Rev c capture portable fuming systems user manual
Technical Reference Summary
Covered Models & Document Scope
Cubre los sistemas de fumigación portátiles Labconco CApture, modelos 3420000 (115 VAC, 60 Hz, 1 Amp) y 3420020, 3420022, 3420024 (230 VAC, 50/60 Hz, 0.5 Amp). Los accesorios opcionales incluyen el kit de conexión de escape (#3425400).
Section-by-Section Technical Breakdown
Enumera los suministros necesarios (pegamento CA, bandejas de calentamiento de aluminio, tratamiento Rain-X o Rain Block) y especifica los requisitos eléctricos y de escape. Los datos eléctricos se tabulan por modelo, y las notas cubren las clasificaciones de los disyuntores, la longitud del cable de alimentación y el funcionamiento con inversor.
- CA glue capacity of filter is approximately 2 oz (57 g); do not use thickening or gap-filling formulations.
- Aluminum heating pans: 10 shipped; bag of 250 available as Labconco part 34252002.
- Exhaust Connection Kit optional; do not add additional duct beyond that supplied.
- Model 3420000: 115 VAC, 60 Hz, 1 Amp; models 34200-20, -22, -24: 230 VAC, 50/60 Hz, 0.5 Amp.
- Circuit breaker: 10 A for 115 V models, 5 A for 230 V models; power cords 3 m (10 ft).
- Inverter operation: 12 VDC to AC, output rating at least 100 W, protected against current overload.
Explica la polimerización del CA, el calentamiento controlado para generar vapores, la cámara transparente de policarbonato, el sistema de filtro de dos etapas y el soplador. Incluye límites de temperatura y datos de eficiencia del filtro.
- CA boiling point approximately 120°F (50°C); decomposition products above 250°F (121°C); hydrogen cyanide can be produced above 400°F (204°C).
- Controller setpoint range: 120°F (50°C) to 350°F (177°C); redundant safety switch limits heater to 350°F (177°C).
- Clear chamber must be treated with Rain-X or Rain Block; protection lasts up to 20 uses if undisturbed.
- First filter stage: polyester loft particulate filter, MERV 11 (65-79% of 1.0-3.0 µm particles; >85% of 3.0-10 µm particles).
- Second filter stage: granular activated carbon (GAC) adsorbs gaseous CA.
- Filter life: minimum 100 fuming cycles (about 2 oz or 57 ml of CA).
Cubre el desembalaje, el transporte, la identificación de componentes, las indicaciones de manipulación del CA, la configuración, el ajuste del calentador, el procedimiento de fumigación y el ajuste de la humedad. Proporciona instrucciones de funcionamiento paso a paso y valores de parámetros clave.
- System weight approximately 22 lbs (10 kg); transport only in transport mode with heater/filter module nested and transport lid locked.
- Setup: release clamps, remove foam insert, lower leveling foot, treat chamber, select hooks/shelves, seal chamber, plug in and switch main power ON.
- Heater adjustment: push control switch to Heat; setpoint range 120-350°F (50-177°C); temperature stabilizes within 3-5 minutes at setpoint ±15°F.
- Fuming: use 4-6 drops of CA (never exceed 8 drops); typical development 3-5 minutes with 4 drops at 350°F (177°C).
- After fuming, switch to Vent and let chamber vent at least 3-5 minutes before opening.
- Humidity: optimum 70-90% RH, best at approximately 80%; one drop of water raises RH by approximately 2%.
Proporciona un programa de mantenimiento rutinario y operaciones de servicio, incluido el reemplazo del filtro y el reinicio del disyuntor. Especifica los intervalos de limpieza y la reaplicación del tratamiento.
- Weekly: clean stainless steel face plate with soap/detergent or LabSolutions wipes (part #1570000); reapply Rain-X or Rain Block if chamber interior cleaned.
- Monthly: clean exterior surfaces with damp cloth or LabSolutions wipes.
- After every 20 development cycles: clean clear chamber with mild detergent or glass cleaner, then treat with Rain-X or Rain Block.
- After every 100 development cycles: replace filter cartridge.
- Filter replacement: unplug, separate module, loosen captive thumbscrew, remove panel, pull cartridge out of seal and flexible connector, install replacement, ensure seal not deformed.
- Circuit breaker reset: press switch on rear of steel module to ON (I) position.
Presenta una tabla de solución de problemas con problemas, causas y acciones correctivas para fallas comunes del sistema. Cubre errores de alimentación, revelado de huellas, generación de vapores, calentador, ventilador y controlador.
