c10g-e097
Technical Reference Summary
Covered Models & Document Scope
Folleto de aplicación sobre la caracterización y el control de calidad de oligonucleótidos mediante instrumentos analíticos Shimadzu: MALDI-MS (MALDI-8030, MALDI-8020, MALDImini-1), LC-MS (LCMS-2050, LCMS-2020, LCMS-9050, LCMS-9030, LCMS-QTOF), HPLC Nexera XS inert / Nexera UC / Nexera Prep, y software LabSolutions / LabSolutions MD / LabSolutions Insight Biologics. No se incluye ninguna hoja de datos de modelo de producto con códigos de pedido.
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Flow rate | 0.3 mL/min p.14 ↗ | LCMS single quadrupole, 70-mer DNA analysis |
| Mobile phase A composition | 95.4 mM HFIP, 7.1 mM TEA in water p.14 ↗ | LCMS single quadrupole, 70-mer DNA analysis |
| Detection wavelength range | 200-400 nm p.14 ↗ | PDA detection, 70-mer DNA analysis |
| Ionization mode | ESI/APCI (DUIS), Negative mode p.14 ↗ | LCMS single quadrupole, 70-mer DNA analysis |
| Interface voltage | -2.0 kV p.14 ↗ | LCMS single quadrupole, 70-mer DNA analysis |
| Flow rate | 0.3 mL/min p.16 ↗ | LCMS single quadrupole, FLP impurity analysis |
| UV detection wavelength range | 190-400 nm p.16 ↗ | LCMS single quadrupole, FLP impurity analysis |
| Injection volume | 6 µL p.16 ↗ | LCMS single quadrupole, FLP impurity analysis |
| Scan mode | Scan (m/z 550-2000) p.16 ↗ | LCMS single quadrupole, FLP impurity analysis |
| Nebulizing gas flow | 3.0 L/min p.16 ↗ | LCMS single quadrupole, FLP impurity analysis |
| Mobile phase A composition | 100 mmol/L HFIP, 10 mmol/L TEA in water p.18 ↗ | Nexera XS inert, oligonucleotide impurity analysis |
| Column temperature | 60 °C p.18 ↗ | Nexera XS inert, oligonucleotide impurity analysis |
| Drying gas flow | 5.0 L/min p.18 ↗ | LCMS single quadrupole, oligonucleotide impurity analysis |
| Column temperature | 60 °C p.13 ↗ | Nexera XS inert, co-injection pretreatment |
| Co-injection volumes | 1, 3, 5 µL p.13 ↗ | Nexera XS inert, co-injection pretreatment |
| Column temperature | 40 °C p.20 ↗ | Reversed-phase ion-pair chromatography optimization |
| Flow rate | 0.35 mL/min p.20 ↗ | Reversed-phase ion-pair chromatography optimization |