Unistat P825w Technical Data

Contenido

  1. Overview p. 1
    Introduces the Unistat P825w unit and its main features
  2. Spec p. 2
    Lists technical data according to DIN 12876 standard
  3. p. 2 table, 26 rows
  4. p. 1 table, 30 rows

Technical data according to DIN 12876

ParameterValue
Circulation pumpMK pump
max. delivery67 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar61 l/min
Delivery at 1,0 bar55 l/min
Delivery at 1,5 bar48 l/min
Delivery at 2,0 bar39 l/min
Delivery at 2,5 bar28 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water connectionG1/2 male
Consumption at water 15°C, flow 20°C180 l/h
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity2,4 l
Filling capacity expansion tank3,3 l
Overall dimensions WxDxH **460x604x1465 mm
Net weight237 kg
Power supply requirement400V 3~ 50Hz
max. current15 A
Fuse3x16A
Pressure equipment categoryI
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S513166

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876

ParameterValue
Operating temperature range-85...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power3 kW
Cooling power with Thermooil at 250°C2,3 kW
Cooling power with Thermooil at 200°C2,3 kW
Cooling power with Thermooil at 100°C2,3 kW
Cooling power with Thermooil at 20°C2,2 kW
Cooling power with Ethanol at 0°C2,2 kW
Cooling power with Ethanol at -20°C2,1 kW
Cooling power with Ethanol at -40°C2 kW
Cooling power with Ethanol at -60°C1,3 kW
Cooling power with Ethanol at -80°C0,3 kW
Refrigeration machinewater-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity0,95 kg
CO2 equivalent2,0 t
Refrigerant 2nd stage (ASHRAE, GHS)R-23 (A1, H280)
Global Warming Potential (GWP)14800
Refrigerant quantity 2nd stage0,39 kg
CO2 equivalent 2nd stage5,8 t

Order-No.: 1079.0012.01; Unistat P825w

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply and breaker/fuse rating needed before installation.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

Cooling water must be supplied at 5–20 °C inlet with at least 1 bar differential pressure and no more than 6 bar, and pipes must be insulated if below dew point.

Defines the cooling water utility requirements and condensation risk at the installation site.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling power drops significantly at low temperatures, from 2.2 kW at 0 °C to only 0.3 kW at -80 °C, which must be considered for load sizing.

Affects selection because available cooling capacity at the target low temperature may be insufficient for the process load.

Source
“at -80°C 0,3 kW”
Performance

The pump delivers 67 l/min max and 3 bar max pressure, with 28 l/min still available at 2.5 bar, so high-pressure-drop loops can be served.

Needed to verify the circulation pump can overcome the pressure drop of the external application.

Source
“max. delivery pressure 3 bar”
Selection

The unit uses R-452A (GWP 2141, 0.95 kg) and R-23 (GWP 14800, 0.39 kg) refrigerants, giving a total CO2 equivalent of about 7.8 t.

Relevant for refrigerant regulations, leak monitoring obligations and environmental compliance when selecting the unit.

Source
“Refrigerant 2nd stage (ASHRAE, GHS) R-23 (A1, H280)”
Operation

The unit is rated IP20 and must be operated between 5 °C and 40 °C ambient, so it cannot be placed in wet or hot environments.

Constrains the installation location and room ventilation/cooling requirements.

Source
“max. ambient temperature 40 °C”
Installation

Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and certain combinations are explicitly not allowed.

Must be checked against the site power quality to avoid damage or malfunction.

Source
“-5% voltage and +2% frequency -> not allowed!”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit has safety classification III / FL and adjustable overtemperature protection according to DIN 12876.

Important for safety assessment and compliance with laboratory temperature-control standards.

Source
“With adjustable overtemperature protection according to DIN 12876.”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality and treatment to prevent corrosion and fouling in the cooling circuit.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation

The pump connection is M30x1.5 male and cooling water connection is G1/2 male, so matching hoses and fittings are required.

Needed to order correct hoses, adapters and fittings before installation.

Source
“Cooling water connection G1/2 male”
Performance

Cooling water consumption is 180 l/h at 15 °C water and 20 °C flow, and higher water temperatures reduce cooling power and increase flow demand.

Determines cooling water supply capacity and utility sizing at the installation site.

Source
“Consumption at water 15°C, flow 20°C 180 l/h”