Product introduction and main features of Unistat T402
Specpp. 1-2
Technical data, performance and operating parameters per DIN 12876
p. 2table, 5 rows
p. 1table, 23 rows
Technical data according to DIN 12876
Characteristic
Value
Pressure equipment category
Art. 4.3 PED
Degree of Protection
IP20
min. ambient temperature
5 °C
max. ambient temperature
40 °C
from Serial-No.:
S622126
Unistat T402 Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
80... 425 °C
Temperature stability at 70°C
0,05 K
temperature set point / display
5,7" colour Touchscreen
Resolution of display
0,01 K
Internal temperature sensor
Pt100
External sensor
Pt100
Interface digital
Ethernet, USB (Host u. Device), RS232
digital input
ECS ONE
digital output
POKO ONE
Alarm message
optic, acoustic, relay
Safety classification
III / FL
Heating power
6 kW
Circulation pump: max. delivery
45 l/min
Circulation pump: max. delivery pressure
0,9 bar
Pump connection
M24x1,5 male
Cooling water connection
G1/2 male
min. filling capacity
1,45 l
Filling capacity expansion tank
2,8 l
Overall dimensions WxDxH **
506x400x765 mm
Net weight
62 kg
Power supply (3 Phase)
400V 3~N 50/60Hz
max. current (3 Phase)
17 A
Fuse (3 phase)
3x20A
** The dimensions specified below are approximate and may be subject to change.
Engineering Facts
14 facts · every claim linked to a page
Selection
The unit requires a three-phase 400 V supply with up to 17 A per phase, so the electrical infrastructure must be sized accordingly.
Determines whether the site has a suitable three-phase supply and correct breaker/fuse rating before purchase.
Source
“Power supply (3 Phase) 400V 3~N 50/60Hz”
Installation
Three-phase devices drawing less than 63 A are delivered with cable but without a plug, so the customer must provide and install the plug.
Affects installation planning and the need to procure a suitable plug and have a qualified electrician connect it.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The pump connection is an M24x1.5 male thread, so mating hoses or adapters must match this thread.
Prevents ordering incompatible hoses or fittings and avoids leaks or delays during commissioning.
Source
“Pump connection M24x1,5 male”
Installation
The cooling water connection is G1/2 male, so the site cooling water supply must be fitted with a matching G1/2 female coupling.
Ensures the cooling water circuit can be connected without adapters and that the correct thread is available on site.
Source
“Cooling water connection G1/2 male”
Installation
The expansion tank lid can be fixed and has connections for a nitrogen blanket at a maximum pressure of 0.2 bar.
Relevant when inert gas blanketing is required; exceeding 0.2 bar could damage the tank or create a safety hazard.
Source
“The lid has connections for a nitrogen blanket (maximum pressure 0,2 bar)”
Operation
The ambient temperature must stay between 5 °C and 40 °C for safe and reliable operation.
Sites outside this range may experience degraded performance, condensation, or overheating and should be assessed before installation.
Source
“min. ambient temperature 5 °C”
Operation
The maximum ambient temperature is 40 °C, so the installation location must not exceed this limit.
Exceeding this limit can reduce cooling capacity and cause overheating or protective shutdowns.
Source
“max. ambient temperature 40 °C”
Safety
The unit has an adjustable overtemperature protection according to DIN 12876, which must be set correctly for the application.
Correct setting of the overtemperature protection is essential to prevent overheating of the heat transfer fluid and a fire hazard.
Source
“With adjustable overtemperature protection according to DIN 12876.”
Safety
The unit is classified as safety class III / FL according to DIN 12876, which defines the permitted fluids and operating conditions.
The safety class determines which heat transfer fluids may be used and what protective measures are required.
Source
“Safety classification III / FL”
Performance
The circulation pump delivers up to 45 l/min at a maximum pressure of 0.9 bar, which must be sufficient for the connected external circuit.
Ensures the pump can overcome the pressure drop of the external loop and maintain the required flow for heat transfer.
Source
“max. delivery 45 l/min”
Performance
The heating power is 6 kW, which determines heat-up time and the maximum heat load that can be handled.
Needed to verify that the unit can reach the required temperature within the desired time and support the process heat load.
Source
“Heating power 6 kW”
Maintenance
The minimum filling capacity is 1.45 l and the expansion tank capacity is 2.8 l, which must be respected when filling the system.
Underfilling can cause pump cavitation and poor temperature control; overfilling can cause overflow when the fluid expands.
Source
“min. filling capacity 1,45 l”
Cumplimiento
The unit is classified as Class A, Group 1 for electromagnetic disturbance according to EN55011, which may require additional filtering in residential areas.
Class A equipment is not intended for residential use without additional measures; this affects site selection and EMC compliance.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The overall dimensions are 506 x 400 x 765 mm and the net weight is 62 kg, so the installation location must provide adequate space and floor loading.
Needed to plan the installation footprint, access for service, and ensure the floor can support the weight.