Unistat P425 Temperature Control System Datasheet

Contenido

  1. Overview p. 1
    Introduces the Unistat P425 temperature control unit and its main features
  2. Spec p. 2
    Lists technical data and performance parameters per DIN 12876
  3. p. 2 table, 17 rows
  4. p. 1 table, 29 rows

Technical data according to DIN 12876

ParameterValue
Delivery at 2,0 bar48 l/min
Delivery at 2,5 bar30 l/min
Pump connectionM30x1,5 male
max. permissible kin. viscosity50 mm²/s
min. filling capacity4 l
Filling capacity expansion tank3,7 l
Overall dimensions WxDxH **460x554x1453 mm
Net weight185 kg
sound pressure level +/- 4 dB(A)65 dB(A)
Power supply factory configured (3 Phase)400V 3~ 50Hz
max. current (3 Phase)12 A
Fuse (3 phase)3x16A
Pressure equipment categoryArt. 4.3 PED
Degree of ProtectionIP20
min. ambient temperature5 °C
max. ambient temperature40 °C
from Serial-No.:S481361

** Please respect space requirements. See operating conditions at www.huber-online.com

Unistat P425 Technical data according to DIN 12876

CharacteristicValue
Operating temperature range-40...250 °C
Temperature stability at -10°C0,01 K
temperature set point / display5,7" colour Touchscreen
Resolution of display0,01 K
Internal temperature sensorPt100
Sensor external connectionPt100
Interface digitalEthernet, USB (Host u. Device), RS232
digital inputECS ONE
digital outputPOKO ONE
Alarm messageoptic, acoustic, relay
Safety classificationIII / FL
Heating power2 kW
Cooling power with Thermooil at 250°C2,8 kW
Cooling power with Thermooil at 200°C2,8 kW
Cooling power with Thermooil at 100°C2,8 kW
Cooling power with Thermooil at 20°C2,8 kW
Cooling power with Ethanol at 0°C2,5 kW
Cooling power with Ethanol at -20°C1,8 kW
Cooling power with Ethanol at -40°C0,1 kW
Refrigeration machineair-cooled, CFC- and HCFC-free
Refrigerant (ASHRAE, GHS)R-452A (A1, H280)
Global Warming Potential (GWP)2141
Refrigerant quantity2,01 kg
CO2-Äquvivalent4,3 t
Circulation pump: MK pump max. delivery97 l/min
max. delivery pressure3 bar
Delivery at 0,5 bar86 l/min
Delivery at 1,0 bar75 l/min
Delivery at 1,5 bar63 l/min

Engineering Facts

14 facts · every claim linked to a page
Installation

The unit requires a three-phase 400 V supply with a 3x16 A fuse, so the electrical installation must be sized accordingly.

Determines the site power supply, cable, and protection device rating before installation.

Source
“Power supply factory configured (3 Phase) 400V 3~ 50Hz”
Installation

Three-phase devices below 63 A are delivered with cable but without a plug, so a suitable plug must be provided on site.

Affects electrical installation planning and the bill of materials for commissioning.

Source
“Three-phase devices with current consumption less than 63A --> with cable, without plug”
Safety

Leakage current exceeds 3.5 mA, which may require additional protective measures such as a suitable RCD.

Important for electrical safety design and selection of residual current protection.

Source
“Attention: leakage current > 3,5mA”
Installation

The pump connections are M30x1.5 male with Y-bore and require ball socket and sleeve nut fittings per DIN 3863/3870.

Determines the mating hose/fitting hardware needed for the external circuit.

Source
“Pump connection M30x1,5 male”
Installation

The pump connection bore and mating fittings must follow DIN 3863 and DIN 3870, so non-compliant fittings should not be used.

Ensures leak-free and mechanically compatible connections to the external loop.

Source
“Bore shape Y (60°) according to DIN 3863, pipework/flexible tempering hoses: Ball socket according to DIN 3863, sleeve nut according to DIN 3870.”
Operation

Cooling capacity is rated at 20 °C room temperature and may drop at higher ambient temperatures, so derating must be considered.

Affects thermal performance margins in warm plant rooms.

Source
“Output data valid for: Room temperature 20°C. If the ambient temperature rises, the cooling capacity may drop.”
Installation

The allowable ambient temperature range is 5 to 40 °C, so the installation location must stay within these limits.

Prevents operation outside the specified environmental envelope.

Source
“min. ambient temperature 5 °C”
Installation

The maximum ambient temperature is 40 °C, so adequate room ventilation or cooling may be needed.

Ensures the refrigeration machine and electronics are not operated beyond rated conditions.

Source
“max. ambient temperature 40 °C”
Cumplimiento

Voltage and frequency tolerances are ±5% and ±2% respectively, but certain combinations are not allowed, so supply quality must be checked.

Protects the compressor and electronics from unacceptable supply conditions.

Source
“Voltage + / - 5% with a simultaneous frequency tolerance of + / - 2%”
Cumplimiento

The equipment is EMC Class A Group 1, so it may cause radio interference in residential areas and is intended for industrial environments.

Influences site selection and EMC coordination with nearby sensitive equipment.

Source
“Classification (disturbance) to EN55011: Class A, Group 1”
Safety

The refrigerant is R-452A with GHS classification H280 and GWP 2141, so refrigerant handling and leak response procedures are required.

Needed for safety planning, refrigerant labeling, and environmental compliance.

Source
“R-452A (A1, H280)”
Selection

The unit uses 2.01 kg of R-452A corresponding to 4.3 t CO2 equivalent, which matters for F-gas reporting and leak-check obligations.

Determines regulatory obligations for refrigerant-containing equipment.

Source
“CO2-Äquvivalent 4,3 t”
Selection

The pump delivers 97 l/min at zero head but only 30 l/min at 2.5 bar, so the external circuit pressure drop must be matched to the pump curve.

Ensures adequate flow through the external application at its actual pressure drop.

Source
“max. delivery 97 l/min”
Selection

The maximum permissible kinematic viscosity is 50 mm²/s, so the heat transfer fluid must be selected to stay within this limit at low temperatures.

Prevents pump overload and poor circulation when using viscous fluids at low temperature.

Source
“max. permissible kin. viscosity 50 mm²/s”