IDP-7 Dry Scroll Vacuum Pump Instructions for Use

Contenido

  1. Quality p. 3
    Warranty terms and conditions for the product
  2. Overview p. 5
    Table of contents listing document sections
  3. Procedure pp. 7-14
    Operating and safety instructions for the pump
  4. Spec pp. 15-26
    Technical specifications, dimensions, and performance data
  5. Services p. 27
    Health and safety form for product return
  6. p. 32
    Agilent vacuum division contact information
  7. p. 19 table, 9 rows
  8. p. 15 table, 30 rows
  9. p. 23 table, 22 rows
  10. p. 24 table, 3 rows
  11. p. 27 table, 4 rows
  12. p. 20 table, 2 rows

Table 2 IDP-7 Power Cord Selection

CountryPower Cord SpecificationOrder
Europe10 A / 220-230 VAC, 2.5 m656494220
Denmark10 A / 220-230 VAC, 2.5 m656494225
Switzerland10 A / 230 VAC, 2.5 m656494235
UK/Ireland13 A / 230 VAC, 2.5 m656494250
India10 A / 220-250 VAC, 2.5 m656494245
Israel10 A / 230 VAC, 2.5 m656494230
Japan12 A / 100 VAC, 2.3 m656494240
North America15 A / 125 VAC, 2.0 m656458203
10 A / 230 VAC, 2.5 m656494255

Table 1 Specifications

SpecificationValue
ModelIDP-7 Dry Scroll Single Hermetic Vacuum Pump
Dimensions248 x 260 x 421 (h x w x l [mm])
Peak pumping speed50 Hz: 120 L/m, 7.2 m3/hr 60 Hz: 152 L/min (9.1 m3/hr)
MediaNo corrosive, explosive or particulate forming gases
Maximum inlet pressure1.0 atmosphere (0 psig)
Maximum recommended continuous inlet pressure150 T (200 mbar)
Maximum outlet pressure115V range: 5 psig, 115 VAC 50/60HZ or greater, derate linearly to 0 psig at 90 VAC below 115 VAC 230V range: 5pisg, 230 VAC 50/60HZ or greater, derate linearly to 0 psig at 180 VAC below 230 VAC
Ultimate pressure*20 mT at 60 Hz, 30 mT at 50 Hz
Inlet connectionNW25
Exhaust connectionNW16
Gas ballastFemale 1/8" National Pipe Thread, (20 micron sintered plug provided) SAE-2 5/16-24
Ambient operating temperature5 °C to 40 °C (41 °F to 104 °F)
Storage temperature–20 °C to 60 °C (–4 °F to 140 °F)
Motor rating3/8 HP 0 (280 W)
Operating voltages**Selectable: 100-120 VAC, 50/60 HZ, 6 A typical; 200-240 VAC 50/60 HZ, 3 A typical; single phase
Motor thermal protectionAutomatic thermal protection
Operating speed50 Hz: 1450 RPM 60 Hz: 1750 RPM
Cooling systemAir-cooled
WeightPump only: 24.5 kg (54 lbs) Shipping weight: 28.1 kg (62 lbs)
Leak rate (with exhaust sealed)<1 x 106 secs helium
Noise Level (per ISO 11201)52 +/- 2 dB(A)
Vibration level at inlet (per ISO 10816-1)Class 1B, 7 mm/s
Fuse TypeM5 x 20 mm, 10 A, Slo-Blo, Littelfuse Model 215010 or equivalent
Hour meterIntegral Hour Meter provided, displays running time to closest 1 hour
Emissions and ImmunityEN 61236-1:2006 as a Class A device
SafetyEN/IEC 61010 3rd Edition CSA certified for US and Canada
Ambient Relative Humidity5-95%; non-condensing
Water Vapor Handling120 g/hr
InstallationIndoor use, Installation Category II, Pollution Degree 2
Altitude2000 m (6562’)

*If the pump and/or vacuum system is new, or has been exposed to atmosphere for an extended period of time, it is possible that water can be present which can affect the pump’s ability to achieve ultimate pressure. To remove water from the system it is recommended to open GB1 for 15 minutes and/or flush with dry nitrogen. **Mains supply voltage fluctuations shall not exceed 10% of the nominal voltage.

Table 4 Troubleshooting Chart

ProblemPossible CausesCorrective Actions
Will not startFuse blownReplace fuse.
Identify cause of overload.
Check the line voltage and the voltage configuration of the pump.
Motor thermal protector openAllow motor to cool.
Identify cause of overload.
Excessive voltage dropCheck size and length of cable.
Defective motorInspect. Contact Agilent.
Poor ultimate pressureSystem leakLocate and repair leak.
Water in pumpFlush pump with air or dry nitrogen.
Gas ballast pluggedReplace breather vent. Contact Agilent.
Solvent in pumpFlush pump with air or dry nitrogen.
Install trap or filter.
Seals worn outRebuild pump.
Poor conductance to pumpReplumb with shorter and/or larger diameter tubing.
Hammering noisePump overheatedCheck ventilation to pump.
Check ambient temperature.
Debris in pumpCheck intake screen.
Flush pump.
Disassemble pump and inspect.
Pump runs intermittentlyMotor thermal protector is cycling open and closed.*Allow the motor to cool.
Identify the cause of the overload.
Service, Maintenance, Tip Seal, or Part ReplacementWear or damageSee Table 5 on page 18 and Table and Table 6 on page 18.

