c10g-e089
Technical Reference Summary
Covered Models & Document Scope
Espectrofotómetro UV-VIS UV-1900i; la sección del documento abarca la cuantificación de ácidos nucleicos (medición de trazas de ADN de doble cadena mediante celdas TrayCell y Nano Stick).
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Sample volume | 10 mm standard cell = 2.5 to 4.0 mL; TrayCell = 0.7 to 10 μL; Nano Stick = 2 μL min. p.3 ↗ | UV-1900i |
| Wavelength range | 190 to 1,100 nm p.3 ↗ | UV-1900i |
| Spectral bandwidth | 1 nm p.3 ↗ | UV-1900i |
| Light source | 20 W halogen lamp and deuterium lamp p.3 ↗ | UV-1900i |
| Monochromator | LO-RAY-LIGH grade blazed holographic grating in Czerny-Turner mounting p.3 ↗ | UV-1900i |
| Detector | Silicon photodiode p.3 ↗ | UV-1900i |
| Sample compartment internal dimensions | W 110 × D 250 × H 115 mm p.3 ↗ | UV-1900i |
| Distance between light beams | 100 mm p.3 ↗ | UV-1900i |
| Dimensions | W 450 × D 501 × H 244 mm p.3 ↗ | UV-1900i |
| Weight | 16.6 kg p.3 ↗ | UV-1900i |
| Output device | USB memory (optional); Extended memory (optional) p.3 ↗ | UV-1900i |
| Display | 24-bit color touch screen p.3 ↗ | UV-1900i |
| Scan speed | up to 29,000 nm/min p.3 ↗ | UV-1900i |
Operating Limits
- Minimum sample volume with Nano Stick: 2 μL min. p.3 ↗
- Minimum sample volume with TrayCell: 0.7 μL p.3 ↗