Unistat T345w HT Technical Data

Contenido

  1. Overview p. 1
    Product introduction and general description of the unit
  2. Spec pp. 1-2
    Technical data according to DIN 12876 and performance figures
  3. p. 2 table, 2 rows
  4. p. 1 table, 34 rows

Technical data according to DIN 12876

ParameterValue
min. ambient temperature5 °C
from Serial-No.:1.1/26

Unistat T345w HT — Technical data according to DIN 12876

ParameterValue
Order-No.1042.0011.01
Operating temperature range65...300 °C
Minimum temperature with water cooling15 °C
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power72 kW
Cooling power at 300°C40 kW
Cooling power at 200°C40 kW
Cooling power at 100°C24 kW
Circulation pumpMK pump
max. delivery200 l/min
max. delivery pressure5,5 bar
Pump connectionM38x1,5 male
max. permissible kin. viscosity50 mm²/s
Cooling water consumption at water temp. 15°C1600 l/h
Cooling water connectionG1 1/4 male
min. cooling water differential pressure1 bar
max. cooling water pressure6 bar
min. filling capacity33 l
Filling capacity expansion tank56 l
Overall dimensions WxDxH **800x1007x1607 mm
Net weight445 kg
Power supply (3 Phase)400V 3~ 50Hz
max. current (3 Phase)115 A
Fuse (3 phase)3x125A
Degree of ProtectionIP20
max. ambient temperature40 °C

** Overall dimensions WxDxH

Engineering Facts

12 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 125 A fuse per phase, so the electrical installation must be sized for a 115 A maximum current.

Determines the site power distribution, cable cross-section and protection device rating before ordering.

Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation

The cooling water circuit needs a minimum differential pressure of 1 bar and can withstand a maximum of 6 bar, with G1 1/4 male connections.

Defines the required cooling water supply pressure and the fittings/adapters needed for connection.

Source
“min. cooling water differential pressure 1 bar”
Performance

Cooling water consumption is 1600 l/h at 15 °C water temperature, so the facility must provide that flow rate continuously.

Sizing of the cooling water supply and drain, and verification that the building loop can deliver this flow.

Source
“Cooling water consumption at water temp. 15°C 1600 l/h”
Operation

The unit has no active refrigeration system, so reaching the 65 °C minimum target depends on adequate heat loss from the application or suitable water cooling.

Prevents specifying the unit for cooling duties it cannot perform without external heat removal.

Source
“These devices do not have an active refrigeration system.”
Installation

The cooling water intake temperature must be between 5 °C and 20 °C, and pipes must be insulated if the water temperature falls below the condensation point.

Ensures the site cooling water temperature is compatible and that condensation is avoided.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Cumplimiento

Voltage and frequency tolerances are ±5 % and ±2 % respectively, but they must not be applied in opposite directions simultaneously.

Confirms whether the site power quality is acceptable for warranty and safe operation.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Safety

The leakage current exceeds 3.5 mA, which requires appropriate protective earthing and possibly a higher-rated residual current device.

Affects the electrical protection concept and RCD selection during installation.

Source
“Attention: leakage current > 3,5mA”
Installation

Three-phase devices drawing more than 63 A are delivered without cable and without plug, so the customer must provide the power connection.

Clarifies the scope of supply and the need to plan for on-site cabling and a suitable isolator.

Source
“Three-phase devices with current consumption greater than 63A --> without cable, without plug”
Installation

The pump connection is M38x1.5 male with a bore shape Y (60°) per DIN 3863, so hoses must use a matching ball socket and sleeve nut.

Ensures the correct mating hose fittings are ordered to avoid leaks or delays.

Source
“Pump connection M38x1,5 male”
Operation

The maximum permissible kinematic viscosity of the thermofluid is 50 mm²/s, which limits the choice of heat transfer fluid at low temperatures.

Prevents selecting a fluid that is too viscous for the pump at the operating temperature.

Source
“max. permissible kin. viscosity 50 mm²/s”
Installation

The unit has IP20 protection and a maximum ambient temperature of 40 °C, so it must be installed in a dry, ventilated indoor location.

Rules out outdoor or wet/dusty environments and informs room ventilation requirements.

Source
“Degree of Protection IP20”
Maintenance

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only compatible cooling water may be used.

Guides water quality specification to avoid corrosion or scaling of the heat exchanger.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”