- Blower and controller won't turn on: check outlet and circuit breaker.
- Prints not developing properly: insufficient humidity; add 1-2 drops of water in a heating pan.
- Little or no fumes: heater temperature too low or pan bottom dented/bent.
- Heater temperature not changing: possible over-temperature switch failure or defective controller.
- Fan does not run: fan or 12 V power supply failure.
- Controller displays 'Er' or 'oo' or '-': controller error.
Enumera piezas de repuesto y accesorios con números de pieza y cantidades. Incluye cámara transparente, bandejas de calentamiento, conector flexible, cartucho de filtro, estante, gancho e inserto de almacenamiento de espuma.
- Clear chamber: part 3421901.
- Heating pans, package of 250: part 3425202.
- Flexible connector: part 3424500.
- Filter cartridge: part 3421801.
- Shelf: part 3422203; Hook: part 3422202.
- Foam storage insert: part 3425500.
Proporciona las dimensiones interiores de la cámara transparente.
- Clear chamber interior: 16" (406 mm) H x 8" (203 mm) W x 8" (203 mm) D.
Enumera los datos eléctricos y las condiciones ambientales de funcionamiento del sistema.
- Electrical: 3420000 115 VAC, 60 Hz, 1 Amp, 75 W; 34200-20, -22, -24 230 VAC, 50/60 Hz, 0.5 Amp, 75 W.
- Indoor use only; maximum altitude 6562 ft (2000 m).
- Ambient temperature range: 41°F to 104°F (5°C to 40°C).
- Maximum relative humidity: 80% up to 88°F (31°C), decreasing linearly to 50% at 104°F (40°C).
- Main supply voltage fluctuations not to exceed ±10% of nominal.
- Transient overvoltages per Installation Category II: up to 1500 V for 115 V models, 2500 V for 230 V models.
- Pollution degree 2 environment.
Enumera accesorios opcionales y de repuesto con números de pieza, descripciones y pesos.
- Clear Chamber (#3421901): replacement clear chamber with latches, 2 lb (0.9 kg).
- Heating Pans (#3425202): package of 250 aluminum disposable pans, 1 lb (0.5 kg).
- Filter Cartridge (#3421801): replacement particulate/GAC filter, lasts 100 cycles, 5 lbs (2.3 kg).
- Exhaust Connection Kit (#3425400): allows remote exhaust or fume hood connection; includes 10 ft (1.2 m) extended, 4" (100 mm) diameter collapsible duct, 1 lb (0.5 kg).
- Replacement Hook (#3422202): for hanging evidence, 1 lb (0.5 kg).
- Replacement Shelf (#3422203): for supporting small evidence, 1 lb (0.5 kg).
- Perforated Shelf Cover (#3426101): stainless steel cover fits over Shelf #3426103, 1 lb (0.5 kg).
Key Integration, Operating & Safety Requirements
| Specification / Condition | Engineering Implication | Aplica a |
|---|---|---|
| El modelo 3420000 requiere 115 VAC, 60 Hz, 1 Amp; los modelos 34200-20, -22, -24 requieren 230 VAC, 50/60 Hz, 0.5 Amp. p.7 ↗ | Select model according to available mains voltage; use a circuit breaker rated 10 A for 115 V models or 5 A for 230 V models. | whole series |
| Los cables de alimentación suministrados tienen 3 metros (10 pies) de longitud. p.7 ↗ | Locate the system within 3 m of a suitable outlet; use an extension cord only if adequately rated. | whole series |
| El sistema puede funcionar con un inversor que convierta 12 VDC a CA, con una potencia nominal de salida de al menos 100 W y protección contra sobrecarga de corriente. p.7 ↗ | Enables mobile operation from a vehicle 12 V supply; ensure inverter meets minimum rating. | whole series |
| El punto de ebullición del CA es aproximadamente 120°F (50°C); los productos de descomposición aparecen por encima de 250°F (121°C); se puede producir cianuro de hidrógeno por encima de 400°F (204°C). p.9 ↗ | Never exceed 350°F (177°C) setpoint; the redundant safety switch limits heater to 350°F (177°C). | whole series |
| El rango de consigna del controlador es de 120°F (50°C) a 350°F (177°C). p.9 ↗ | Program setpoint within this range; temperatures stabilize within 3-5 minutes at setpoint ±15°F. | whole series |
| La cámara transparente debe tratarse con Rain-X o Rain Block; la protección dura hasta 20 usos si no se altera. p.9 ↗ | Reapply treatment after every 20 development cycles or whenever chamber interior is cleaned. | whole series |
