Unistat GL P560w Technical Data Sheet

Contenido

  1. Overview p. 1
    Product introduction and general description of the Unistat GL P560w
  2. Spec pp. 1-2
    Technical data and performance specifications according to DIN 12876
  3. p. 2 table, 17 rows
  4. p. 1 table, 29 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 5,0 bar42 l/min
Pump connectionM38x1,5 male
max. permissible kin. viscosity100 mm²/s
Cooling water connectionG3/4 male
min. cooling water differential pressure1 bar
max. cooling water differential pressure4 bar
max. cooling water pressure6 bar
max. System pressure6 bar
min. filling capacity8,9 l
Filling capacity expansion tank19,9 l
Overall dimensions WxDxH **918x963x1771 mm
Power supply requirement400V 3~ 50Hz
Pressure equipment categoryII
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:1.0/24

** Please respect space requirements. See operating conditions at www.huber-online.com

Technical data according to DIN 12876 — Unistat GL P560w (Order-No.: 5026.0002.01)

ParameterValue
Operating temperature range-50...200 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Interface analogoptional
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power24 kW
Cooling power with Thermooil at 200°C41 kW
Cooling power with Thermooil at 100°C41 kW
Cooling power with Thermooil at 20°C41 kW
Cooling power with Ethanol at 0°C37 kW
Cooling power with Ethanol at -20°C30 kW
Cooling power with Ethanol at -40°C15 kW
Cooling power with Ethanol at -50°C8 kW
Refrigeration machinewater-cooled, natural refrigerant
Refrigerant (ASHRAE, GHS)R-744 (A1, H280)
Global Warming Potential (GWP)1
Circulation pump: MK pump max. delivery201 l/min
Circulation pump: MK pump max. delivery pressure5,3 bar
Delivery at 1,0 bar179 l/min
Delivery at 2,0 bar154 l/min
Delivery at 3,0 bar124 l/min
Delivery at 4,0 bar87 l/min

Значения приведены дословно по тексту страницы; строки 'Cooling power with Thermooil' и 'Cooling power with Ethanol' объединены с соответствующими температурными точками.

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a 400 V three-phase 50 Hz supply, so the site must provide a suitable three-phase power connection.

Determines electrical infrastructure and whether a plug or hard wiring is needed.

Source
“Power supply requirement 400V 3~ 50Hz”
Installation

The cooling water circuit must be supplied at 1–4 bar differential pressure and no more than 6 bar, with G3/4 male connections.

Cooling water supply pressure and connection size must be verified before installation.

Source
“min. cooling water differential pressure 1 bar max. cooling water differential pressure 4 bar max. cooling water pressure 6 bar”
Operation

The unit is designed for cooling water intake between 5 °C and 20 °C, and lower temperatures require insulating the cooling water pipes.

Cooling water temperature affects cooling capacity and condensation risk.

Source
“This temperature control unit is designed to operate with cooling water intake temperature between 5°C and 20°C.”
Performance

Cooling capacity drops as cooling water temperature rises, and cooling water flow rate may need to increase.

Performance must be derated if the available cooling water is warmer than the reference conditions.

Source
“As the cooling water temperature increases, drop in the cooling power should be expected, and also an increased cooling water flow rate possible.”
Installation

The cooling water circuit contains copper, stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM, so only suitable cooling water may be used.

Water quality must be compatible with these materials to avoid corrosion or fouling.

Source
“Materiels used in the cooling water circuit include; copper, Stainless steel 1.4401, MS, PA, PPE, PTFE and EPDM.”
Cumplimiento

Voltage tolerance is ±5% only with a simultaneous frequency tolerance of ±2%, and opposite-sign deviations are not allowed.

Site power quality must be checked against these limits to avoid invalid operating conditions.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation scope and whether a plug must be sourced separately.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

The unit has safety classification III / FL and adjustable overtemperature protection according to DIN 12876.

Safety class and overtemperature protection are relevant for risk assessment and application suitability.

Source
“Safety classification III / FL”
Installation

The pump connection is M38x1.5 male, and pipework or hoses must use ball socket per DIN 3863 and sleeve nut per DIN 3870.

Ensures correct mating of hoses and pipework to the circulation pump.

Source
“Pump connection M38x1,5 male”
Operation

The maximum permissible kinematic viscosity of the medium is 100 mm²/s.

The heat transfer fluid must be selected within this viscosity limit for proper pump operation.

Source
“max. permissible kin. viscosity 100 mm²/s”
Installation

The unit requires a minimum filling capacity of 8.9 l and has a 19.9 l expansion tank.

Determines the required heat transfer fluid volume and expansion allowance.

Source
“min. filling capacity 8,9 l Filling capacity expansion tank 19,9 l”
Installation

The unit must be installed in an ambient temperature range of 5 °C to 40 °C.

Ambient conditions affect reliable operation and must be respected at the installation site.

Source
“min. ambient temperature 5 °C max. ambient temperature 40 °C”
Cumplimiento

The equipment is classified as EN55011 Class A, Group 1 for electromagnetic disturbance.

Class A equipment may require EMC precautions in residential or sensitive environments.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Maintenance

The refrigeration machine is water-cooled and uses natural refrigerant R-744 with GWP 1.

Cooling water supply and refrigerant type affect maintenance and environmental compliance.

Source
“Refrigeration machine water-cooled, natural refrigerant”