Product introduction and main features of Unistat P905w
Specpp. 2-3
Technical data according to DIN 12876 standard
p. 2table, 25 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
max. delivery
65 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
60 l/min
Delivery at 1,0 bar
54 l/min
Delivery at 1,5 bar
47 l/min
Delivery at 2,0 bar
39 l/min
Delivery at 2,5 bar
29 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G1/2 male
Consumption at water 15°C, flow 20°C
580 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,2 l
Filling capacity expansion tank
5,3 l
Overall dimensions WxDxH **
540x654x1500 mm
Net weight
261 kg
Power supply requirement
460V 3~ 60Hz
max. current
23 A
Fuse
3x25A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S454887
** Overall dimensions WxDxH
Technical data according to DIN 12876 — Unistat P905w (Order-No.: 1054.0013.01)
Characteristic
Value
Operating temperature range
-90...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
4,2 kW
Cooling power with Thermooil at 200°C
4,2 kW
Cooling power with Thermooil at 100°C
4,2 kW
Cooling power with Ethanol at 0°C
4,4 kW
Cooling power with Ethanol at -20°C
4,4 kW
Cooling power with Ethanol at -40°C
4 kW
Cooling power with Ethanol at -60°C
2,3 kW
Cooling power with Ethanol at -80°C
0,5 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-452A (A1, H280)
Global Warming Potential (GWP)
2141
Refrigerant quantity
1,4 kg
CO2 equivalent
3,0 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
0,75 kg
CO2 equivalent 2nd stage
11,1 t
Circulation pump:
MK pump
Таблица приведена как напечатана (характеристика — значение); строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' объединяют подзаголовки температур с соответствующими значениями мощности.
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit covers a very wide temperature range from -90 to 250 °C, so it can serve both deep-freeze and high-temperature processes with one circulator.
Defines whether the unit can meet the process temperature envelope before any other sizing work.
Source
“Operating temperature range -90...250 °C”
Performance
Cooling capacity collapses to only 0.5 kW at -80 °C, so deep-cold loads must be sized against the low-temperature end, not the nominal 4.4 kW.
Prevents oversizing the process load based on mid-range cooling capacity and underestimating pull-down time.
Source
“at -80°C 0,5 kW”
Installation
The refrigeration machine is water-cooled and needs 580 l/h at 15 °C with at least 1 bar differential pressure, so facility cooling water must be verified before installation.
Cooling water flow and pressure are hard installation constraints that can block commissioning.
Source
“Consumption at water 15°C, flow 20°C 580 l/h”
Installation
Cooling water inlet must stay between 5 and 20 °C, and pipes must be insulated if condensation can occur, so the water supply temperature must be controlled.
Out-of-range cooling water temperature degrades capacity and creates condensation risk on the unit.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
The unit requires a 460 V 3~ 60 Hz supply with 23 A max current and 3x25 A fusing, so the electrical feed and branch protection must be planned accordingly.
Electrical supply mismatch is a common cause of failed site acceptance.
Source
“Power supply requirement 460V 3~ 60Hz”
Safety
Leakage current exceeds 3.5 mA and UL 489 branch circuit protection is required, so residual-current protection and breaker selection must be coordinated.
Affects RCD selection and compliance with UL installation requirements.
Source
“Attention: leakage current > 3,5mA”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide the site plug and connection.
Avoids assuming a ready-to-plug unit and missing the mating plug at installation.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Operation
The pump delivers 65 l/min at zero head but only 29 l/min at 2.5 bar, so external pressure drop must be checked against the pump curve.
High-pressure-drop loops may starve flow if only the maximum delivery figure is used for sizing.
Source
“max. delivery pressure 3 bar”
Selection
The P-model pump is intended for high discharge pressure applications with high pressure drops such as flow-through chemistry or semiconductor industry.
Confirms the model family is appropriate only when the external circuit actually has high pressure drop.
Source
“Circulating pumps with a high discharge pressure for applications with high pressure drops”
Operation
Maximum permissible kinematic viscosity is 50 mm²/s, so the heat transfer fluid must be checked at the lowest operating temperature.
Too-viscous fluid at low temperature overloads the pump and reduces heat transfer.
Source
“max. permissible kin. viscosity 50 mm²/s”
Maintenance
The refrigerant charges are R-452A (1.4 kg, GWP 2141) and R-23 (0.75 kg, GWP 14800), giving a combined CO2 equivalent of about 14.1 t, which matters for leak reporting and service.
Determines refrigerant leak-check obligations and environmental compliance during service.
Source
“CO2 equivalent 2nd stage 11,1 t”
Installation
The unit is IP20 and must be installed at ambient temperatures between 5 and 40 °C, so it cannot be placed in wet or hot environments.
Prevents installation in washdown areas or locations outside the allowed ambient range.
Source
“Degree of Protection IP20”
Cumplimiento
Voltage tolerance is ±5% only with simultaneous frequency tolerance of ±2%, and the combination -5% voltage with +2% frequency is not allowed.
Grid quality must be checked because some voltage/frequency combinations are explicitly prohibited.
Source
“-5% voltage and +2% frequency -> not allowed!”
Installation
The unit is 540x654x1500 mm and weighs 261 kg, so floor loading, transport route and service clearance must be planned.
Affects rigging, doorway clearance and structural floor assessment before delivery.