c190-e092
Technical Reference Summary
Covered Models & Document Scope
Sistema Nexera™ GPC (sistema de cromatografía de permeación en gel), configuración estándar: controlador de sistema CBM-40, unidad de suministro de disolvente LC-40D, unidad de desgasificación DGU-403, automuestreador SIL-40C, horno de columna CTO-40C, detector de índice de refracción RID-20A, estación de trabajo LabSolutions con software LabSolutions GPC; columnas de la serie Shim-pack GPC-800/80M/2000 listadas por separado.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| System configuration | Modular p.12 ↗ | Nexera GPC System standard |
| Mobile phase flow rate accuracy | No more than 1% or 2 µL/min, whichever is greater (0.01 to 2 mL/min) p.12 ↗ | Nexera GPC System standard |
| Mobile phase flow rate precision | No more than 0.06% RSD or 0.02 min SD, whichever is greater p.12 ↗ | Nexera GPC System standard |
| Detector cell volume | 9 µL p.12 ↗ | RID-20A detector |
| Detector cell temperature control range | 30 to 60 °C p.12 ↗ | RID-20A detector |
| Calibration curve approximation formulae | Linear, 3rd-order, 3rd-order + hyperbolic curve, 5th-order, 5th-order + hyperbolic p.12 ↗ | LabSolutions GPC Software |
| Data output method | ASCII format p.12 ↗ | LabSolutions GPC Software |
| Mobile phase containers monitored | up to twelve containers p.4 ↗ | 1-liter containers; up to 4 with 2- to 5-liter containers |
| Solvent usage savings | 30 % p.7 ↗ | RID-20A GPC analysis |
| Stabilization time | After 30 Minutes p.7 ↗ | Power ON, RID-20A |
Operating Limits
- Detector cell temperature control range: 30 to 60 °C p.12 ↗
- Mobile phase flow rate range for accuracy specification: 0.01 to 2 mL/min p.12 ↗
- Mobile phase containers monitored: up to 12 (1 L) or up to 4 (2–5 L) p.4 ↗
Ordering Data
| Ordering Code | Part No. | Variant / Code | Aplica a | Note |
|---|---|---|---|---|
| 228-65502-58 p.12 ↗ | 228-65502-58 | System Controller CBM-40, qty 1 | ||
| 228-65002-58 p.12 ↗ | 228-65002-58 | Solvent Delivery Unit LC-40D, qty 1 | ||
| 228-65018-58 p.12 ↗ | 228-65018-58 | Degassing Unit DGU-403, qty 1 | ||
| 228-65101-58 p.12 ↗ | 228-65101-58 | Autosampler SIL-40C, qty 1 | ||
| 228-65202-58 p.12 ↗ | 228-65202-58 | Column Oven CTO-40C, qty 1 | ||
| 228-65306-58 p.12 ↗ | 228-65306-58 | Refractive Index Detector RID-20A, qty 1 | ||
| 223-62700-92 p.12 ↗ | 223-62700-92 | LC Workstation LabSolutions, qty 1 | ||
| 223-19189-92 p.12 ↗ | 223-19189-92 | GPC Software LabSolutions GPC Software, qty 1 | ||
| 228-20803-91 p.11 ↗ | 228-20803-91 | Shim-pack GPC-801 column, 300 × 8.0 mm, exclusion limit 1.5 × 10³ | ||
| 228-20808-91 p.11 ↗ | 228-20808-91 | Shim-pack GPC-805 column, 300 × 8.0 mm, exclusion limit 4 × 10⁶ |