* The IDP-7 is equipped with an auto-reset thermal motor protector. This protector will automatically shut down the pump when it detects an overload condition, and will automatically restart the pump when the motor has cooled to within an acceptable temperature range.

Table 5 Accessories and Replacement Kits

Part NumberDescriptionContents
X3807-67002Front Cowling AssemblyReplacement front cowling material to replace a broken plastic cowling.
X3807-67003Rear Cowling AssemblyReplacement rear cowling material to replace a broken plastic cowling.
X3807-68008Gas Ballast KitSpare Parts

RETURN THE COMPLETED REQUEST FOR RETURN FORM TO YOUR NEAREST LOCATION:

EUROPE:NORTH AMERICA:PACIFIC RIM:
Fax: 00 39 011 9979 330Fax: 1 781 860 9252please visit our website for individual office information
Fax Free: 00 800 345 345 00Toll Free: 800 882 7426http://www.agilent.com
Toll Free: 00 800 234 234 00vpl-ra@agilent.com
vpt-customercare@agilent.com

Table 3 Typical IDP-7 Maximum Starting Current (A)

FrequencyHigh Voltage Range 100 V120 V127 V200 V230 V240 V
60 Hz22252512N/A14
50 Hz23N/AN/A121415

Engineering Facts

14 facts · every claim linked to a page
Selection

The pump is only suitable for air or inert gases and must not be used with corrosive, explosive, or particulate-forming gases.

Process gas compatibility determines whether this pump can be selected for the application at all.

Source
“This pump is suitable for pumping air or inert gases. The pump is not intended to pump corrosive, explosive, or particulate-forming gases.”
Installation

The pump must be powered through an IEC-320 cord rated at least 10 A.

Undersized power cords can overheat or cause voltage drop during startup.

Source
“Connect the pump to the power supply using an IEC-320 style power cord of at least 10 A capacity”
Installation

The pump requires at least 60 cm clear space on all sides for ventilation, maintenance access, and breaker access.

Insufficient clearance causes overheating and prevents safe access to the power disconnect.

Source
“There must be a clear space of at least 60 cm (24") on all sides of the equipment.”
Installation

The pump is delivered configured for high voltage and must be reconfigured for low-voltage mains.

Connecting a high-voltage-configured pump to low-voltage mains will cause malfunction or damage.

Source
“The pump as delivered from the factory is configured for high voltage.”
Installation

The pump must be grounded through a properly installed grounding plug; adapters are not permitted.

Proper grounding is required to reduce electric shock risk in case of short circuit.

Source
“This product must be grounded.”
Installation

Extension cords, if used, must be at least 16-gage wire and no longer than 25 feet (7.6 m).

Undersized or overly long extension cords cause voltage drop and overheating.

Source
“Agilent recommends using only extension cords with a minimum of 16-gage wire and a maximum length of 25 feet (7.6 m).”
Performance

Maximum starting current can reach 25 A at 120 V, so supply and protection must be sized accordingly.

Inrush current affects breaker selection and verifies the supply can handle startup without nuisance trips.

Source
“Table 3 lists the typical IDP-7 maximum starting current and duration.”
Installation

Use short, large-diameter NW25 or larger inlet hardware and a bellows to isolate vibration and relieve strain.

Long or narrow inlet lines reduce effective pumping speed and transmit vibration to the chamber.

Source
“Insert a bellows between the pump and vacuum chamber to provide both vibration isolation and strain relief.”
Installation

Do not restrict exhaust flow with long small-diameter tubing; use short NW16 or larger hardware to avoid overheating.

Exhaust restriction raises back pressure and can overheat the pump.

Source
“To avoid overheating the pump, do not restrict the exhaust flow with long lengths of small diameter tubing.”
Operation

Gas ballast is required when pumping water vapor or condensable gases; both ports together handle up to 120 g/hr.

Exceeding water vapor handling capability shortens tip seal life and degrades ultimate pressure.

Source
“With both gas ballast 1 and gas ballast 2 in use, up to 120 g/hr water vapor can be handled.”
Operation

To reach maximum vacuum, the pump must run at least 24 hours with the inlet sealed.

Acceptance testing before this stabilization period may falsely indicate poor performance.

Source
“In order to reach maximum vacuum, the pump must be left running for at least 24 hours with the inlet sealed.”
Maintenance

Newly installed tip seals require several hours of run time to seat and meet speed and base pressure specifications.

Performance should not be judged immediately after tip seal replacement.

Source
“Newly installed Tip Seals may require several hours of run time to seat properly and enable the pump to meet speed and base pressure specifications.”
Safety

The outer surface can exceed 60 °C; allow the pump to cool before maintenance.

Hot surfaces present a burn hazard during maintenance after extended operation.

Source
“allow it to cool as the temperature of the outer surface may be in excess of 60 °C.”
Safety

Always disconnect electrical power before beginning maintenance work.

Prevents electric shock and accidental start during service.

Source
“Always disconnect your power supply to the pump before beginning maintenance work.”