Improving Food Safety Analysis with LC Q TOF Feb2020
Technical Reference Summary
Covered Models & Document Scope
Ebook patrocinado de LCGC (febrero de 2020) que abarca flujos de trabajo de análisis de seguridad alimentaria con el Agilent 6546 LC/Q-TOF, el Agilent 1290 Infinity II LC, MassHunter Quant 10.1 y LC/Q-TOF Screener Tool; incluye análisis de pesticidas en matrices alimentarias complejas, cribado no dirigido de adulterantes/contaminantes y pruebas de autenticidad alimentaria con MassHunter Classifier. No se indican códigos de pedido de instrumentos ni variantes de tamaño.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Mass resolution type | high mass resolution p.3 ↗ | Agilent 6546 LC/Q-TOF |
| Mass accuracy | high mass accuracy p.3 ↗ | Agilent 6546 LC/Q-TOF |
| Mass accuracy requirement for identification | less than 5 ppm p.4 ↗ | SANTE/12682/2019, two ions |
| Minimum number of ions for identification | two ions with at least one fragment ion p.4 ↗ | SANTE/12682/2019 |
| Minimum acquisition rate for quantitation | 2 Hz p.5 ↗ | 12 points over 6 s chromatographic peak |
| Data acquisition speed | 6 Hz p.7 ↗ | All Ions acquisition, strawberry method |
| Chromatographic peak width | about six s p.5 ↗ | typical peak |
| Dynamic range channels | two channels (low-gain and high-gain), dual-gain 10-GHz spectrum p.5 ↗ | Agilent 6546 LC/Q-TOF |
| LC back pressure rating | 1,300 bar p.3 ↗ | Agilent 1290 Infinity II LC |
| Detector data rate | 240 Hz p.3 ↗ | VWD and DAD |
| Autosampler carryover | less than 9 ppm p.3 ↗ | Agilent 1290 Infinity II LC |
| Autosampler sample capacity | 6,144 samples in 16x384 well plates p.3 ↗ | Agilent 1290 Infinity II LC |
| SureMass data conversion time | 3–5 minutes per sample p.6 ↗ | post-acquisition script |
| Number of pesticides quantified | 180 p.6 ↗ | proof of concept, four matrices |
| Number of pesticides screened | over 200 p.6 ↗ | proof of concept, four matrices |
| Mass error achieved | less than 2 ppm p.8 ↗ | almost all analytes at two concentrations |
| Column length | 150 mm p.7 ↗ | C18 column with guard column |
| Extraction solvent volume | 10 mL p.7 ↗ | acetonitrile, strawberry QuEChERS |
Operating Limits
- minimum mass accuracy for identification: less than 5 ppm p.4 ↗
- minimum acquisition rate for quantitation: 2 Hz p.5 ↗
- LC maximum back pressure: 1,300 bar p.3 ↗
- autosampler carryover: less than 9 ppm p.3 ↗