Unistat T305w HT Temperature Control System Datasheet
Contenido
Overviewp. 1
Product introduction and key features of the Unistat T305w HT
Specpp. 1-2
Technical data, performance figures and operating limits per DIN 12876
p. 2table, 7 rows
p. 1table, 32 rows
Technical data according to DIN 12876
Parameter
Value
max. current (3 Phase)
15 A
Fuse (3 phase)
3x16A
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
max. ambient temperature
40 °C
min. ambient temperature
5 °C
from Serial-No.:
S604611
Unistat T305w HT — Technical data according to DIN 12876
Parameter
Value
Operating temperature range
65...300 °C
Minimum temperature with water cooling
15 °C
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
Sensor external connection
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power at 240V
3 kW
Heating power at 230V
3 kW
Heating power at 220V
3 kW
Heating power at 208V
2,7 kW
Heating power at 200V
2,5 kW
Cooling power at 300°C
10 kW
Cooling power at 200°C
10 kW
Cooling power at 100°C
10 kW
Circulation pump: max. delivery
45 l/min
Circulation pump: max. delivery pressure
0,9 bar
Pump connection
M24x1,5 male
max. permissible kin. viscosity
50 mm²/s
Cooling water consumption at water temp. 15°C
240 l/h
Cooling water connection
G1/2 male
min. cooling water differential pressure
1 bar
max. cooling water pressure
6 bar
min. filling capacity
3,5 l
Filling capacity expansion tank
4,5 l
Overall dimensions WxDxH **
426x249x631 mm
Net weight
42 kg
Power supply requirement
200-240V 1~/2~ 50/60Hz
** Overall dimensions WxDxH
Engineering Facts
14 facts · every claim linked to a page
Installation
The unit requires cooling water with inlet temperature between 5°C and 20°C and at least 1 bar differential pressure, so the site water supply must be verified before installation.
Cooling water outside this range will reduce cooling power or cause condensation, affecting selection and site services.
Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Installation
If cooling water temperature falls below the condensation point, the cooling water pipes in the unit must be insulated.
Prevents condensation and corrosion/electrical risk during installation.
Source
“When cooling water temperatures fall below the condensation point, it is essential to insulate the cooling water pipes in the unit.”
Selection
The cooling water circuit uses copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so the cooling water must be compatible with these materials.
Water chemistry must be checked to avoid corrosion or fouling of the heat exchanger.
Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Installation
Voltage tolerance is ±10% only if frequency tolerance does not run in the opposite direction, so supply quality must be verified.
Prevents selecting a unit for a supply that violates the allowed combined voltage/frequency tolerance.
Source
“Voltage + / - 10%, as long as the frequency tolerance does not run in the opposite direction.”
Installation
Single/two-phase devices are supplied with a power cable and country-specific plug that must be specified at order.
Ordering without specifying the plug may result in a non-compliant or unusable power connection.
Source
“Single / two-phase devices (100V to 240V) --> with power cable and country-specific plug (please specify when ordering)”
Installation
Three-phase devices below 63 A are delivered with cable but without plug, so the customer must provide the plug and branch protection.
Affects electrical installation scope and cost.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Cumplimiento
External branch circuit protection according to UL 489 is required.
Mandatory for compliant electrical installation in UL jurisdictions.
Source
“External branch circuit protection according to UL 489 required.”
Installation
Pump connections use bore shape Y (60°) per DIN 3863 and mating hoses need ball socket per DIN 3863 and sleeve nut per DIN 3870.
Ensures correct mating fittings and leak-free hose connection.
Source
“Pump connections: Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation
Cooling power ratings are valid only at 20°C room temperature, 15°C cooling water inlet and 1 bar differential pressure.
Actual cooling capacity will differ if site conditions deviate from these reference values.
Source
“Output data valid for: Room temperature 20°C, cooling water inlet 15°C and 1 bar differential pressure between cooling water inlet and - outlet.”
Operation
Higher cooling water temperature reduces cooling power and may require increased cooling water flow rate.
Must be considered when sizing the cooling water supply and predicting performance.
Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Safety
The unit is classified Class A, Group 1 for electromagnetic disturbance per EN55011.
Class A equipment may not be suitable for residential environments without additional EMC measures.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit has IP20 protection and must be installed in an environment free of water ingress and within 5–40°C ambient.
IP20 limits installation location and requires protection from moisture and dust.
Source
“Degree of Protection IP20”
Selection
The unit is a pressure equipment category Art. 4.3 PED device, which affects documentation and conformity requirements.
Relevant for compliance and procurement documentation in the EU.
Source
“Pressure equipment category Art. 4.3 PED”
Operation
Water exit temperature is limited to 55°C and no steam exits when cooling at circulation temperatures above 100°C.
Important for safe operation and for planning the cooling water drain/return.
Source
“Water exit temperature limited to 55°C. No steam exits on cooling with circulation temperatures over 100°C.”