Agilent Inert Plus GC/MS System with Hydrogen Carrier Gas
Contenido
Overviewpp. 1-3
Introduction to hydrogen carrier gas transition and system benefits
Qualitypp. 4-6
Spectral fidelity and library match results with hydrogen
Qualitypp. 7-8
Source maintenance reduction and sensitivity performance data
Specpp. 9-17
Application results for hydrocarbons, PAHs, and pesticides
Specp. 18
VOC analysis performance with hydrogen carrier
Orderingpp. 19-20
Upgrade kit configurations and complete source options
p. 21
Summary of hydrogen carrier gas benefits
p. 20table, 10 rows
p. 7table, 21 rows
p. 9table, 7 rows
p. 19table, 17 rows
p. 4table, 2 rows
p. 11table, 22 rows
p. 12table, 40 rows
p. 18table, 4 rows
Table 9. Agilent HydroInert triple quadrupole GC/MS upgrade kit for the Agilent 7000C/D/E triple quadrupole mass spectrometers, with part numbers and item numbers indicated in Figure 15.
Item Number (Figure 15)
Part Name
Part Number
NA
HydroInert GC/TQ upgrade
5505-0084
17
Insert*
G7078-20901
12
Repeller*
G7078-20902
3
Extractor source body*
G7078-20903
9**
Entrance lens –extended*, **
G7006-60926
10
Ion focus lens*
G7078-20905
4
Extractor lens–9 mm*
G7078-20909
16
H2 EI heater/sensor assembly*
G7078-67920
NA
Wire, extractor lens
G7000-60827
11
Repeller insulator (x2)
G1099-20133
* HydroInert-specific parts. ** The extended entrance lens for the HydroInert MS/MS system is longer than the entrance lens of a single quadrupole MS system.
Table 2. Comparison of LMS for a mixture of SVOCs, pesticides, and fragrance compounds using an extractor source with He carrier gas, an extractor source with H2 carrier gas, and an Agilent HydroInert source with H2 carrier gas.
Analyte
Extractor He LMS
Extractor H2 LMS
HydroInert H2 LMS
Aniline
98.6
92.3
97.7
1,4-dichlorobenzene-d4
96.7
94.7
97.5
Nitrobenzene
95.7
80.9
94.3
Benzoic acid
93.3
87.7
97.2
Naphthalene-d8
96.9
93.9
97.6
Acenaphthene-d10
98.3
97.3
97.8
2,4-dinitrophenol
95.6
90.6
94.8
4-nitrophenol
89.5
83.1
94.8
4,6-dinitro-2-methylphenol
94.3
93.6
95.7
Pentachlorophenol
90.7
85.4
89.1
4-aminobiphenyl
96.9
95.8
97.9
Phenanthrene-d10
97.7
93.3
97.2
Musk ambrette
86.6
86.3
89.5
Fenitrothion
95.6
88.3
97.0
Musk ketone
95.8
89.8
98.2
Benzidine
92.6
70.1
97.5
3,3'-dichlorobenzidine
97.3
89.8
95.3
Chrysene-d12
96.0
84.8
93.5
Benzo[b]fluoranthene
97.8
70.1
98.4
Benzo[k]fluoranthene
97.8
96.8
98.4
Perylene-d12
94.4
79.4
93.1
Table 3. Tested compound classes, including example compounds, with performance abilities for hydrogen carrier gas with a typical Inert Plus source and the Agilent HydroInert source.
– "Unacceptable" was declared when mass spectral infidelity or compound misidentification was observed. – "Neutral" was determined when results were similar across the normal GC/MS system and HydroInert source in the presence of hydrogen carrier gas. – "Differentiating" was stated when the HydroInert source retained mass spectral fidelity and compounds were identified correctly with a helium library as reference.
Table 7. List of parts in a fully assembled Agilent HydroInert source for an Agilent Inert Plus GC/MS system.
Item Number (Figure 15)
Part Name
Part Number
1
Setscrews
G3870-20446
2
Screws
G3870-20021
3
Extractor source body*
G7078-20903
4
Extractor lens–9 mm*
G7078-20909
5
Extractor lens insulator
G3870-20445
6
Filament
G7005-60061
7
Spring washer
Flat washer
3050-1301
3050-0982
8
Lens insulator
G3870-20530
9
Entrance lens*
G7078-20904
10
Ion focus lens*
G7078-20905
11
Repeller insulator
G1099-20133
12
Repeller*
G7078-20902
13
Flat washer
3050-0891
14
Belleville spring washer
3050-1301
15
Repeller nut
0535-0071
16
H2 EI heater/sensor assembly*
G7078-67910
17
Insert*
G7078-20901
* HydroInert-specific parts.
Table 1. Comparison of extractor and Agilent HydroInert sources when using H2 carrier gas for identification of DDT and DDT breakdown products.
Source with H2 Carrier Gas
Retention Time (min)
Identified Compound
Library Match Score
Extractor
13.04
DDD
79
HydroInert
13.04
DDT
85
Table 5. Response Factors (RFs) for select compounds from EPA method 8270E (Table 4 in the EPA method)4, GC/MS single quadrupole analysis with He carrier gas, GC/MS single quadrupole analysis with Agilent HydroInert source and H2 carrier gas, and GC/MS/MS tandem quadrupole analysis with HydroInert source and H2 carrier gas.
