Data Sheet Model MKF COâ 240
Technical Reference Summary
Covered Models & Document Scope
Cámara climática dinámica de CO₂ modelo MKF 240 para cambios rápidos de temperatura con control de humedad. Dos variantes: MKF240CO2-400V (n.º de art. 9020-0514, 400 V 3~/N/PE 50 Hz) y MKF240-480V (n.º de art. 9020-0516, 480 V 3~/N/PE 60 Hz).
Technical Data
| Parameter | Value | Condición |
|---|---|---|
| Temperature range | -50...180 °C p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Temperature range | -50...180 °C p.3 ↗ | MKF240-480V, 60 Hz |
| Temperature uniformity dependent on set | 0.1...1.2 ± K p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Temperature fluctuation dependent on set | 0.1...0.4 ± K p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Cooling down time from 180°C to -40°C | 60 min p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Temperature range with humidity | 10...95 °C p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Humidity range | 10...98 % RH p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Max. heat compensation at 25°C and 90% RH | 900 W p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Rated voltage | 400 V p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Rated voltage | 480 V p.3 ↗ | MKF240-480V, 60 Hz |
| Power frequency | 50 Hz p.3 ↗ | MKF240CO2-400V |
| Power frequency | 60 Hz p.3 ↗ | MKF240-480V |
| Nominal power | 6,8 kW p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Unit fuse | 16,0 A p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Phase (nominal voltage) | 3~/N/PE p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Interior volume | 228 L p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Net weight of the unit (empty) | 327 kg p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Net weight of the unit (empty) | 335 kg p.3 ↗ | MKF240-480V, 60 Hz |
| Load per rack | 30 kg p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Permitted load | 70 kg p.3 ↗ | MKF240CO2-400V, 50 Hz |
| Width net | 1,115 mm p.4 ↗ | MKF240CO2-400V, 50 Hz |
| Height net | 1,713 mm p.4 ↗ | MKF240CO2-400V, 50 Hz |
| Depth net | 925 mm p.4 ↗ | MKF240CO2-400V, 50 Hz |
| Sound-pressure level | 60 dB(A) p.4 ↗ | MKF240CO2-400V, 50 Hz |
| Energy consumption at 25°C and 60% RH | 1,400 Wh/h p.4 ↗ | MKF240CO2-400V, 50 Hz |
Operating Limits
- Temperature range: -50...180 °C p.3 ↗
- Temperature range with humidity: 10...95 °C p.3 ↗
- Humidity range: 10...98 % RH p.3 ↗
- Wall clearance back: 300 mm p.4 ↗
- Wall clearance sidewise: 200 mm p.4 ↗
- Permitted load: 70 kg p.3 ↗
Type Code Decoder
| Position | Token | Meaning |
|---|---|---|
| 1-3 p.1 ↗ | MKF | Series: dynamic climate chamber with humidity control |
| 4-6 p.1 ↗ | 240 | Model size (interior volume class, 228 L) |
| suffix p.1 ↗ | CO2 | Refrigerant R744 (natural CO₂) |
| voltage/frequency field p.2 ↗ | 400V | 400 V 3~/N/PE, 50 Hz variant (Art.-No. 9020-0514) |
| voltage/frequency field p.2 ↗ | 480V | 480 V 3~/N/PE, 60 Hz variant (Art.-No. 9020-0516) |
Ordering Data
| Ordering Code | Part No. | Variant / Code | Aplica a | Note |
|---|---|---|---|---|
| MKF240CO2-400V p.2 ↗ | 9020-0514 | 400 V 3~/N/PE, 50 Hz, Standard version | Main unit variant | |
| MKF240-480V p.2 ↗ | 9020-0516 | 480 V 3~/N/PE, 60 Hz, Standard version | Main unit variant |