Product introduction and main features of CS 200-H
Specp. 2
Technical data according to DIN 12876
p. 1table, 32 rows
Technical data according to DIN 12876
Characteristic
Value
Operating temperature range
-20...85 °C
temperature set point / display
colour LED Touchscreen
Internal temperature sensor
Pt100
Temperature stability at -10°C
0,5 K
Interface digital
RS232
Safety classification
I / NFL
Heating power
6 kW
Cooling power at ambient temperature 20°C
.
at 15°C
28 kW
at 10°C
20 kW
at 0°C
18 kW
at -10°C
15 kW
at -20°C
9,5 kW
Refrigeration machine
air-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)
R-449A (A1, H280)
Refrigerant quantity
12 kg
CO2 equivalent
16,76 t
Circulation pump: max. delivery
105 l/min
max. delivery pressure
4,7 bar
Pump connection
G1 male
min. filling capacity
225 l
max. filling capacity
245 l
Overall dimensions WxDxH **
1400x1000x1800 mm
Net weight
515 kg
sound pressure level +/- 4 dB(A)
78 dB(A)
Power supply (3 Phase)
400V 3~ 50Hz
max. current (3 Phase)
23 A
Fuse (3 phase)
3x32A
Degree of Protection
IP52
min. ambient temperature
5 °C
max. ambient temperature
32 °C
from Serial-No.:
1.0/24
** Technical details and dimensions are subject to change. No liability is accepted for errors or omissions. Illustrations can deviate from the original.
Engineering Facts
14 facts · every claim linked to a page
Installation
The chiller requires a three-phase 400 V supply with a 3x32 A fuse, so the site electrical infrastructure must be sized accordingly.
Determines the electrical connection, cable sizing and protection device needed at the installation site.
Source
“Power supply (3 Phase) 400V 3~ 50Hz”
Installation
Three-phase devices drawing less than 63 A are delivered with a cable but without a plug, so the customer must provide and fit the plug.
Affects installation planning and the bill of materials for the electrical hook-up.
Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Installation
The unit must be operated within an ambient temperature window of 5 to 32 °C, which constrains where it can be installed.
Room conditions at the installation site must be verified to stay within the allowed ambient range.
Source
“min. ambient temperature 5 °C”
Performance
Cooling capacity is rated at 20 °C room temperature and drops if ambient temperature rises, so capacity must be derated for warmer rooms.
Prevents undersizing the chiller when the actual installation room is warmer than 20 °C.
Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation
Voltage and frequency tolerances are coupled: ±5 % voltage is only allowed with ±2 % frequency, and certain combinations are explicitly forbidden.
The site power quality must be checked against EN60034-1 tolerances before commissioning.
Source
“-5% voltage and +2% frequency -> not allowed!”
Safety
The refrigerant R-449A is classified A1 but carries GHS hazard H280 (pressurised gas), and the unit contains 12 kg with a CO2 equivalent of 16.76 t.
Relevant for refrigerant handling, leak reporting obligations and safety labelling at the site.
Source
“R-449A (A1, H280)”
Operation
The recommended heat transfer fluid is a 50:50 water–monoethylene glycol mixture.
Guides fluid selection to avoid freezing, corrosion or performance issues.
Source
“Recommended thermofluid: Water - Monoethylene Glycol 50:50”
Operation
The tank filling capacity must be between 225 l and 245 l, so the fluid loop volume must be planned within this range.
Determines the required fluid volume and expansion volume when designing the loop.
Source
“min. filling capacity 225 l”
Installation
The pump connection is a G1 male thread, which dictates the mating fittings for the hydraulic connection.
Needed to select correct hoses and adapters for the circuit hook-up.
Source
“Pump connection G1 male”
Operation
Flow rate can be adjusted via a manual bypass valve on the back of the chiller, allowing the user to match the loop resistance.
Important for balancing the hydraulic circuit after installation.
Source
“The flow rate can be adjusted via the manual bypass valve on the backside of the chiller.”
Cumplimiento
The equipment is classified Class A, Group 1 under EN55011 for electromagnetic disturbance, which has implications for the installation environment.
Class A equipment is intended for industrial environments and may require EMC measures in residential areas.
Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Installation
The unit has IP52 protection, so it is protected against dust and dripping water but not against water jets or immersion.
Constrains cleaning methods and placement relative to water sources.
Source
“Degree of Protection IP52”
Maintenance
The condenser is air-cooled with a high-efficiency fan, so adequate air flow and cleanliness around the condenser must be maintained.
Air-cooled condensers need free space and periodic cleaning to sustain cooling capacity.
Source
“Condenser in air-cooled design, performance- optimized by a built-in high-efficiency fan motor.”
Selection
The chiller has a sound pressure level of 78 dB(A) ±4 dB(A), which must be considered for the working environment.
Noise level affects placement and possible need for hearing protection or acoustic measures.