Induztr
Resumen técnico Visor de documentos Modelos relacionados

Data Sheet Model VDL 56

Technical Reference Summary

Covered Models & Document Scope

Cámara de secado por vacío BINDER VDL 56 para disolventes inflamables, volumen interior de 55 L, en dos variantes de tensión: VDL056-230V (Art.-No. 9630-0010, 200...230 V 1~ 50/60 Hz, enchufe CEE 7/7) y VDL056-120V (Art.-No. 9630-0014, 100...120 V 1~ 50/60 Hz, enchufe NEMA 5-15), ambas en la versión de modelo Standard.

Technical Data

Parameter Value Condición
Temperature range +9 °C above ambient temperature to 110 °C p.2 ↗ VDL 56
Temperature uniformity at 100°C 1.2 ± K p.2 ↗ VDL 56
Temperature fluctuation at 100°C 0.1 ± K p.2 ↗ VDL 56
Heating up time to 100°C 140 min p.2 ↗ VDL 56
Leak rate 0,01 bar/h p.2 ↗ VDL 56
Pressure fluctuation over time at 50 mbar 1 ± mbar p.2 ↗ VDL 56
Pressure fluctuation over time at 400 mbar 70 ± mbar p.2 ↗ VDL 56
Rated voltage 200...230 V p.2 ↗ VDL056-230V
Rated voltage 100...120 V p.2 ↗ VDL056-120V
Power frequency 50/60 Hz p.2 ↗ VDL 56
Nominal power 1,4 kW p.2 ↗ VDL 56
Unit fuse 8,0 A p.2 ↗ VDL056-230V
Unit fuse 16,0 A p.2 ↗ VDL056-120V
Phase (nominal voltage) 1~ p.2 ↗ VDL 56
Vacuum connection with small flange 16 DN mm p.2 ↗ VDL 56
Measuring access port with small flange 16 DN mm p.2 ↗ VDL 56
Interior volume 55 L p.2 ↗ VDL 56
Net weight of the unit (empty) 104 kg p.2 ↗ VDL 56
Load per rack 20 kg p.2 ↗ VDL 56
Permitted load 60 kg p.2 ↗ VDL 56
Wall clearance back 100 mm p.2 ↗ VDL 56
Wall clearance sidewise 70 mm p.2 ↗ VDL 56
Width net 638 mm p.3 ↗ VDL 56
Height net 815 mm p.3 ↗ VDL 56
Depth net 461 mm p.3 ↗ VDL 56

Operating Limits

  • Temperature range: +9 °C above ambient temperature to 110 °C p.2 ↗
  • ATEX classification: EX II 2/3/- G IIB T3 Gb/Gc/- X p.1 ↗
  • Wall clearance back: 100 mm p.2 ↗
  • Wall clearance sidewise: 70 mm p.2 ↗
  • Permitted load: 60 kg p.2 ↗
  • Load per rack: 20 kg p.2 ↗

Type Code Decoder

Position Token Meaning
1-6 p.1 ↗ VDL056 Model series VDL vacuum drying chamber, size 56
8-10 p.1 ↗ 230 Voltage variant 200...230 V 1~ 50/60 Hz, unit fuse 8,0 A, CEE 7/7 plug, Art.-No. 9630-0010
8-10 p.1 ↗ 120 Voltage variant 100...120 V 1~ 50/60 Hz, unit fuse 16,0 A, NEMA 5-15 plug, Art.-No. 9630-0014

Ordering Data

Ordering Code Part No. Variant / Code Aplica a Note
VDL056-230V p.1 ↗ 9630-0010 200...230 V 1~ 50/60 Hz, 8,0 A, CEE 7/7, Standard
VDL056-120V p.1 ↗ 9630-0014 100...120 V 1~ 50/60 Hz, 16,0 A, NEMA 5-15, Standard