Unistat P525 Technical Data Sheet

Contenido

  1. Overview p. 1
    Introduces the Unistat P525 unit and its main features
  2. Spec p. 2
    Lists technical data according to DIN 12876 standard
  3. p. 2 table, 22 rows
  4. p. 1 table, 28 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 1,0 bar66 l/min
Delivery at 1,5 bar57 l/min
Delivery at 2,0 bar46 l/min
Delivery at 2,5 bar32 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity5,1 l
Filling capacity expansion tank18 l
Overall dimensions Temperature Control Unit WxDxH1290x795x1377 mm
Dimensions of External Transformer WxDxH253x560x602 mm
Weight External Transformer135 kg
sound pressure level +/- 4 dB(A)56 dB(A)
Power supply factory configured (3 Phase)208V 3~ 60Hz
Input Voltage External Transformer208V 3~ 60Hz
Output Voltage External Transformer440V 3~ 60Hz
Max. current temp. control unit with external transformer45 A
Fuse input voltage external transformer3x50A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.1/24

Delivery at 1,0 bar; Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.

Unistat P525 Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power6 kW
Cooling power with Thermooil at 250°C10 kW
Cooling power with Thermooil at 200°C10 kW
Cooling power with Thermooil at 100°C10 kW
Cooling power with Thermooil at 20°C10 kW
Cooling power with Ethanol at 0°C7 kW
Cooling power with Ethanol at -20°C4,2 kW
Cooling power with Ethanol at -40°C1,5 kW
Cooling power with Ethanol at -50°C0,65 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
UN-numberUN 2857
Circulation pumpMK pump
max. delivery82 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar75 l/min

Engineering Facts

14 facts · every claim linked to a page
Selection

The unit is a refrigerated heating circulator with a magnetic-coupled stainless steel pump and is suitable for both externally closed and externally open applications.

Determines whether the circulator can serve open bath or external loop configurations without additional hardware.

Source
“Refrigerated Heating Circulator with air-cooled cooling machine and optical level indicator. Magnetic coupled circulation pump made of stainless steel.”
Selection

The P models use a high-discharge-pressure pump intended for applications with high pressure drops such as flow-through chemistry or semiconductor industry.

Confirms the pump head is adequate for restrictive external circuits with significant pressure loss.

Source
“Unistat „P“ Models: Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation

The temperature control unit requires an external transformer supplied as standard to provide the required voltage.

The transformer must be accommodated and wired on site, affecting footprint and electrical installation.

Source
“the required voltage is provided using a external transformer (supplied as standard)”
Installation

The external transformer is a separate heavy component with its own dimensions and weight that must be planned for.

Space and floor loading must be reserved for the transformer in addition to the main unit.

Source
“Dimensions of External Transformer WxDxH 253x560x602 mm”
Installation

The external transformer weighs 135 kg and must be handled and positioned accordingly.

Requires suitable lifting and floor support during installation.

Source
“Weight External Transformer 135 kg”
Installation

The input voltage of the external transformer is 208V 3~ 60Hz and the output is 440V 3~ 60Hz.

Site power must match the 208V three-phase input; the 440V output feeds the temperature control unit.

Source
“Output Voltage External Transformer 440V 3~ 60Hz”
Installation

The input side of the external transformer requires 3x50A fuses.

Branch circuit protection must be sized to 50A per phase on the transformer input.

Source
“Fuse input voltage external transformer 3x50A”
Installation

The maximum current of the temperature control unit with external transformer is 45 A.

Cable and protection ratings on the transformer output must accommodate 45 A.

Source
“Max. current temp. control unit with external transformer 45 A”
Installation

External branch circuit protection according to UL 489 is required.

The installer must provide UL 489 listed branch protection; it is not included.

Source
“External branch circuit protection according to UL 489 required.”
Safety

Leakage current exceeds 3.5 mA, which affects protective earth and RCD selection.

Standard 30 mA RCDs may nuisance-trip; earth connection and protection must be designed for high leakage.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices drawing less than 63A are supplied with cable but without a plug.

A suitable plug or direct connection must be procured and installed on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

Cooling capacity is rated at 20°C room temperature and may drop if ambient temperature rises.

Performance margins must account for actual ambient conditions in the installation room.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation

The permissible ambient temperature range is 5°C to 40°C.

The room must stay within this range for reliable operation and rated performance.

Source
“max. ambient temperature 40 °C”
Installation

The pump connection is M30x1.5 male with bore shape Y (60°) per DIN 3863 and mating ball socket and sleeve nut per DIN 3863/3870.

External hoses and fittings must match this thread and DIN geometry to avoid leaks.

Source
“Pump connection M30x1,5 male”