Unistat P527w Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat P527w unit
  2. Spec p. 2
    Technical data according to DIN 12876 standard
  3. p. 2 table, 25 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
Consumption at water 15°C, flow 0°C880 l/h
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,2 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **730x804x1738 mm
Net weight444 kg
sound pressure level +/- 4 dB(A)57 dB(A)
Power supply requirement400V 3~ 50Hz
max. current31,62 A
Fuse3x32A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S602229

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-55...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power12 kW
Cooling power with Thermooil at 250°C12 kW
Cooling power with Thermooil at 200°C12 kW
Cooling power with Thermooil at 100°C12 kW
Cooling power with Thermooil at 20°C12 kW
Cooling power with Ethanol at 0°C12 kW
Cooling power with Ethanol at -20°C6 kW
Cooling power with Ethanol at -40°C2 kW
Cooling power with Ethanol at -50°C1 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2,9 kg
CO2 equivalent6,2 t
Circulation pump:MK pump
max. delivery201 l/min
max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min

Order-No.: 5001.0002.01; Unistat P527w

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x32 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable cross-section and protection device needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz max. current 31,62 A Fuse 3x32A”
Installation

The cooling water circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar, with 880 l/h consumption at 15 °C.

Sizes the cooling water supply, return lines and pressure regulation needed for reliable operation.

Source
“Consumption at water 15°C, flow 0°C 880 l/h min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Operation

Cooling water inlet must stay between 5 °C and 20 °C, and pipes must be insulated if the water can drop below the dew point.

Prevents condensation, corrosion and loss of cooling performance in the water circuit.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C. When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance

Cooling capacity drops sharply at low temperatures with ethanol, from 12 kW at 0 °C to 1 kW at -50 °C.

Allows the engineer to check whether the required heat load can be removed at the lowest process temperature.

Source
“at 0°C 12 kW at -20°C 6 kW at -40°C 2 kW at -50°C 1 kW”
Selection

The pump delivers 201 l/min maximum and 42 l/min at 5.0 bar, so high-pressure-drop loops must be checked against the pump curve.

Confirms whether the circulation pump can overcome the pressure drop of the external circuit.

Source
“max. delivery 201 l/min max. delivery pressure 5,3 bar Delivery at 5,0 bar 42 l/min”
Safety

The refrigerant R-452A is classified A1 with GHS H280, and the charge is 2.9 kg corresponding to 6.2 t CO2 equivalent.

Affects refrigerant handling, leak detection and F-gas reporting obligations.

Source
“R-452A (A1, H280) Global Warming Potential (GWP) 2141 Refrigerant quantity 2,9 kg CO2 equivalent 6,2 t”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and opposite-sign deviations are not allowed.

Defines the acceptable power quality at the installation site and protects the compressor and electronics.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2% Example -5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the plug and connection.

Clarifies the scope of supply and what the electrical contractor must provide on site.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient temperature.

Restricts installation location and prevents operation in wet or hot environments.

Source
“Degree of Protection IP20 min. ambient temperature 5 °C max. ambient temperature 40 °C”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible water quality may be used.

Guides water treatment and avoids corrosion or fouling of the heat exchanger.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM. Please use suitable cooling water.”
Selection

The maximum permissible kinematic viscosity of the heat transfer fluid is 100 mm²/s.

Limits the choice of heat transfer fluid, especially at low temperatures where viscosity rises.

Source
“max. permissible kin. viscosity 100 mm²/s”
Installation

The unit needs at least 8.2 l filling capacity and has a 19.9 l expansion tank.

Determines the minimum fluid charge and the expansion volume available for the external loop.

Source
“min. filling capacity 8,2 l Filling capacity expansion tank 19,9 l”