| La primera etapa del filtro es MERV 11, que elimina el 65-79% de las partículas de 1.0-3.0 µm y >85% de las partículas de 3.0-10 µm. p.9 ↗ | Provides particulate filtration; replace filter cartridge after 100 development cycles or when odor is detected. | whole series |
| La vida útil del filtro es de un mínimo de 100 ciclos de fumigación usando aproximadamente 2 oz (57 ml) de CA. p.9 ↗ | Plan filter replacement after 100 cycles; part number 3421801. | whole series |
| Nunca exceda las 8 gotas de CA por ciclo de revelado; el procedimiento típico utiliza 4-6 gotas. p.11 ↗ | Exceeding 8 drops may overload filter and produce excessive fumes; use 4 drops for typical 3-5 minute development at 350°F (177°C). | whole series |
| La humedad relativa óptima para el revelado de huellas dactilares es del 70-90%, mejor a aproximadamente 80%; una gota de agua aumenta la HR en aproximadamente 2%. p.20 ↗ | Use the humidity chart to add water drops: e.g., to raise RH by 30%, add 15 drops. | whole series |
| El tiempo de revelado típico es de 3-5 minutos con 4 gotas de CA a 350°F (177°C). p.21 ↗ | Monitor development and switch to Vent when prints are developed as desired. | whole series |
| Después de la fumigación, deje ventilar la cámara durante al menos 3-5 minutos antes de abrirla. p.22 ↗ | Ensures CA fumes are removed before retrieving evidence. | whole series |
| El peso del sistema es de aproximadamente 22 lbs (10 kg). p.11 ↗ | Follow safe-lifting guidelines when transporting; use transport mode with heater/filter module nested and transport lid locked. | whole series |
| Dimensiones interiores de la cámara transparente: 16" (406 mm) de alto x 8" (203 mm) de ancho x 8" (203 mm) de profundidad. p.28 ↗ | Limits evidence size; ensure evidence fits within chamber. | whole series |
| Límites ambientales: uso en interiores, altitud máxima 6562 ft (2000 m), ambiente 41-104°F (5-40°C), HR máxima 80% hasta 88°F (31°C) disminuyendo al 50% a 104°F (40°C). p.29 ↗ | Install in a location meeting these conditions; voltage fluctuations not to exceed ±10%. | whole series |
| El kit de conexión de escape (#3425400) incluye un conducto plegable extendido de 10 ft (1.2 m), de 4" (100 mm) de diámetro; no agregue conducto adicional. p.6 ↗ | Additional duct reduces exhaust flow and may cause incomplete venting. | whole series |
| Las bandejas de calentamiento deben desecharse después del segundo uso; no use bandejas con fondos abollados o deformados. p.16 ↗ | Residue buildup affects heat transfer; damaged pans heat CA inefficiently. | whole series |
| Almacene el pegamento CA abierto en un congelador o recipiente hermético con gel de sílice para evitar que fragüe. p.16 ↗ | Extends usable life of CA by limiting moisture exposure. | whole series |
Faults & Troubleshooting
| Código / Síntoma | Possible Cause | Remedy | Note |
|---|---|---|---|
| Blower and controller won’t turn on p.26 ↗ | Unit not plugged into outlet | Plug the cabinet into appropriate electrical service. | |
| Blower and controller won’t turn on p.26 ↗ | Circuit breaker tripped | Reset circuit breaker. | |
| Prints are not developing properly p.26 ↗ | Insufficient humidity in the clear chamber | Place 1-2 drops of water in a heating pan, and place on the heater. | |
| Little or no fumes are being produced p.26 ↗ | Heater temperature is too low. | Increase heater temperature | |
| Little or no fumes are being produced p.26 ↗ | Bottom of heater pan is dented/bent | Discard damaged pan, and use one with flat bottom. | |
| Heater temperature not changing on controller (“out” indicator is on) p.26 ↗ | Heater /Over temperature switch failure | Contact Labconco Product Service Department. | |
| Heater temperature not changing on controller (out indicator is off) p.26 ↗ | Controller is defective | Contact Labconco Product Service Department. | |
| Fan does not run p.26 ↗ | Fan or 12 volt power supply failure | Contact Labconco Product Service Department. | |
| Controller displays “Er” or “oo” or “-” p.26 ↗ | Controller error | Contact Labconco Product Service Department. |