Covered Models & Document Scope
Nota de aplicación que cubre un método rápido de GC/MS/MS para 13 esteroides anabólicos androgénicos en orina utilizando una columna Agilent J&W FactorFour VF-5ms (10 m x 0.15 mm x 0.15 µm, número de referencia CP9034) y cartuchos de SPE BondElut C18 (200 mg).
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Column | VF-5ms, 10 m x 0.15 mm x 0.15 µm (part number CP9034) p.2 ↗ | |
| SPE cartridge | BondElut C18, 200 mg p.2 ↗ | |
| Sample volume | 3 µL p.2 ↗ | |
| Carrier gas flow | 0.5 mL/min p.2 ↗ | Helium, constant flow |
| Injector temperature | 250 ºC p.2 ↗ | split ratio 1:10 |
| Temperature gradient initial | 170 ºC p.2 ↗ | for 0.5 min |
| Temperature gradient ramp 1 | 10 ºC/min p.2 ↗ | to 260 ºC |
| Temperature gradient ramp 2 | 50 ºC/min p.2 ↗ | to 320 ºC (1 min) |
| Detector | Triple quadrupole GC, 70 eV EI Mode, ion source 250 ºC p.2 ↗ | |
| Urine sample volume | 2 mL p.2 ↗ | |
| Sample pH | 6 p.2 ↗ | buffered |
| Incubation temperature | 55 ºC p.2 ↗ | 1 hour after addition of 50 µL ß-glucuronidase |
| Derivatization temperature | 60 ºC p.2 ↗ | 20 minutes with 50 µL MSTFA/NH4I/dithioerythritol |
| Internal standard concentration | 200 ng/mL p.2 ↗ | 17a-Methyltestosterone (ISTD) |
| Detection level Clenbuterol | 2 ng/mL p.2 ↗ | |
| Detection level 19-Norandrosterone | 2 ng/mL p.2 ↗ | |
| Detection level Epimethenediol | 2 ng/mL p.2 ↗ | |
| Detection level 19-Noretiocholanolone | 2 ng/mL p.2 ↗ | |
| Detection level 17-Epimethanedienone | 2 ng/mL p.2 ↗ | |
| Detection level 5a-Methyltestosterone | 2 ng/mL p.2 ↗ | |
| Detection level 5b-Methyltestosterone | 2 ng/mL p.2 ↗ | |
| Detection level Norethandrolone metabolite | 5 ng/mL p.2 ↗ | |
| Detection level Ethisterone | 2 ng/mL p.2 ↗ | |
| Detection level Bolasterone | 5 ng/mL p.2 ↗ | |
| Detection level Calusterone | 5 ng/mL p.2 ↗ |
Operating Limits
- Maximum GC oven temperature in gradient: 320 ºC p.2 ↗
- Minimum GC oven temperature in gradient: 170 ºC p.2 ↗
- Incubation temperature: 55 ºC p.2 ↗
- Derivatization temperature: 60 ºC p.2 ↗
- Sample pH: 6 p.2 ↗