Product introduction and general description of the Unistat P915w unit
Specpp. 2-3
Technical data according to DIN 12876 including performance parameters
p. 2table, 28 rows
p. 1table, 30 rows
Technical data according to DIN 12876
Parameter
Value
CO2 equivalent 2nd stage
18,2 t
Circulation pump
MK pump
max. delivery
110 l/min
max. delivery pressure
3 bar
Delivery at 0,5 bar
89 l/min
Delivery at 1,0 bar
75 l/min
Delivery at 1,5 bar
60 l/min
Delivery at 2,0 bar
43 l/min
Delivery at 2,5 bar
24 l/min
Pump connection
M30x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water connection
G3/4 male
Consumption at water 15°C, flow 0°C
720 l/h
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,9 l
Filling capacity expansion tank
6,5 l
Overall dimensions WxDxH **
630x704x1565 mm
Net weight
369 kg
sound pressure level +/- 4 dB(A)
64 dB(A)
Power supply requirement
400V 3~ 50Hz
max. current
27 A
Fuse
3x32A
Pressure equipment category
I
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S419528
** Overall dimensions WxDxH
Technical data according to DIN 12876
Parameter
Value
Operating temperature range
-90...250 °C
Temperature stability at -10°C
0,01 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Cooling power with Thermooil at 250°C
6,5 kW
Cooling power with Thermooil at 200°C
7,5 kW
Cooling power with Thermooil at 150°C
9 kW
Cooling power with Thermooil at 100°C
11 kW
Cooling power with Ethanol at 0°C
11 kW
Cooling power with Ethanol at -20°C
11 kW
Cooling power with Ethanol at -40°C
8,2 kW
Cooling power with Ethanol at -60°C
4,2 kW
Cooling power with Ethanol at -80°C
1,3 kW
Cooling power with Ethanol at -90°C
0,6 kW
Refrigeration machine
water-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-507 (A1, H280)
Global Warming Potential (GWP)
3985
Refrigerant quantity
2,9 kg
CO2 equivalent
11,6 t
Refrigerant 2nd stage (ASHRAE, GHS)
R-23 (A1, H280)
Global Warming Potential (GWP)
14800
Refrigerant quantity 2nd stage
1,23 kg
Order-No.: 1080.0008.01; Unistat P915w
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit covers a very wide temperature range from -90 to 250 °C, so it can serve both deep-freeze and high-temperature processes with one circulator.
Defines whether the unit can cover the full process window without a second thermostat.
Source
“Operating temperature range -90...250 °C”
Performance
Cooling capacity collapses at low temperatures (0.6 kW at -90 °C versus 11 kW at 0 °C with ethanol), so deep-cold loads must be sized against the low-temperature end.
Prevents oversizing expectations and ensures heat loads are checked at the coldest operating point.
Source
“at -90°C 0,6 kW”
Installation
The refrigeration machine is water-cooled and requires cooling water between 5 and 20 °C, so a suitable chilled-water supply must be available at the installation site.
Site services and cooling-water temperature must be verified before purchase and installation.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
Cooling water differential pressure must be at least 1 bar and must not exceed 6 bar, which dictates the supply pump and any pressure-reducing valve.
Ensures adequate flow without overpressuring the unit's cooling circuit.
Source
“min. cooling water differential pressure 1 bar”
Installation
The unit draws up to 27 A on a 400 V three-phase supply and needs 3x32 A fusing, so the electrical feed and protection must be sized accordingly.
Determines cable cross-section, breaker rating and whether the existing site supply is sufficient.
Source
“max. current 27 A”
Installation
Three-phase devices below 63 A are delivered with cable but without a plug, so the customer must provide and fit the site plug.
Affects electrical installation planning and the bill of materials for commissioning.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety
Leakage current exceeds 3.5 mA, which requires proper earthing and may call for a higher-rated residual current device.
Impacts electrical safety design and RCD selection during installation.
Source
“Attention: leakage current > 3,5mA”
Cumplimiento
Voltage and frequency tolerances are ±5 % and ±2 % respectively, but the combination of -5 % voltage with +2 % frequency is not allowed.
Prevents operation on unstable or generator-supplied mains that could damage the compressor.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant R-507 has a GWP of 3985 and a CO2 equivalent of 11.6 t, and the second stage R-23 has a GWP of 14800 and 18.2 t CO2e, creating significant F-gas reporting obligations.
Affects environmental compliance, leak-check frequency and total cost of ownership.
Source
“Global Warming Potential (GWP) 3985”
Operation
The pump delivers 110 l/min maximum and 3 bar maximum pressure, with flow falling to 24 l/min at 2.5 bar, so external circuit pressure drop must be matched to the pump curve.
Ensures the external loop and hoses do not exceed the available pump head.
Source
“max. delivery pressure 3 bar”
Installation
The unit requires at least 3.9 l filling capacity and has a 6.5 l expansion tank, so the external volume plus thermal expansion must fit within this budget.
Prevents overflow or insufficient fill when connecting external consumers.
Source
“min. filling capacity 3,9 l”
Installation
The unit weighs 369 kg and measures 630x704x1565 mm, so floor loading, door widths and lifting equipment must be planned.
Determines transport route, floor reinforcement and rigging requirements.
Source
“Net weight 369 kg”
Operation
Ambient temperature must stay between 5 and 40 °C, and the sound pressure level is 64 dB(A), which constrains the room environment and operator exposure.
Ensures the room ventilation and acoustic environment are suitable for continuous operation.
Source
“max. ambient temperature 40 °C”
Maintenance
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used to avoid corrosion.
Guides water quality specification and prevents premature fouling or corrosion of the heat exchanger.