Introduces the Unistat P815w unit and its main features
Specp. 2
Lists technical data according to DIN 12876 standard
p. 2table, 25 rows
p. 1table, 31 rows
Technical data according to DIN 12876
Parameter
Value
max. delivery
67 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
61 l/min
Delivery at 1,0 bar
55 l/min
Delivery at 1,5 bar
48 l/min
Delivery at 2,0 bar
39 l/min
Delivery at 2,5 bar
28 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 0°C
130 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,2 l
Filling capacity expansion tank
3,4 l
Overall dimensions WxDxH **
460x604x1465 mm
Net weight
231 kg
Power supply requirement
400V 3~ 50Hz
max. current
12,5 A
Fuse
3x16A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S535218
** Please respect space requirements. See operating conditions at www.huber-online.com
Unistat P815w — Technical data according to DIN 12876
Characteristic
Value
Order-No.
1053.0010.01
Operating temperature range
-85...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
2 kW
Cooling power with Thermooil at 250°C
1,5 kW
Cooling power with Thermooil at 200°C
1,5 kW
Cooling power with Thermooil at 100°C
1,5 kW
Cooling power with Ethanol at 0°C
1,5 kW
Cooling power with Ethanol at -20°C
1,5 kW
Cooling power with Ethanol at -40°C
1,4 kW
Cooling power with Ethanol at -60°C
1,2 kW
Cooling power with Ethanol at -80°C
0,2 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
0,95 kg
CO2 equivalent
2,0 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
0,39 kg
CO2 equivalent 2nd stage
5,8 t
Circulation pump:
MK pump
Engineering Facts
12 facts · every claim linked to a page
Installation
The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.
Determines the site power supply and protection device rating before installation.
Source
“Power supply requirement 400V 3~ 50Hz”
Installation
The cooling water circuit must be supplied at 1–6 bar differential pressure and 5–20 °C inlet temperature, so the facility water system must meet these limits.
Insufficient or excessive cooling water pressure will prevent proper operation and may damage the unit.
Source
“min. cooling water differential pressure 1 bar”
Installation
Cooling water pipes must be insulated when the water temperature can fall below the dew point to avoid condensation.
Prevents condensation and corrosion on the cooling water circuit.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Performance
Cooling capacity drops significantly at low temperatures, from 1.5 kW at -20 °C to only 0.2 kW at -80 °C, which must be considered when sizing the load.
The available cooling power at the lowest temperatures limits the permissible heat load.
Source
“at -80°C 0,2 kW”
Selection
The pump delivers 67 l/min maximum and 3 bar maximum pressure, with 28 l/min still available at 2.5 bar, so external circuit pressure drop must be checked.
Ensures the circulation pump can overcome the pressure drop of the external application.
Source
“max. delivery pressure 3 bar”
Safety
The refrigerant R-452A has a GWP of 2141 and the second-stage R-23 has a GWP of 14800, giving a total CO2 equivalent of about 7.8 t that must be reported and managed.
F-gas regulations require knowledge of refrigerant charge and CO2 equivalent for operation and maintenance.
Source
“Global Warming Potential (GWP) 14800”
Safety
The leakage current exceeds 3.5 mA, so a suitable residual current device and protective earthing must be provided.
Affects the choice of RCD type and earthing arrangement for safe operation.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices below 63 A are delivered with a cable but without a plug, so the customer must provide and install the plug.
The site electrician must supply a suitable plug and connect the unit.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The unit is designed for cooling water intake between 5 °C and 20 °C, and higher water temperatures reduce cooling power and increase water flow.
Facility cooling water temperature directly affects achievable cooling performance.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The ambient temperature must stay between 5 °C and 40 °C for correct operation.
Site conditions outside this range may cause overheating or control problems.
Source
“max. ambient temperature 40 °C”
Cumplimiento
Voltage tolerance is ±5 % only with a simultaneous frequency tolerance of ±2 %, and certain combinations such as -5 % voltage with +2 % frequency are not allowed.
Ensures the supply quality at the installation site meets the unit's requirements.
Source
“-5% voltage and +2% frequency -> not allowed!”
Maintenance
The cooling water circuit contains copper, stainless steel 1.4401, brass, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.
Prevents corrosion and fouling of the cooling water circuit.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”