Overview of the 4500 series FTIR instruments and their features
Procedurepp. 7-16
Packing list, power connection, power cord selection, startup, and diagnostics
Procedurepp. 17-38
Sample analysis and accessory cleaning for TumblIR, ATR, and DialPath
Orderingpp. 37-38
Spare parts list for the instrument and accessories
Specpp. 39-41
Technical specifications and safety information
p. 14table, 6 rows
p. 37table, 2 rows
p. 12table, 8 rows
Table 2. Diagnostic values
Value
Optimal (green)
Marginal (yellow)
Critical (red)
Comment
Energy (centerburst)
26,000–21,000
>30,000 or <15,000
>31,000 or <10,000
Indicates the overall alignment and proper gain adjustment for the system.
Battery life
AC or >30 min
30–5 min
<5 min
Source
1.9 A
>2.2 or <1.6
>2.5 or <1.0
Indicates a problem with the source control voltage or that the source has burnt out. Both voltage and amperage are displayed; however, amperage provides a sufficient diagnostic.
Laser signal
15,000–3,000
>15,000 or <3,000
>17,000 or <2,000
Can verify gross alignment errors even if the reflectance cap is not in place.
Detector temperature
35–44
<35 or >44
<30 or >48
Indicates a problem with the cooling circuitry or an ambient temperature over the specified range.
CPU temperature
10–75
>75
>80
Indicates the ambient temperature over the specified range.
Table
Part number
Description
430 - 0001
Agilent 4500 Series FTIR power supply
430 - 0018
Agilent 4500 Series FTIR USB cable
Table 1. Power switch LED status
Status
LED color
Action (% duty cycle)
System OFF
No LED illumination
N/A
System OFF
Red
100%
System startup
Red/green
Green 0.5 s on / Red 0.5 s on
System ON
Green
100%
Low battery
Green
50% (0.5 s on / 0.5 s off)
Critically low battery
Red
50% (0.5 s on / 0.5 s off)
Dead battery
No LED illumination
N/A
Firmware update
Red/green
Green blinks rapidly twice then red blinks rapidly twice
Engineering Facts
14 facts · every claim linked to a page
Safety
The instrument is not intrinsically safe, so it must only be used in atmospheres verified free of flammable materials.
Determines whether the analyzer can be installed in a classified or potentially explosive area.
Source
“4500 Series FTIR systems are NOT intrinsically safe. Use the system only in atmospheres that have been tested for flammable materials.”
Installation
The shipment must be allowed to equilibrate to room temperature for several hours before opening to prevent condensation on the optics.
Prevents condensation damage to the interferometer and optics during initial setup.
Source
“place the shipment in a room-temperature environment and allow several hours for the contents of the container to reach the room ambient temperature.”
Installation
The instrument must be placed on a flat, stable surface away from hot surfaces and electromagnetic interference sources.
Defines siting constraints that affect measurement stability and interference.
Source
“place it on a flat, stable surface. The instrument must be kept away from hot surfaces and any sources of electromagnetic interference.”
Installation
The power cord must be connected to the spectrometer first and then to the outlet to avoid startup damage.
Wrong connection order can cause power startup damage to the electronics.
Source
“connect the power cord to the spectrometer system first and then to the power outlet.”
Installation
A UL-approved surge protection strip must always be used between the outlet and the system power cord.
Protects the instrument from power surges and faulty power sources.
Source
“always use a UL approved power surge protection strip between the power outlet source and the system power cord.”
Operation
The power button is a momentary switch requiring a two-second hold to turn the system on or off.
Prevents accidental power-off and explains why a short press does nothing.
Source
“In order to turn the system on or off, hold the button down for two (2) seconds.”
Operation
The spectrometer needs a 5-minute warmup before reliable analysis, and 30 minutes before performance validation tests.
Sets the minimum wait time before collecting valid data after power-on.
Source
“The spectrometer must have a warmup period of 5 minutes before reliable analysis may begin.”
Performance
Diagnostic values should be checked daily or weekly, and performance validation quarterly or bi-annually, more often in regulated industries.
Defines the routine verification schedule needed to trust measurement results.
Source
“the diagnostic values should be checked on a daily or weekly basis.”
Maintenance
ZnSe windows are damaged by samples outside pH 4 to 9, and only soft cotton swabs with acetone should be used for cleaning.
Determines which samples are compatible and how to clean without destroying the windows.
Source
“ZnSe is relatively chemically-resistant to materials with a pH range between 4 and 9, but there are some materials, such as strong acids or heavily basic materials, that can harm ZnSe.”
Operation
The TumblIR must not be used with solid samples because they will damage the ZnSe windows or alter the fixed pathlength.
Prevents accessory damage and loss of the factory-set 100 µm pathlength.
Source
“the TumblIR is not to be used with any solid samples, such as hard graphite lubricant. Use of the TumblIR with solid samples will damage the ZnSe windows.”
Selection
The single-reflection ATR suits high-absorbing, powder and solid samples and limited sample amounts, while the triple-reflection ATR suits low-absorbing liquids.
Guides the choice between single- and triple-reflection ATR configurations for the sample type.
Source
“The single reflection ATR is most suitable for higher absorbing samples such as rubber, polymers, paints and fibers.”
Operation
The ATR sample press applies 15 pounds of force assuming a 0.03 in sample thickness and must not be used on sharp or pointed samples.
Defines the contact pressure needed for good spectra and the risk of cracking the thin diamond.
Source
“The spring-loaded sample press tip is set to apply 15 pounds of force assuming a 0.03 in sample thickness.”
Operation
The background must be measured at the same pathlength as the sample on the DialPath.
Prevents invalid results from mismatched background and sample pathlengths.
Source
“The background measurement should be taken at the same pathlength that the sample will be measured at.”
Maintenance
Breaking the spectrometer seal to clean interior surfaces voids the warranty.
Warns service personnel not to open the enclosure for cleaning.
Source
“Do not break the spectrometer seal and attempt to clean interior surfaces. Breaking the seal will void the warranty.”