Unistat P510w Temperature Control System Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unistat P510w unit and its main features
  2. Spec pp. 1-2
    Performance, temperature range, and technical data per DIN 12876
  3. p. 2 table, 22 rows
  4. p. 1 table, 29 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,5 bar74 l/min
Delivery at 2,0 bar55 l/min
Delivery at 2,5 bar33 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 0°C355 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity4,1 l
Filling capacity expansion tank3,6 l
Overall dimensions WxDxH **460x554x1453 mm
Net weight182 kg
sound pressure level +/- 4 dB(A)60 dB(A)
Power supply requirement400V 3~ 50Hz
max. current13 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S450083

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat P510w (Order-No.: 1070.0013.01)

ParameterValue
Operating temperature range-50...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C5,3 kW
Cooling power with Thermooil at 200°C5,3 kW
Cooling power with Thermooil at 100°C5,3 kW
Cooling power with Thermooil at 20°C5,3 kW
Cooling power with Ethanol at 0°C5,3 kW
Cooling power with Ethanol at -20°C2,8 kW
Cooling power with Ethanol at -40°C0,9 kW
Cooling power with Ethanol at -50°C0,25 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity1,45 kg
CO2 equivalent3,1 t
Circulation pump: MK pump max. delivery119 l/min
Circulation pump: MK pump max. delivery pressure3 bar
Circulation pump: MK pump Delivery at 0,5 bar106 l/min
Circulation pump: MK pump Delivery at 1,0 bar91 l/min

Таблица приведена как напечатана (характеристика — значение); строки Cooling power сгруппированы по теплоносителю (Thermooil / Ethanol) с указанием температур.

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable and breaker rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit must be supplied at 1 bar minimum differential pressure and must not exceed 6 bar.

Ensures the water-cooled condenser receives adequate flow without overpressurizing the circuit.

Source
“min. cooling water differential pressure 1 bar”
Installation

Cooling water intake must stay between 5 °C and 20 °C, and pipes must be insulated if water can drop below the dew point.

Prevents condensation, corrosion and loss of cooling performance at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity falls sharply at low temperatures, from 5.3 kW at 0 °C to only 0.25 kW at -50 °C with ethanol.

Critical for sizing the unit when the process requires deep cooling near the lower limit.

Source
“at -50°C 0,25 kW”
Selection

The pump delivers 119 l/min at zero head but only 33 l/min at 2.5 bar, so high-pressure-drop loops need careful pump curve review.

Matches pump capability to the pressure drop of the external application circuit.

Source
“max. delivery 119 l/min”
Safety

Leakage current exceeds 3.5 mA, which affects residual-current device selection and earthing requirements.

Needed to choose a suitable RCD type and ensure protective earthing at the installation.

Source
“Attention: leakage current > 3,5mA”
Cumplimiento

Voltage and frequency tolerances are ±5 % and ±2 % but must not combine in the same direction, e.g. -5 % voltage with +2 % frequency is not allowed.

Defines acceptable supply quality and protects the compressor and electronics from out-of-tolerance operation.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a plug must be sourced and fitted on site.

Affects site electrical preparation and commissioning schedule.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The refrigerant R-452A has a GWP of 2141 and a CO2 equivalent of 3.1 t, which matters for leak monitoring and F-gas obligations.

Relevant for environmental compliance and refrigerant leak-check intervals.

Source
“Global Warming Potential (GWP) 2141”
Operation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in a wet or hot environment.

Constrains the installation location and enclosure requirements.

Source
“Degree of Protection IP20”
Selection

The maximum permissible kinematic viscosity of the heat transfer fluid is 50 mm²/s, limiting fluid choice at low temperatures.

Prevents pump overload and ensures adequate circulation with the selected thermal fluid.

Source
“max. permissible kin. viscosity 50 mm²/s”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, brass, PA, PPE, PTFE and EPDM, so only compatible water quality may be used.

Guides water treatment and avoids corrosion or scaling damage to the condenser.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”