Compound
Response Factor (RF) from EPA Method 8270E4
RF He GC/MS
RF H2 HydroInert GC/MS
RF H2 HydroInert GC/MS/MS
Acenaphthene
0.9
1.3
1.1
0.2
Acenaphthylene
0.9
1.9
1.4
0.1
Acetophenone
0.01
1.2
0.4
1.0
Anthracene
0.7
1.1
1.0
0.9
Benzo(a)anthracene
0.8
1.4
1.5
1.0
Benzo(a)pyrene
0.7
1.2
0.9
0.9
Benzo(b)fluoranthene
0.7
1.4
1.2
1.2
Benzo(g,h,i)perylene
0.5
1.1
1.0
1.3
Benzo(k)fluoranthene
0.7
1.2
1.2
1.3
Bis(2-chloroethoxy)methane
0.3
0.4
0.3
0.7
Bis(2-chloroethyl)ether
0.7
0.8
0.6
0.5
Bis-(2-ethylhexyl)phthalate
0.01
0.8
0.5
0.2
4-Bromophenyl-phenyl ether
0.1
0.3
0.2
0.2
Butyl benzyl phthalate
0.01
0.6
0.3
0.1
4-Chloroaniline
0.01
0.4
0.4
0.6
4-Chloro-3-methylphenol
0.2
0.3
0.2
0.4
2-Chloronaphthalene
0.8
2.4
1.0
0.8
2-Chlorophenol
0.8
0.8
0.7
0.5
4-Chlorophenyl-phenyl ether
0.4
0.7
0.5
0.3
Chrysene
0.7
1.2
1.1
0.4
Dibenz(a,h)anthracene
0.4
1.1
1.0
0.2
Dibenzofuran
0.8
1.7
1.5
1.4
Response Factor (RF) from EPA Method 8270E4
Compound
Response Factor (RF) from EPA Method 8270E4
RF He GC/MS
RF H2 HydroInert GC/MS
RF H2 HydroInert GC/MS/MS
Di-n-butyl phthalate
0.01
1.3
0.8
0.8
3,3'-Dichlorobenzidine
0.01
0.5
0.4
0.1
2,4-Dichlorophenol
0.2
0.3
0.2
0.4
Diethyl phthalate
0.01
1.4
1.0
0.6
Dimethyl phthalate
0.01
1.4
1.0
0.8
2,4-Dimethylphenol
0.2
0.3
0.3
0.4
4,6-Dinitro-2-methylphenol
0.01
0.2
0.1
0.01
2,4-Dinitrophenol
0.01
0.2
0.1
0.02
2,4-Dinitrotoluene
0.2
0.4
0.2
0.03
2,6-Dinitrotoluene
0.2
0.3
0.2
0.03
Di-n-octyl phthalate
0.01
1.3
0.8
0.4
Fluoranthene
0.6
1.2
1.2
0.4
Fluorene
0.9
1.3
1.2
1.4
Hexachlorobenzene
0.1
0.3
0.3
0.4
Hexachlorobutadiene
0.01
0.2
0.2
0.4
Hexachlorocyclopentadiene
0.05
0.3
0.1
0.03
Hexachloroethane
0.3
0.2
0.1
0.14
Indeno(1,2,3-cd)pyrene
0.5
1.2
1.2
1.0
Isophorone
0.4
0.6
0.4
0.3
2-Methylnaphthalene
0.4
0.7
0.7
1.7
2-Methylphenol
0.7
0.7
0.6
0.5
4-Methylphenol
0.6
1.0
0.3
0.7
Naphthalene
0.7
1.1
1
0.9
2-Nitroaniline
0.01
0.4
0.23
0.05
3-Nitroaniline
0.01
0.3
0.18
0.03
4-Nitroaniline
0.01
0.3
0.21
0.13
Nitrobenzene
0.2
0.3
0.2
0.3
2-Nitrophenol
0.1
0.2
0.11
0.1
4-Nitrophenol
0.01
0.2
0.14
0.05
N-Nitroso-di-n-propylamine
0.5
0.4
0.4
0.03
N-Nitrosodiphenylamine
0.01
2.05
0.9
2.3
2,2'-Oxybis-(1-chloropropane)
0.01
0.45
0.54
0.03
Pentachlorophenol
0.05
0.18
0.1
0.1
Phenanthrene
0.7
1.2
1.1
1.1
Phenol
0.8
0.9
0.7
0.3
Pyrene
0.6
1.3
1.2
0.4
1,2,4,5-Tetrachlorobenzene
0.01
0.38
0.31
0.23
2,3,4,6-Tetrachlorophenol
0.01
0.36
0.17
0.07
2,4,5-Trichlorophenol
0.2
0.3
0.3
0.2
2,4,6-Trichlorophenol
0.2
0.3
0.2
0.2
Table 6. NIST helium library scores for select low-level (µg/L) trihalomethanes analyzed with H2 carrier gas and the Agilent HydroInert source.
