Product introduction and general description of Unistat P815
Specp. 2
Technical data according to DIN 12876 standard
p. 2table, 22 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
max. delivery
67 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
61 l/min
Delivery at 1,0 bar
55 l/min
Delivery at 1,5 bar
48 l/min
Delivery at 2,0 bar
39 l/min
Delivery at 2,5 bar
28 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
min. filling capacity
3,8 l
Filling capacity expansion tank
3,3 l
Overall dimensions WxDxH **
460x604x1465 mm
Net weight
241 kg
sound pressure level +/- 4 dB(A)
64 dB(A)
Power supply factory configured (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
12 A
Fuse (3 phase)
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S290580
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P815 — Technical data according to DIN 12876
Parameter
Value
Order-No.
1053.0009.01
Operating temperature range
-85...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
2 kW
Cooling power with Thermooil at 250°C
1,3 kW
Cooling power with Thermooil at 200°C
1,3 kW
Cooling power with Thermooil at 100°C
1,3 kW
Cooling power with Ethanol at 0°C
1,5 kW
Cooling power with Ethanol at -20°C
1,5 kW
Cooling power with Ethanol at -40°C
1,4 kW
Cooling power with Ethanol at -60°C
1,2 kW
Cooling power with Ethanol at -80°C
0,2 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
1,8 kg
CO2-Äquvivalent
3,9 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
0,35 kg
CO2-Äquvivalent 2. Stufe
5,2 t
Circulation pump:
MK pump
Значения приведены дословно по тексту страницы; строки Cooling power сгруппированы по теплоносителю (Thermooil / Ethanol) и температуре.
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit is a refrigerated heating circulator suitable for both externally closed and externally open applications, which determines whether an isolation sleeve may be needed.
Open external applications require additional accessories such as an isolation sleeve, affecting selection and cost.
Source
“As well as for externally closed and also externally open applications.”
Selection
The 'P' model pump is designed for high pressure drops, making it suitable for flow-through chemistry or semiconductor applications.
Confirms the pump variant matches high-pressure-drop loops; a standard pump may be insufficient.
Source
“Circulating pumps with a high discharge pressure for applications with high pressure drops, e.g. in the Flow-Through chemistry or in the Semicon industry.”
Installation
The unit requires a 3-phase 400 V supply with 12 A max current and 3x16 A fusing, so the electrical installation must be sized accordingly.
Determines breaker, cable and supply requirements before installation.
Source
“max. current (3 Phase) 12 A”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the plug.
Affects site preparation and commissioning; a plug must be sourced and installed.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
Pump connections are M30x1.5 male with bore shape Y (60°) per DIN 3863, requiring matching ball-socket hoses and sleeve nuts per DIN 3870.
Ensures correct hose/adaptor selection and leak-free connection at commissioning.
Source
“Pump connection M30x1,5 male”
Operation
Cooling capacity is rated at 20 °C room temperature and may drop if ambient temperature rises, so performance must be derated in warm rooms.
Prevents undersizing cooling capacity in installations with ambient above 20 °C.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Operation
The permissible ambient temperature range is 5 to 40 °C, limiting where the unit can be installed and operated.
Site conditions outside this range void performance assumptions and may affect reliability.
Source
“max. ambient temperature 40 °C”
Safety
Leakage current exceeds 3.5 mA, which requires appropriate protective earthing and RCD selection.
Affects electrical safety design; standard 30 mA RCDs may nuisance-trip or be unsuitable.
Source
“Attention: leakage current > 3,5mA”
Safety
The refrigerant R-452A is classified A1 with GHS H280 (pressurised gas), and the second stage R-23 is also A1/H280, so handling and ventilation precautions apply.
Informs refrigerant safety classification and site ventilation/emergency planning.
Source
“R-452A (A1, H280)”
Cumplimiento
The equipment is EMC Class A Group 1, so it is intended for industrial environments and may cause interference in residential areas.
Determines EMC environment suitability and possible need for filtering or separation.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Cumplimiento
Voltage and frequency tolerances are ±5% and ±2% respectively, but certain combinations (e.g. -5% voltage with +2% frequency) are not allowed.
Prevents operation on marginal supplies that could damage the compressor or electronics.
Source
“-5% voltage and +2% frequency -> not allowed!”
Maintenance
The refrigeration machine is air-cooled and CFC/HCFC-free, so adequate air flow and condenser cleaning are required.
Air-cooled condensers need clearance and periodic cleaning; poor airflow reduces capacity.
Source
“air-cooled, CFC- and HCFC-free”
Performance
Cooling capacity drops sharply at low temperature with ethanol, from 1.2 kW at -60 °C to 0.2 kW at -80 °C, so low-temperature loads must be small.
Critical for sizing low-temperature processes near the -85 °C limit.
Source
“at -80°C 0,2 kW”
Selection
The pump delivers 67 l/min max and 3 bar max pressure, with flow falling to 28 l/min at 2.5 bar, so system pressure drop must be checked against the pump curve.
Ensures the circulation loop can achieve required flow at the actual pressure drop.