0°C37 kW
Cooling power with Ethanol at -20°C22 kW
Cooling power with Ethanol at -40°C10 kW
Cooling power with Ethanol at -50°C6 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
Refrigerant quantity8,7 kg
CO2 equivalent0,0 t
Circulation pump: MK pump
max. delivery201 l/min
max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min

Значения приведены дословно по тексту страницы; строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' в оригинале напечатаны как группирующие заголовки с подстроками температур.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V 3-phase 50 Hz supply with a 3x63 A fuse, so the electrical installation must be sized for a 61.5 A maximum current.

Determines the site power distribution, cable cross-section and breaker rating needed before delivery.

Source
“Power supply requirement 400V 3~ 50Hz max. current 61,5 A Fuse 3x63A”
Installation

Cooling water must be supplied at 5–20 °C with 1–4 bar differential pressure and no more than 6 bar static pressure, so the site water system must be checked against these limits.

Insufficient or excessive cooling water pressure will prevent rated cooling performance or damage the water circuit.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Installation

The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used to avoid corrosion.

Guides water quality and material compatibility decisions for the plant cooling water supply.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Installation

If cooling water temperature can fall below the condensation point, the cooling water pipes inside the unit must be insulated.

Prevents condensation and corrosion inside the unit when cold cooling water is used.

Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Installation

Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the site plug.

Affects electrical installation planning and the parts needed on site before commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The leakage current exceeds 3.5 mA, so protective earthing and RCD selection must account for this higher leakage.

Impacts the choice of residual current device and earthing arrangement for safe operation.

Source
“Attention: leakage current > 3,5mA”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and certain combinations such as -5% voltage with +2% frequency are not allowed.

The site power quality must be verified against this combined tolerance before installation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Cumplimiento

The equipment is EMC classified as Class A, Group 1 to EN55011, which is intended for industrial environments.

Class A equipment may not be suitable for residential or light-industrial locations without additional measures.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Selection

The pump delivers 201 l/min at zero head but only 42 l/min at 5.0 bar, so the external circuit pressure drop must be matched to the pump curve.

Ensures the selected model can actually achieve the required flow through a high-pressure-drop external loop.

Source
“Delivery at 5,0 bar 42 l/min”
Selection

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s, which limits fluid choice at low temperatures.

Prevents pump overload and poor circulation when selecting a cold heat transfer fluid.

Source
“max. permissible kin. viscosity 100 mm²/s”
Installation

The unit needs at least 8.9 l filling capacity and has a 19.9 l expansion tank, so the external loop volume must be planned accordingly.

Determines the minimum fluid charge and expansion volume for the installed system.

Source
“min. filling capacity 8,9 l Filling capacity expansion tank 19,9 l”
Installation

The unit weighs 799 kg and measures 918x963x1771 mm, so floor loading and access routes must be verified.

Needed for transport planning, doorway clearance and floor load assessment.

Source
“Overall dimensions WxDxH ** 918x963x1771 mm Net weight 799 kg”
Operation

The unit is rated for ambient temperatures from 5 °C to 40 °C, so the installation room must stay within this range.

Operating outside this range may reduce performance or trip the unit.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Selection

The unit uses R-744 (CO2) with GWP 1 and 8.7 kg charge, giving a CO2 equivalent of 0.0 t.

Relevant for refrigerant regulations, safety classification and environmental reporting.

Source
“Refrigerant (ASHRAE, GHS) R-744 (A1, H280) Global Warming Potential (GWP) 1 Refrigerant quantity 8,7 kg CO2 equivalent 0,0 t”