Retention Time (minutes)
NIST LMS
Trichloromethane
2.107
93
Bromodichloromethane
2.792
86
Dibromochloromethane
3.526
88
Tribromomethane
4.267
80
Engineering Facts
14 facts · every claim linked to a page
Installation
Stainless steel tubing is strongly recommended for hydrogen service; if only copper is available, fresh copper lines must be installed.
Material compatibility and cleanliness of the gas path directly affect hydrogen carrier gas reliability and contamination risk.
Source
“It is highly recommended that stainless steel tubing be used with H2. If only copper tubing is available, install fresh copper lines.”
Installation
Gas filters are mandatory, especially with hydrogen generators, using the specified Agilent trap and carrier gas purifier part numbers.
Filtration protects the column and MS source from moisture, oxygen, and hydrocarbons that degrade performance with H2.
Source
“Always use gas filters, especially with H2 generators.”
Operation
Hydrogen's lower viscosity reduces inlet pressure, so column dimensions should be changed or carrier flow increased, with 0.5–1.0 mL/min optimal and 2.0 mL/min maximum for turbo pumps.
Flow and pressure must be re-optimized when switching from helium to avoid poor chromatography or pump overload.
Source
“Optimal range: 0.5 to 1.0 mL/min”
Operation
For GC/MS/MS with hydrogen, quench gas must be disabled by removing helium plumbing and setting quench gas to 0.00 mL/min in the method.
Leaving helium quench gas connected with hydrogen carrier will cause method errors and incorrect MS/MS operation.
Source
“Please remove the helium plumbing from the back of the electronics pressure control module (EPC) and set the quench gas to 0.00 mL/min in any GC/MS/MS method.”
Installation
HydroInert source parts must never be mixed with extractor source parts, especially when using hydrogen carrier gas.
Part intermixing can cause hydrogenation/dechlorination and invalidate mass spectral fidelity.
Source
“Do not mix the HydroInert source parts with the extractor source parts; this can cause unexpected and unwanted results, especially when using hydrogen carrier gas.”
Selection
The HydroInert source is not compatible with older MS models such as the 5975 Series, nor with stainless steel or inert sources.
This determines whether an existing instrument can be upgraded or a new system is required.
Source
“The HydroInert source is not compatible with older mass spectrometer models, such as the Agilent 5975 Series GC/MSD, nor with stainless steel or inert sources.”
Selection
Diffusion pump single quadrupole instruments are not supported; a turbo pump is required for the HydroInert source.
Pump type is a hard selection criterion before purchasing or installing the HydroInert source.
Source
“Diffusion pump single quadrupole instruments are not supported. A turbo pump is required.”
Installation
Before installing the HydroInert source, verify hardware support and ensure required components (transfer line tip seal, source wires, green and brown wires) are present on older systems.
Missing electrical isolation or wiring components will prevent proper source installation and operation.
Source
“Before installing the HydroInert ion source, verify that the instrument hardware supports use of this source.”
Maintenance
With hydrogen and the HydroInert source, 5,200 matrix injections were completed before source failure versus 365 for helium, and the extraction lens and repeller are recommended for replacement when dirty.
This sets realistic maintenance intervals and spare-parts planning for high-throughput hydrogen operation.
Source
“For the HydroInert source, 5,200 matrix injections were completed before the source failed acceptance criteria.”
Performance
The HydroInert source retains mass spectral fidelity so existing helium-based libraries such as NIST can still be used with hydrogen carrier gas.
Library compatibility avoids rebuilding compound libraries and simplifies method transfer.
Source
“HydroInert source with H2 carrier gas retains mass spectral fidelity and can allow users to continue to use existing He-based mass spectral libraries.”
Performance
For GC/MS/MS with hydrogen, four OFN IDL datasets averaged 2.9 fg, all passing the 4.0 fg maximum allowable IDL criterion.
Sensitivity performance must be verified against method IDL requirements before adopting hydrogen.
Source
“The four IDL values were 2.1, 4.0, 2.0, and 3.5 fg for an average of 2.9 fg IDL, all of which pass the criteria.”
Operation
Not every compound, matrix, or analysis will improve with hydrogen; a limited-capacity test is advised before transferring multiple instruments.
Pilot testing prevents costly site-wide conversion failures for incompatible applications.
Source
“Not every compound, matrix, nor analysis will have the same or improved results, and should be tested in a limited capacity before transferring multiple instruments to hydrogen.”
Selection
For VOCs, a 6 mm HydroInert extraction lens is recommended, though a 9 mm HydroInert lens may also be an option.
Lens choice affects sensitivity and spectral fidelity for the target compound class.
Source
“For VOCs, a 6 mm HydroInert extraction lens is recommended, but a 9 mm HydroInert extraction lens may also be an option.”
Installation
Triple quadrupole and single quadrupole HydroInert parts must be kept separate because entrance lenses differ in length and must not be swapped.
Swapping lenses between instrument types causes installation errors and degraded performance.
Source
“the triple quadrupole parts should be kept separate from the single quadrupole parts as the entrance lenses are different lengths and should not be swapped.”